We present herein the most complete characterization of microneedle (MN) potentiometric sensors for pH transdermal measurements for the time being. Initial in vitro assessment demonstrated suitable analytical performances (e.g., Nernstian slope, linear range of response from 8.5 to 5.0, and fast response time) in both buffer media and artificial interstitial fluid (ISF). Excellent repeatability and reproducibility together with adequate selectivity and resiliency facilitate the appropriateness of the new pH MN sensor for transdermal ISF analysis in healthcare. The ability to resist skin insertions was evaluated in several ex vivo setups using three different animal skins (i.e., chicken, pork, and rat). The developed pH MN sensor was able to withstand from 5 to 10 repetitive insertions in all the skins considered with a minimal change in the calibration graph (<3% variation in both slope and intercept after the insertions). Ex vivo pH measurements were validated by determining the pH with the MN sensor and a commercial pH electrode in chicken skin portions previously conditioned at several pH values, obtaining excellent results with an accuracy of <1% and a precision of <2% in all cases. Finally, pH MN sensors were applied for the very first time to transdermal measurements in rats together with two innovative validation procedures: (i) measuring subcutaneous pH directly with a commercial pH microelectrode and (ii) collecting ISF using hollow MNs and then the pH measurement of the sample with the pH microelectrode. The pH values obtained with pH MN sensors were statistically more similar to subcutaneous measurements, as inferred by a paired sample t-test at 95% of confidence level. Conveniently, the validation approaches could be translated to other analytes that are transdermally measured with MN sensors.
We present herein the most complete characterization of microneedle (MN) potentiometric sensors for pH transdermal measurements for the time being. Initial in vitro assessment demonstrated suitable analytical performances (e.g., Nernstian slope, linear range of response from 8.5 to 5.0, and fast response time) in both buffer media and artificial interstitial fluid (ISF). Excellent repeatability and reproducibility together with adequate selectivity and resiliency facilitate the appropriateness of the new pH MN sensor for transdermal ISF analysis in healthcare. The ability to resist skin insertions was evaluated in several ex vivo setups using three different animal skins (i.e., chicken, pork, and rat). The developed pH MN sensor was able to withstand from 5 to 10 repetitive insertions in all the skins considered with a minimal change in the calibration graph (<3% variation in both slope and intercept after the insertions). Ex vivo pH measurements were validated by determining the pH with the MN sensor and a commercial pH electrode in chicken skin portions previously conditioned at several pH values, obtaining excellent results with an accuracy of <1% and a precision of <2% in all cases. Finally, pH MN sensors were applied for the very first time to transdermal measurements in rats together with two innovative validation procedures: (i) measuring subcutaneous pH directly with a commercial pH microelectrode and (ii) collecting ISF using hollow MNs and then the pH measurement of the sample with the pH microelectrode. The pH values obtained with pH MN sensors were statistically more similar to subcutaneous measurements, as inferred by a paired sample t-test at 95% of confidence level. Conveniently, the validation approaches could be translated to other analytes that are transdermally measured with MN sensors.
Entities:
Keywords:
in vivo monitoring; ion-selective electrodes; microneedle transdermal sensing; pH sensing; validation
Microneedle (MN) sensors are
attracting increasing attention linked to the need for new wearable
diagnostic tools within the healthcare system.[1,2] Indeed,
the intrinsic ability of MN sensors to perform minimally invasive
on-body and real-time measurements is crucial to accelerate the provision
of meaningful observations in view of the next generation of personalized
and preventive medicine. Unlike other formats of wearable chemical
sensors that target “excretable” biological fluids such
as sweat, tears, and saliva (e.g., sweat bands and patches, smart
watches, glasses, and contact lenses),[3−6] MN sensors are primarily conceived for transdermal
monitoring of interstitial fluid (ISF). Advantageously, the ISF composition
has claimed to be very similar to blood, as a consequence of the existing
equilibrium between ISF and plasma carried out by several small molecules
(e.g., albumin, CO2, and phosphates).[2] This similarity in composition is appealing for clinical
purposes in order to substitute current (and inconvenient for the
patient) blood analysis while providing the same physiological outcomes.
In addition, biofouling in ISF is known to be lower than in blood
due to lower concentration of proteins and other large molecules.[2,7,8]The analysis of ISF entails
several challenges rarely addressed
in the literature such as resistance to skin penetration and strict
analytical validation (mostly absent) of the obtained analyte levels.[2] It is essential to unequivocally demonstrate
that the chemical sensing element embedded in the MN is not altered
during skin insertion prior to claim any suitability for further in vivo applications; otherwise, the developed device may
create unreal expectations. Special attention should be paid when
the MN is externally modified with the sensing element because this
is totally exposed to the skin penetration process. Therefore, external
modification presents an increasing probability of detachment/alteration
together with incremental cytotoxicity risk in the monitored subject
compared to internal modification of the MN.[2]Concerning the analytical validation of transdermal measurements
through MNs, the main problem lies in the difficulty of extracting
enough volume of ISF to be measured using a reference analytical technique.
Several strategies have been reported for this purpose including hollow
MNs, effusion, dialysis, sonication, and reverse iontophoresis.[2,9] The latter has been demonstrated for the successful analysis of
glucose but is not valid for all analytes because the iontophoresis
process per se modifies the composition of the collected ISF.[9] By contrast, extraction with hollow MNs preserves
ISF composition as far as pertinent precautions are fulfilled to avoid
sample evaporation and/or external contamination, which is a common
issue whatever the collection method. The main limitation of ISF collection
with MNs is that the volume is very low, usually less than 10 μL.[10,11] Thus, in view of this problem, most in vivo studies
using MN sensors attempt a validation by inducing changes in analyte
concentration[12,13] or comparing with concentration
in blood, which has been widely applied in the case of MNs for glucose
detection.[14,15] Evidently, these strategies cannot
be considered as pure analytical validation methods because they are
either qualitative or do not consider the same fluid.A search
into the literature reveals that in vivo implementation
of MN chemical sensors primarily targets glucose,
with pH being the second analyte most widely detected.[2] While the interest of measuring glucose to manage diabetes
is undeniable, it has been proved that acid/base disorders are closely
related to many diseases such as renal failure, ischemia, multiple
sclerosis, and psychiatric disorders. Therefore, monitoring in vivo dynamic changes in pH has been claimed to be highly
relevant from a clinical perspective.[16] While a great number of pH MN sensors can be found in the literature,
with electrochemical readout but also others,[2,17,18] to the best of our knowledge, only few works
reported on real applications involving the in vivo determination of pH and transdermal measurements in ISF totally
lack.In this context, Mani et al. analyzed pH in mouse cerebrospinal
fluid (CSF) and bladders using tungsten-modified MNs (W/ZnO)[16] while Zuliani et al. developed an iridium oxide
(IrOx) MN sensor for mapping pH distributions in rat heart.[19] Zhou et al. developed a pH sensor based on molybdenum
disulfide (MoS2) nanosheets and polyaniline (PANI)-functionalized
acupuncture needles for real-time monitoring of pH changes in rat
brain.[20] However, these papers did not
cover the determination of pH in ISF and lack the evaluation of a
possible detachment of the sensing element with the transdermal use
of the MN and how this may affect the monitored individual. In this
sense, it is important to mention that all the tungsten compounds
are regarded as highly toxic compounds,[21] whereas very little information is available for IrOx and the PANI-MoS2 tandem despite careful manipulation of these compounds being
generally advised.[22,23] In addition, the high-cost and
high-temperature processing associated to metal oxide-based pH sensors
is known to limit the fabrication of disposable MN sensors.[24]In analogy to solid-contact electrochemical
sensors already well-entrenched
for pH detection in diverse samples, potentiometric ion-selective
electrodes (ISEs) based on plasticized polymeric membranes should
be good candidates toward effective transdermal monitoring of pH in
ISF, beyond the use of inorganic coverings that are pH-responsive
(i.e., W/ZnO, IrOx, and PANI-MoS2).[24,25] However, to the best of our knowledge, this approach has not been
translated to the MN configuration yet. Importantly, our group has
recently demonstrated the huge potential of the ISE technology toward
potassium detection in ISF.[26] Commercially
available stainless-steel solid MNs were externally modified, first
with the solid ion-to-electron transducer and then with the ion-selective
membrane (ISM), following a procedure adapted from the traditional
all-solid-state ISE fabrication.[27] The
proposed MN technology demonstrated excellent resiliency to skin penetration
and the absence of biofouling (in the time scale of hours).[26] In addition, cytotoxicity studies using human
dermal fibroblasts revealed a connection with the leaching of the
ionophore from the ISM to the cell culture with any found toxic effect.[28] Thus, while a toxic impact was detected in the
24–36 h experimental time frame for potassium and ammonium
ISMs (based on valinomycin and nonactin, respectively), no effect
appeared for other ionophores, including hydrogen ionophore I (typically
used in pH sensors).[28] In principle, these
results pointed out the suitability of pH-responsive ISMs to be implemented
into MN technology for transdermal ISF analysis. This approach is
expected to exceed the sensor biocompatibility compared to already
existing pH MN sensors commented above.[2,16,19,20]We present herein
the very first MN sensor applied to the transdermal
detection of pH in ISF in rats. It herein shows a deep characterization
and pioneering validation protocol for pH MN sensors, which is schematized
in Figure and could
pave the way as a general strategy for any MN chemical sensor validation.
In addition, the present work innovates toward proper inquiry of MN
sensing technology through the in vitro, ex vivo, and in vivo journey aiming at
reaching on-body measurements. Initial in vitro assessment
is performed to evaluate the analytical performances of the developed
pH MN sensors in controlled buffer media and artificial ISF (AISF).
Subsequently, the ability of the MN sensor to perform transdermal
measurements is evaluated using several animal skins and following
two protocols. The first one compares pre- and postinsertion calibrations
in the skin to ensure that the sensing element is not damaged. The
second protocol aims at testing the accuracy of transdermal pH measurements
in skin pieces conditioned at different pHs, that is, with known analyte
concentration. Finally, pH MN sensors are tested in rats together
with two different validation approaches based on subcutaneous measurements
with a commercially available microelectrode and ISF extraction with
bare hollow MNs. Advantageously, the procedure to fabricate and validate
the pH MN sensors can be tailored for other ions and molecules, therefore
prospecting fast advances in accurate multianalyte monitoring in ISF
toward in vivo measurements in animals and humans.
Figure 1
Scheme
for the protocol used in the present paper for the characterization
and validation of the pH MN patch.
Scheme
for the protocol used in the present paper for the characterization
and validation of the pH MN patch.
Experimental Section
Preparation of the MN Patch
for the Potentiometric Determination
of pH
The MN patch for pH sensing consists of two solid MNs
acting as working (WE) and reference (RE) electrodes, that are fixed
in a silicon rubber substrate. This configuration is particularly
suitable to access to the wearable technology owing to its flexibility
and biocompatibility. The manufacturing process is illustrated in Figure a. More details on
substrate fabrication and MN fixation are provided in the Supporting Information. The WE is a potentiometric
pH electrode based on a three-layer structure of carbon ink, functionalized
multiwalled carbon nanotubes (f-MWCNTs) as an ion-to-electron transducer
and a hydrogen-selective membrane (HSM) (see Figure b). The RE consists of a Ag/AgCl layer covered by a poly(vinyl
butyral) reference membrane (RM) cocktail (see the Supporting Information), which provides a high and constant
chloride concentration in the solid-state RE, as previously demonstrated
elsewhere.[29,30] The layer-by-layer structure
of the RE is displayed in Figure c, whereas optical images of WE and RE are shown in Figure d,e, respectively.
Transdermal pH measurements in euthanized rats were performed by coupling
the developed MN patch with a portable potentiometer (Figure f). Note that all the rat-based
experiments were carried out at the Karolinska University Hospital
(Stockholm, Sweden) and assisted by the Operation Manager and Karolinska
Experimental Research and Imaging Centre (KERIC) personnel. The rats
were donated by KERIC and consisted of euthanized specimens that were
previously used for other research purposes at KERIC. Importantly,
the animals were not specifically euthanized for the purposes of our
investigations but used as donated.
Figure 2
(a) Manufacturing procedure for the MN
patch containing both the
pH-selective electrode and reference electrode: (1) stainless-steel
solid MNs are modified with carbon and Ag/AgCl ink coatings to further
produce the working (WE) and reference electrode (RE), respectively;
(2) modified MNs are assembled into the silicon substrate; (3) deposition
of f-MWCNTs on top of the carbon ink in the WE, (4) deposition of
the pH-selective membrane (HSM) in the WE; and (5) deposition of the
RM and PU external layer in the RE electrode. (b) Cross-sectional
image of the layers’ arrangement in the WE, namely, stainless-steel
core, carbon ink, f-MWCNTs, and HSM. (c) Cross-sectional image of
the layers’ arrangement in the RE: stainless-steel core, Ag/AgCl
ink, RM, and PU. (d) Image obtained by optical microscopy for the
WE. (e) Image obtained by optical microscopy for the RE. (f) Arrangement
of the entire device composed of the MN patch connected to the hand-made
wireless potentiometric board.
(a) Manufacturing procedure for the MN
patch containing both the
pH-selective electrode and reference electrode: (1) stainless-steel
solid MNs are modified with carbon and Ag/AgCl ink coatings to further
produce the working (WE) and reference electrode (RE), respectively;
(2) modified MNs are assembled into the silicon substrate; (3) deposition
of f-MWCNTs on top of the carbon ink in the WE, (4) deposition of
the pH-selective membrane (HSM) in the WE; and (5) deposition of the
RM and PU external layer in the RE electrode. (b) Cross-sectional
image of the layers’ arrangement in the WE, namely, stainless-steel
core, carbon ink, f-MWCNTs, and HSM. (c) Cross-sectional image of
the layers’ arrangement in the RE: stainless-steel core, Ag/AgCl
ink, RM, and PU. (d) Image obtained by optical microscopy for the
WE. (e) Image obtained by optical microscopy for the RE. (f) Arrangement
of the entire device composed of the MN patch connected to the hand-made
wireless potentiometric board.
Results and Discussion
The fabrication process for the pH
WE was adapted from a recipe
previously reported by our group.[26] Thus,
the deposition of the HSM was optimized considering two different
approaches to incorporate the membrane cocktail: dip-coating and drop-casting.
For both approaches, different membrane thicknesses were evaluated,
which were obtained by changing the number of layers in dip-coating
(3, 5, 10, 15, and 300) and both the number of layers (1–3)
and volume per layer (0.5–3 μL) in drop-casting.Figure a,b depicts
the calibration graphs obtained by separate buffer solutions and using
electrodes prepared by each different deposition protocol. As observed
in Figure a, none
of the proposed configurations fabricated via dip-coating provided
a Nernstian response, with the highest slope (34.3 mV) obtained using
15 layers of the membrane cocktail. Conversely, drop-casting deposition
provided better results for both the slope and the linear range of
response (LRR) (see Figure b). A Nernstian response was found for the configuration prepared
by drop-casting three layers of 1 μL of the membrane cocktail.
This MN sensor presented 54.4 mV dec–1 and 8.5–5.0
as slope and LRR, respectively. The differences between both methods
are likely explained by a better control of the layers’ homogeneity
by drop-casting rather than dip-coating, as suggested by images obtained
using optical microscopy (see Figure S1 in the Supporting Information).
Figure 3
Calibration graphs obtained with two different
deposition methods
for the HSM incorporation: (a) dip-coating and (b) drop-casting. (c)
Calibration curves obtained with the pH MN sensor against a commercial
Ag/AgCl RE and the developed MN RE. (d) Dynamic response and average
calibration graph for the carry-over test (ten consecutive calibrations
by alternatively increasing and decreasing the pH in buffer solutions).
(e) Calibration curves obtained with the MN patch in two different
backgrounds: buffer solutions and AISF at increasing pH values.
Calibration graphs obtained with two different
deposition methods
for the HSM incorporation: (a) dip-coating and (b) drop-casting. (c)
Calibration curves obtained with the pH MN sensor against a commercial
Ag/AgCl RE and the developed MN RE. (d) Dynamic response and average
calibration graph for the carry-over test (ten consecutive calibrations
by alternatively increasing and decreasing the pH in buffer solutions).
(e) Calibration curves obtained with the MN patch in two different
backgrounds: buffer solutions and AISF at increasing pH values.The optimized WE was combined with the MN RE to
conform the final
potentiometric MN-based patch. Notably, the performance of the MN
RE was exhaustively characterized elsewhere, even proving its mechanical
resilience during transdermal insertions.[26]Figure c compares
the calibration graph (n = 3, three successive measurements
using the same electrode) for the pH MN sensor against both a commercial
Ag/AgCl RE (blue line) and the MN RE (black line). A very similar
average slope was found in both cases: 54.6 ± 0.6 mV dec–1 for the MN RE and 53.8 ± 0.9 mV dec–1 for the commercial RE. An offset of 100 mV between both sensors
was observed, which is evidently associated to the change of the reference
electrode as reported elsewhere.[29] As a
result, the MN RE was used in all further measurements.First,
the analytical performance of the developed MN patch was
characterized using an in vitro approach in buffer
media. Between-electrode reproducibility was evaluated by carrying
out one calibration using three different MN patches, obtaining a
% RSD of 0.4 for the slope. Response repeatability, calculated from
three consecutive calibration graphs performed with the same sensor,
showed a % RSD of 2.5 and 4.1 for the slope and intercept, respectively.
Reversibility was evaluated recording 10 successive calibration curves
with increasing and decreasing pH values alternatively (Figure d). The sensor displayed a
variation for the slope and the intercept of 3.5 and 2.4%, respectively,
which reflect a rather good reversibility considering the entire experiment
time scale (almost 2 h) and the rather large pH changes in the experiment.
Fast response time (t95 < 5 s) and
very low long-term drift (1.2 ± 2.1 mV h–1,
in 10 mM HCl for a 16 h experiment, n = 3 sensors,
see Figure S2a in the Supporting Information) were also achieved. Overall, in view of all these results, the
pH MN patch presented excellent analytical features.Next, the
suitability of the MN patch for ISF analysis was evaluated
through a selectivity study considering the major interferences found
in this biological fluid, namely, Na+, K+, Mg2+, Ca2+, glucose, and urea. Table S1 collects the logarithmic selectivity coefficients
(n = 3), calculated by the separate solution method.[31] For all the interferences tested, selectivity
coefficients were lower than the minimum value required for accurate
measurements in ISF, which was estimated considering the highest concentration
of each interference traditionally expected in ISF.[32] Furthermore, the lack of synergistic effects between interferences
was demonstrated by recording several calibration curves in AISF solution.
Thus, Figure e presents
the calibration graphs (n = 3) obtained in either
buffer media (black line) or AISF (blue line). Similar slopes were
observed in both media (54.2 ± 2.9 mV dec–1 in AISF and 54.6 ± 0.6 mV dec–1 in buffer)
with the same LRR (from 8.5 to 5.0), which is indeed wide enough to
cover expected pH values in ISF. Normal pH values in ISF range within
7.35 and 7.45 but may be altered due to the influence of diseases
or health disorders/infections. For example, in the case of cancer,
ISF is known to vary from 6.2 to 6.9.[33] In addition, the long-term drift in AISF was rather acceptable (0.6
mV h–1 for 2 h experiment), as shown in Figure S2b
in the Supporting Information.Subsequently, ex vivo evaluation of the MN patch
for transdermal pH detection in the ISF of different types of animal
skins was accomplished to investigate two key aspects: (i) the response
of the developed MN sensors is not altered and/or the external modification
detached from the MN with the skin insertion; and (ii) appropriate
accuracy of the measurements by means of a carefully designed validation
protocol. One approach widely proposed in the literature to assess
MN sensor resistance to skin penetration consists of using agarose
hydrogels in contact with AISF to mimic the real skin system.[34,35] The resiliency of MN sensors is evaluated by comparing the calibration
parameters before and after the insertion into the hydrogel. Then,
the suitability to use an external calibration for transdermal quantification
of the analyte is assessed by measuring different concentrations in
AISF that reach the hydrogel by diffusion. However, this approach
does not provide a realistic evaluation of the ability of MN sensors
to perform effective skin insertion because the hydrogel is softer
and easier to penetrate than any real skin. In contrast, a closer
approach to in vivo analysis utilizes animal skin
instead.[26,36,37] This methodology
allows for a better evaluation of viscoelasticity and the adherence
of undesired substances to the MN (including biofouling), thus providing
close conditions to the further on-body use in animals and/or humans.In this work, sensor resilience to skin penetration was assessed
by manually inserting the MN patch in pieces of chicken, porcine,
and rat skin several times (1, 3, 5, and 10 times) and registering
the calibration curve in AISF afterward. This was compared with an
initial calibration recorded prior to any insertion. Figure a shows the results obtained
in chicken skin, which is the thinnest skin and the easiest to penetrate
out of the three tested types of skin. No significant changes were
observed in the slope after ten insertions (the initial slope was
52.0 mV dec–1 vs a slope of 51.4 mV dec–1 after ten insertions, a variation coefficient of 1.8%). Furthermore,
if all the calibrations are considered, the obtained % RSD for the
slope and intercept was 2.1 and 1.8, respectively. Thus, it is concluded
that the response of the MN patch is not altered during insertion
in chicken skin. Furthermore, the found RSD lies within the variation
observed in the repeatability studies of the electrodes in the in vitro characterization (2.5 and 4.1 for the slope and
intercept (n = 3) respectively). Accordingly, variations
found in the standard electrode potential of the MNs with consecutive
skin insertions can be likely attributed to an inherit variation of
the sensor response rather than a perturbation due to the insertion
cycles.
Figure 4
Dynamic responses and the corresponding calibration graphs observed
for pH before and after several insertions into (a) chicken, (b) porcine,
and (c) rat skin.
Dynamic responses and the corresponding calibration graphs observed
for pH before and after several insertions into (a) chicken, (b) porcine,
and (c) rat skin.Analogous results were
obtained in porcine skin (Figure b), with the initial slope
of 51.3 mV dec–1 versus a slope of 52.7 mV dec–1 after ten insertions, % RSD of 1.5 and 1.2 for the
slope and intercept considering all the calibration graphs. Significantly,
the appropriate results found in porcine skin are particularly promising
for future application of MN sensors in humans because this has striking
similarities to the human skin in terms of general structure, thickness,
hair follicle content, pigmentation, collagen, and lipid composition.[38] Finally, Figure c displays the results obtained in rat skin. Similar
calibration curves were also obtained after up to five insertions:
with the initial slope of 51.1 mV dec–1 versus a
slope of 50.7 mV dec–1 after five insertions and
% RSD of 2.3 and 2.8 for the slope and intercept considering all the
calibration graphs.When using rat skin in the ex vivo studies (Figure c), the intercept
was found to slightly shift to less positive values after each insertion.
Although the variation after 5 insertions was acceptable, higher number
of insertions would require a recalibration of the sensor to correct
this drift. Notably, on-body experiments in rats herein presented
are based in three skin insertions and the electrodes were calibrated
before and after the transdermal measurements to minimize any error
in the pH quantification arising from a change in the calibration
graph.The absence of any attachment of biological material
in the MNs
due to skin insertion and/or contact with the tissues was confirmed
by microscopic images: a comparison of the images taken before (Figure S1b) and after (Figure S1c) penetration in chicken skin is presented in the Supporting Information. As observed, no particles
or pieces are attached to the MN sensor after the insertion. In addition,
no sign about the deterioration and/or detachment of the sensing element
was detected. This confirms, in turn, the absence of marked variations
in the electrode calibration after several insertions (Figure a), hence pointing out the
absence of biofouling effects.The second part of the ex vivo assays focused
on the validation of transdermal pH measurements with the developed
MN patch once inserted in animal skin and therefore demonstrating
the suitability of an external calibration of the sensor for the pH
quantification in further on-body studies. We designed these tests
on the basis of skin pieces that were conditioned 24 h at different
pH. This experiment is not trivial, and one should not assume that
the intradermal concentration of any analyte can be easily modified
just by direct contact of the skin with an external solution of the
desired analyte concentration. For example, Senel et al. reported
a lower chronoamperometric response for urea when MN sensors were
measuring inside phantom gel than that expected according to the concentration
fixed in external AISF.[39] This behavior
was attributed to the slow diffusion of urea in phantom gel. Similar
conclusions were drawn in our previous studies regarding the determination
of K+ in chicken skin using potentiometric MN sensors.[26] It was demonstrated that the lower potassium
concentration determined inside the skin as compared to the concentration
in the AISF solution that soaked the skin was due to the diffusion
and distribution of the ion driven by the concentration gradient at
the skin–AISF interface.[26]In this paper, we opted for performing experiments by conditioning
the pieces of animal skin with a known analyte concentration for a
24 h process and then measuring the transdermal pH with the MN patch
(Figure a). Alternatively,
the concentration inside the skin can be measured by mincing and digestion
followed by analysis with reference analytical techniques.[37] However, this option is not suitable for the
particular case of pH determination because acid digestion will affect
the measurement. To confirm that 24 h conditioning is enough to reach
the desired pH inside the skin, we collected the ISF by means of a
custom-made system based on hollow MNs and measured the pH with a
commercial micro-pH electrode. The collection device was based on
a commercially available hub with hollow MNs connected to a peristaltic
pump (see the Supporting Information for
more details). We found that the pH in the ISF was very similar to that fixed in the conditioning solution
(see Table S2 in the Supporting Information).
In addition, the pH of the conditioning solution was measured by means
of a commercial pH meter to ensure that pH was retained during 24
h.
Figure 5
(a) Setup for ex vivo measurements performed with
the MN patch inserted in chicken skin conditioned at different pH
values. Inset: scheme of the MN insertion. (b) External calibration
graph and three consecutive potentiometric responses obtained during
the measurement of chicken skin at pH 6.0. (c) Measured and expected
values in ex vivo pH measurements of four chicken
skin pieces conditioned at different pH values.
(a) Setup for ex vivo measurements performed with
the MN patch inserted in chicken skin conditioned at different pH
values. Inset: scheme of the MN insertion. (b) External calibration
graph and three consecutive potentiometric responses obtained during
the measurement of chicken skin at pH 6.0. (c) Measured and expected
values in ex vivo pH measurements of four chicken
skin pieces conditioned at different pH values.Transdermal potentiometric responses inside each type of skin were
recorded by triplicate and pH values were calculated by extrapolating
average potentials into an external calibration performed in AISF
(see Figure b). Figure c displays the results
obtained for four chicken skin pieces conditioned at different pH
values, with black circles representing the values measured by the
MN patch and red lines indicating the expected pH values. The accuracy
and precision of the transdermal measurements using the MN results
in the ex vivo experiments are the perfect preamble
for further animal-based tests based on the on-body use of the new
MN patch.Prior to any in vivo tests in rats
and humans,
it is mandatory to demonstrate the biocompatibility level of the MN
patch ensuring that no toxic effect is caused by direct contact, leaching,
or detachment of the sensing element. In this context, we have recently
reported on the cytotoxicity of all the materials used in the pH MN
sensor herein developed.[26,28] Briefly, cell viability
and cell proliferation studies using fibroblasts were carried out
with bare MNs, MNs coated with either carbon or Ag/AgCl ink, f-MWCNT,
and the HSM. In all cases, after 96 h of incubation, the total number
of counted cells was comparable to control conditions (fibroblasts
immersed in culture media and without the presence of material/compound),
indicating the absence of cytotoxicity effects. However, this outcome
should not be generalized to any MN-based ISE because, unlike HSM,
other membrane compounds, such valinomycin-based membranes for potassium
or nonactin membranes for ammonium, revealed a certain level of cytotoxicity
after 96 h.[28]Then, in vivo tests should come accompanied by
validated measurements using a gold standard technique, although the
literature is really scarce in this issue. Indeed, the implementation
of this validation is not trivial due to the difficulty of extracting
enough volume of the same ISF that is measured by the MN. We explored
two different validation protocols for the pH MN tests in rats. The
first approach measures subcutaneous pH by means of a commercial pH
microelectrode after opening the specimen. The second protocol is
based on ISF extraction using a home-made system based on a commercially
available hollow MN hub (see the Supporting Information for more details). Interstitial pH was measured in seven rats with
different genders and ages (see Table S3 in the Supporting Information) using the newly developed MN patch
and the two validation protocols. All measurements were performed
sequentially in one rat before moving to the next one to minimize
pH changes due to evaporation and/or stop in blood flow. Notably,
the rats were donated by KERIC and consisted of euthanized specimens
that were previously used for other research purposes at KERIC.To implant the MN sensors, the back of the rats was shaved to facilitate
MN insertion and visual inspection (Figure a). Subsequently, the MN patch was inserted
in the rat back and secured with a ring of polyurethane (PU) to prevent
any movement of the MNs during the on-body measurements. Finally,
the patch was connected to a wireless potentiometric board and the
response was recorded until the steady-state potential was reached,
40 s approx. (see Figure b). Three consecutive insertions were performed to obtain
the measurement in triplicate and to confirm that the sensing element
is not detached from the MN.
Figure 6
Pictures of on-body measurements in a euthanized
rat specimen:
(a) shaved rat prepared for the insertion of the MN patch. (b) pH
MN patch coupled with the potentiometric electronic board and providing
transdermal measurements in the rat back. (c) Subcutaneous measurement
of pH using a micro-pH meter. (d) Home-made tool for ISF collection
based on a hollow MN hub connected to a peristaltic pump.
Pictures of on-body measurements in a euthanized
rat specimen:
(a) shaved rat prepared for the insertion of the MN patch. (b) pH
MN patch coupled with the potentiometric electronic board and providing
transdermal measurements in the rat back. (c) Subcutaneous measurement
of pH using a micro-pH meter. (d) Home-made tool for ISF collection
based on a hollow MN hub connected to a peristaltic pump.A total of 7 rats were tested, with the rat number 5 monitored
with two different (and twin) MN patches (Table S3). In none of the cases, we could visually detect any modification
of the MN after transdermal measurements. No significant alteration
was detected in the potentiometric response upon increasing insertions
either. Moreover, we found variation coefficients of less than 1%
in all the tested rats, except for the first measurement round in
rat number 5 and in rat number 6, where ca. 5% of variation was observed.
This higher variation in triplicate measurements specifically obtained
in two specimens is likely associated to the quality of the hand-made
fabricated MN patches: the three measurements accomplished in rat
5 provided values (in this order) of 7.51, 7.03, and 6.60. Then, when
measurements were repeated in the same rat but with a different MN
patch, the provided values were 6.36, 6.48, and 6.48, with a variation
coefficient in the range of the 1% as in the majority of the rats.
Thus, in the case of the first measurement round, the real pH is seemingly
within the two last values, with the very first measurement interpreted
as an outlier.After the transdermal measurements, the first
validation method
was carried out by means of an incision in the rat’s back,
exposing the subcutaneous ISF and measuring the pH with a commercial
microelectrode (Figure c). For the second validation method, ISF was directly extracted
from the rat back with the MN hub (Figure d, see the Supporting Information for more details). ISF extraction was prolonged
for 30 min, obtaining significantly different sample volumes for each
rat, ranging from nonmeasurable volume (almost zero) to more than
100 μL (Table S3 in the Supporting
Information). The pH of the collected ISF was measured with an ultramicro
pH electrode, specifically designed for low-volume sample analysis
(∼0.5 μL). Only for one rat, it was not possible to extract
enough ISF to be measured with the pH meter. Interestingly, the highest
volumes of ISF were extracted from the youngest rats (2 months), which
is likely attributed to both smoother skin and higher hydration level.
This combination enhances hydraulic conductivity of tissue, leading
to spacing tissue fibers that facilitate ISF extraction.[40]Individual correlations between the three
types of measurements
were statistically evaluated to quantify the accuracy of the pH MN
patch: separate paired sample t-tests (also known
as the dependent sample t-test) were carried out.
Notably, since measurements performed on the same rat are treated
as a single pair of observation, differences in pH among different
rats will not influence the test. Graphical representations of the
paired sample t-test are displayed in Figure , with Figure a presenting the direction in the variation
of the pH measurements in each single rat (comparing the MN with each
validation strategy), and Figure b being a box and whisker plot of the measured differences
between the three methods. Attending first to the comparison of MN
results with subcutaneous measurements and considering a 95% of confidence
level, the tcalc = 1.2 was lower than tstat = 2.5 and hence there were no statistically
meaningful differences between both methods (accepting the null hypothesis,
see Figure b). This
result is easily visualized in Figure a, where no clear direction in variation appears. Contrarily,
the paired sample t-test indicated that there was
a statistically significant difference between MN results and the
pH measurements in the collected ISF (tcalc = 3.27 > tstat = 2.57).
Figure 7
(a) Comparison
of pH measurements with the MN patch and two different
validation procedures: subcutaneous pH measurements with the micro-pH
meter and ISF extraction. Connected lines represent values obtained
for the same rat. (b) Paired sample t-test box plot
representing statistically analyzed differences in pH measurements
with the MN patch, subcutaneous data, and ISF extraction. Gray dotted
line represents the null hypothesis.
(a) Comparison
of pH measurements with the MN patch and two different
validation procedures: subcutaneous pH measurements with the micro-pH
meter and ISF extraction. Connected lines represent values obtained
for the same rat. (b) Paired sample t-test box plot
representing statistically analyzed differences in pH measurements
with the MN patch, subcutaneous data, and ISF extraction. Gray dotted
line represents the null hypothesis.A further comparison between the pH provided by MNs and the subcutaneous
measurements can be accomplished by directly comparing the differences
in values (see Table S3). In all the cases,
this difference is lower than 0.3 pH units, except for rat number
4 and number 7 (ca. 1 pH unit). Despite the results being within the
bounds of experimental variance, ISF evaporation is hard to be controlled
in subcutaneous measurements when the skin is exposed to the air,
and this may result in significant variations compared to the MN data.
Transdermal measurements are indeed advantageous compared with the
two validation techniques, which is required for sample and/or specimen
manipulation. On the other hand, as observed in Figure a, ISF extraction tends to provide pH values
higher than the MN patch. Thus, the two validation methods are not
providing the same information and evidently, at least one of them
presents some systematic error. Moreover, comparison of the two validation
methods results in tcalc = 4.4 > tstat = 2.6, meaning that the results are statistically
different (see Figure b). A plausible explanation of this result is that ISF extraction
takes place for a long time period (30 min) in which blood flow has
been stopped because the rat is no longer alive, therefore affecting
normal diffusion processes between blood and ISF.The same observation
was reported by Mani et al. for anaesthetized
mice: pH values measured with MN sensors in CSF and bladder were slightly
lower than those measured using commercial pH electrodes after fluid
extraction.[16] Authors attributed the observed
differences to a change in the sample pH caused by the removal of
biological fluids from their buffered (physiological) conditions.
Also, subcutaneous measurements are directly accomplished in the rat,
and, therefore, less marked alterations in the sample are expected.
As a result, subcutaneous measurements seem to be more reliable as
the validation procedure. Nevertheless, it should be noted that this
method requires surgery, and therefore, future efforts should be devoted
to the development of noninvasive and reliable validation protocols
toward an adequate validation of pure in vivo measurements
in animals and humans.
Conclusions
A fully validated potentiometric
pH MN sensor for transdermal pH
measurements in ISF has been presented. This is the first time that
such a sensor is demonstrated for on-body measurements in rats after
a deep in vitro and ex vivo characterization
at the laboratory scale. Stainless-steel solid MNs were modified following
a layer-by-layer approach to provide robust and reliable pH measurements.
Analytical characterization of the MN patch under in vitro conditions demonstrated suitable Nernstian response, excellent repeatability
and reproducibility, fast response time, adequate drift, and an LRR
wide enough to cover both physiological and abnormal pH levels in
ISF. Ex vivo assays using chicken, porcine, and rat
skin revealed that the new MN patch is resilient to skin insertion
and provides high precision and accuracy, as inferred from an ex vivo validation using preconditioned pieces of animal
skin. The MN patch was successfully implemented into on-body assays
in rats, measuring interstitial pH in seven specimens. Two different
validation protocols were evaluated, concluding that subcutaneous
pH measurements are more reliable than the analysis of extracted ISF
samples. Overall, this work provides a clear guide on how to properly
characterize and validate the analytical performance of newly developed
MN sensors for any type of analyte (ions and biomolecules), starting
from in vitro conditions and reaching on-body measurements. This is indeed a key aspect in making meaningful advances
in the development and application of wearable eHealth devices based
on minimally invasive MNs.
Authors: Bao Quoc Tran; Philip R Miller; Robert M Taylor; Gabrielle Boyd; Phillip M Mach; C Nicole Rosenzweig; Justin T Baca; Ronen Polsky; Trevor Glaros Journal: J Proteome Res Date: 2017-12-15 Impact factor: 4.466
Authors: Marc Parrilla; María Cuartero; Sara Padrell Sánchez; Mina Rajabi; Niclas Roxhed; Frank Niklaus; Gastón A Crespo Journal: Anal Chem Date: 2019-01-04 Impact factor: 6.986
Authors: Mohaddeseh Aref; Elias Ranjbari; Juan José García-Guzmán; Keke Hu; Alicia Lork; Gaston A Crespo; Andrew G Ewing; Maria Cuartero Journal: Anal Chem Date: 2021-11-16 Impact factor: 6.986