Qianyu Wang1, Agueda Molinero-Fernandez1, Ana Casanova1, Joep Titulaer2, Jonatan C Campillo-Brocal3, Åsa Konradsson-Geuken2, Gaston A Crespo1, Maria Cuartero1. 1. Department of Chemistry, School of Engineering Sciences in Chemistry, Biotechnology and Health, KTH Royal Institute of Technology, Teknikringen 30, SE-100 44 Stockholm, Sweden. 2. Section of Neuropharmacology and Addiction Research, Department of Pharmaceutical Biosciences, Uppsala University, SE-751 05 Uppsala, Sweden. 3. Department of Genetics and Microbiology, University of Murcia, Campus Universitario de Espinardo, 30100 Murcia, Spain.
Abstract
Glycine (GLY) is gaining importance in medical diagnoses due to its relationship with multiple physiological functions. Today, GLY is exclusively analyzed using instrumentation centralized in clinical labs, and a tangible point-of-care tool that gathers real-time data from the patient for effective and fast evaluations is lacking. Relevant clinical advances are expected as soon as the rapid provision of both punctual and continuous measurements is possible. In that context, this work presents a microneedle (MN)-based biosensor for intradermal GLY detection in interstitial fluid (ISF). The MN tip is externally tailored to detect GLY levels through the hydrogen peroxide formed in its reaction with a quinoprotein-based GLY oxidase enzyme. The analytical performance of the MN biosensor indicates a fast response time (<7 s); acceptable reversibility, reproducibility, and stability; as well as a wide linear range of response (25-600 μM) that covers the physiological levels of GLY in ISF. The MN biosensor conveniently exhibits high selectivity for GLY over other compounds commonly found in ISF, and the response is not influenced by temperature, pH, or skin insertions. Validated intradermal measurements of GLY were obtained at the in vitro (with pieces of rat skin), ex vivo (on-body tests of euthanized rats) and in vivo (on-body tests of anesthetized rats) levels, demonstrating its ability to produce accurate physiological data. The developed GLY MN biosensor is skin-wearable and provides reliable, real-time intradermal GLY measurements in ISF by means of a minimally invasive approach.
Glycine (GLY) is gaining importance in medical diagnoses due to its relationship with multiple physiological functions. Today, GLY is exclusively analyzed using instrumentation centralized in clinical labs, and a tangible point-of-care tool that gathers real-time data from the patient for effective and fast evaluations is lacking. Relevant clinical advances are expected as soon as the rapid provision of both punctual and continuous measurements is possible. In that context, this work presents a microneedle (MN)-based biosensor for intradermal GLY detection in interstitial fluid (ISF). The MN tip is externally tailored to detect GLY levels through the hydrogen peroxide formed in its reaction with a quinoprotein-based GLY oxidase enzyme. The analytical performance of the MN biosensor indicates a fast response time (<7 s); acceptable reversibility, reproducibility, and stability; as well as a wide linear range of response (25-600 μM) that covers the physiological levels of GLY in ISF. The MN biosensor conveniently exhibits high selectivity for GLY over other compounds commonly found in ISF, and the response is not influenced by temperature, pH, or skin insertions. Validated intradermal measurements of GLY were obtained at the in vitro (with pieces of rat skin), ex vivo (on-body tests of euthanized rats) and in vivo (on-body tests of anesthetized rats) levels, demonstrating its ability to produce accurate physiological data. The developed GLY MN biosensor is skin-wearable and provides reliable, real-time intradermal GLY measurements in ISF by means of a minimally invasive approach.
Point-of-care (POC) testing
has become an important concept in healthcare in terms of enabling
diagnostics and quick feedback of many clinical cases, featuring ease
of function near the patient or even self-examination. POC testing
not only relieves congestion in emergency rooms and delays in data
provision but also paves the way toward more sustainable medical management.[1] Accordingly, many POC devices are already available
to provide information about an individual’s health. Most of
the parameters currently being monitored are physical in nature, such
as temperature, breathing frequency, blood pressure, pulse, and heart
rate.[2] Despite the valuable information
that (bio)chemical parameters can provide, they are more difficult
to be conceived at the POC level and hence are still underrepresented
in the field.One of the most successful examples of a POC device
is the glucometer,
which reflects the excellence that is sought in POC tools: almost
no restrictions on patient use anywhere, while offering total reliability
at minimal cost. Yet, the glucometer’s configuration and benefits
are continuously evolving, with research and innovation heading primarily
in four directions: (i) reducing calibration requirements, (ii) data
collection for personalized telemedicine assistance, (iii) automating
measurements for continuous monitoring, and (iv) avoiding blood collection
(finger prick).[3] Within this context, microneedle
(MN)-based (bio)sensor technology is rapidly evolving into wearable
POC tools for both punctual and continuous intradermal measurements.
The core hypothesis for the clinical use of MN sensors is that the
dermal interstitial fluid (ISF) provides the exact same health-related
information as blood. Wearable MN intradermal sensors are in a privileged
position of being able to combine valuable clinical information with
painless ISF access and thus enabling real-time digitalization, which
contrasts sharply with traditional blood tests.While initial
efforts were focused on ISF extraction with hollow
structures connected to external sensors and chips (hence, the analytical
measurements occurred off-skin), recent research has considered the
implementation of electrochemical (bio)sensing concepts directly in
the MN tip (external or internal modification).[3] Seemingly, ISF extraction is not a straightforward task,
in addition to the fact that the collected volume is very low (ca.
1 μL) to provide reliable measurements.[4] Tremendous advances have been achieved in both the design and targets
of MN (bio)sensors, with glucose sensors at the forefront of the list.[5,6] Yet, there are several challenges to be addressed, including new
analytes, proper calibration, reliable validation of on-body data,
and the managing of ethical permits, which results in only few investigations
reaching true in vivo measurements.[7−9]Amino acids (AAs)
are potential targets of MN biosensors, as the
clinical interest in their detection has increased in recent years
owing to new discoveris.[10] Indeed, fruitful
insights are expected along with the development of analytical tools
(e.g., MN biosensors) able to provide discrete and continuous body
measurements. Glycine (GLY), the smallest AA, is of particular relevance
because it accounts for ca. 12% of the total AAs and 20% of AA nitrogen
in body proteins, resulting in its significant association with many
diseases and physiological states.[11] While
the potential and benefits of GLY are yet under study, some investigations
have been dedicated to the correlations between GLY levels and certain
body states, e.g., low blood concentrations of GLY are correlated
with obesity, diabetes, insomnia, gout, and schizophrenia,[12] whereas high concentrations of GLY are associated
with nonketotic hyperglycinemia (also known as glycine encephalopathy)
and cancer progression.[13,14]Today, the determination
of GLY (mainly in blood) is carried out
through laboratory-based techniques, including liquid chromatography,
mass spectroscopy, and fluorescence assay kits.[15,16] While these approaches provide reliable results, the main drawbacks
are delayed results and rather significant startup overhead, maintenance,
and personnel costs, together with the difficulty of obtaining results
in the ISF. Hence, the clinical field would benefit from a wearable
MN device that provides on-body, real-time (continuous and/or discrete)
GLY measurements, something that, to the best of our knowledge, is
lacking in the market and scientific literature.Herein, we
present a solid MN biosensor for the intradermal analysis
of GLY in ISF by means of an epidermal patch. The MN tip is externally
tailored via a multilayering approach to provide indirect GLY analysis
through the detection of hydrogen peroxide (H2O2) formed in its enzymatic reaction with the glycine oxidize (GLY-Ox)
enzyme (a quinoprotein).[17] Validated intradermal
measurements of GLY in ISF are accomplished through the following
scenarios: (i) in vitro measurements involving rat skin pieces conditioned
in solutions with known GLY concentrations, (ii) ex vivo on-body measurements
involving euthanized rats, and (iii) in vivo on-body measurements
involving anesthetized rats. Notably, the developed experimental path
to investigate and validate the GLY MN biosensor may be conveniently
followed by other new MN (bio)sensors toward reliable analytical characterization,
which is a common challenge.[3,9] Then, the presented
GLY MN biosensor has enormous potential in terms of providing new
clinical advances in the prevention and monitoring of illnesses in
the context of contemporary medication therapy. It will also accelerate
medical research involving GLY, such as restricted cancer treatment,
requiring either discrete or dynamic assessment of GLY in real-time.
Experimental Section
GLY measurements in ISF were accomplished
by means of a silicon
rubber substrate (Ecoflex 00-50 platinum cure, USA) with two modified
MNs (the GLY MN biosensor patch, Figure a,b): a working electrode (WE) and a counter/reference
electrode (C/RE). To address the issue of skin deformation and viscoelasticity
during MN insertion for the in vitro, ex vivo, and in vivo tests,
we selected a soft substrate with exceptional flexibility for on-body
testing. Commercial stainless-steel alloy MNs were chosen to develop
the GLY biosensor due to their simplicity, high mechanical strengths,
and low cost.[3] The total length of the
modified MNs (1000 μm) enabled them to penetrate the stratum
corneum layer of the skin, thus reaching the dermis ISF and avoiding
the blood vessels and nerve endings located at around a depth of 2000
μm (Figure b).[18] The substrate was 13 mm in diameter and 0.5
mm thick. The unmodified stainless-steel solid MNs (Dermaroller, Sweden)
were 1.5 mm in length and 150 μM in diameter.
Figure 1
(a) Illustration of the
MN patch. (b) Image of the MN patch inserted
into the skin. (c) Different layers of the WE and the C/RE. (d) Working
GLY biosensor mechanism. GLY = glycine, SST = solid stainless steel,
C = carbon, PB=Prussian Blue, RM = reference membrane, PU = polyurethane,
PW = Prussian White.
(a) Illustration of the
MN patch. (b) Image of the MN patch inserted
into the skin. (c) Different layers of the WE and the C/RE. (d) Working
GLY biosensor mechanism. GLY = glycine, SST = solid stainless steel,
C = carbon, PB=Prussian Blue, RM = reference membrane, PU = polyurethane,
PW = Prussian White.First, the MNs were dip-coated with the corresponding
ink (C for
the WE, and Ag/AgCl for the C/RE), then cured in an oven (120 °C,
10 min). The MNs were subsequently inserted in the substrate and glued
with Loctite Super Glue (Henkel Norden AB) in the upper part (last
0.5 mm in length). From this upper part is where the required electrical
connections are further created to the reader. After drying the glue
for at least 4 h at room temperature, functionalization was completed
in the bottom parts of the MNs. The resulting multilayered MNs are
illustrated in Figure c.For the WE, a Prussian Blue (PB) layer was electrodeposited
on
the C-MN surface by 10 cyclic voltammetry cycles from −0.5
to 0.6 V at 50 mV s–1 in a solution comprising 2.5
mM FeCl3, 2.5 mM K3[Fe(CN)6], 100
mM KCl, and 100 mM HCl. This deposition was followed by a 1-h curing
at 100 °C in the oven. Next, a GLY-Ox/chitosan (GLY-Ox/CHI) mixture
(1:2 v/v ratio of GLY-Ox:CHI) was prepared by diluting the parental
GLY-Ox solution (0.97 ± 0.01 U GLY-Ox mL–1)
three times in phosphate buffer and combining it with CHI (1%, aq.)
in 7.15% acetic acid, then drop cast onto the PB-C-MN. This enzyme
layer was allowed to dry in the fume hood for at least 1 h. Afterward,
2 μL of 1% wt. Nafion solution (0.2 mL of 5% wt. Nafion in 0.8
mL of water) was drop-casted and allowed to dry for 30 min. Finally,
the modified MN was stored in a phosphate buffered saline solution
(PBS) at 4 °C, also for between-day measurements.The C/RE
was prepared as reported elsewhere.[19] Briefly,
3 μL of poly (vinyl butyral) (PVB)/NaCl
/MeOH solution (78 mg of PVB, 50 mg of NaCl in 1 mL of methanol) was
drop cast on top of the Ag/AgCl film and conditioned in 3 M KCl overnight.
Afterward, an external polyurethane (PU) membrane was applied by drop-casting
2 μL of PU solution (20 mg PU in 1 mL THF). All the layers were
cured in the fume hood until they were completely dry. The C/RE MN
was stored in 3 M KCl.The bare solid stainless-steel MNs as
well as finalized WE and
C/RE MNs were characterized using scanning electron microscopy (SEM)
images (Figure S1). The dimensions of the
initial MNs in terms of length plus widths at the base and tip were
confirmed (1.5 mm, 150 and 15 μm, respectively). In the case
of the WE, the widths at the base and tip of the MN were found to
have increased to 200–220 μm and 24–35 μm,
respectively, with the addition of the C, PB, GLY-Ox, and Nafion layers.
For the C/RE, the widths at the base and the tip increased to 180–206
μm and ca. 30 μm, respectively, with the implementation
of the Ag/AgCl, PVB, and PU layers. The homogeneities of the layers,
together with the absence of any physical deterioration in the WE
and C/RE MNs after skin insertion, were confirmed. When accompanied
by a proper insertion angle and force during the penetration process,
the dimensions of our modified MNs should permit painless and minimally
invasive skin puncture.[20]
Results and Discussion
The biosensor patch is composed
of two MN-based electrodes, namely,
the WE and the C/RE, which are necessary for the amperometry readout
(two-electrode system). Figure d illustrates the working mechanism for GLY detection. The
sample’s GLY is partitioned in the Nafion membrane and diffuses
until reaching the GLY-Ox enzyme, where it is then converted into
glyoxylate, ammonium, and H2O2 via a stoichiometric
reaction. Meanwhile, for the PB, the original film is first reduced
to Prussian White (PW) by applying a constant potential of −50
mV to the electrode, i.e., Fe3+/Fe2+ moieties,
then the generated H2O2 is spontaneously reduced
while the PW is oxidized back to PB. The PB is reduced again to PW
via the applied potential, which generates a Faradaic current near
the electrode surface in dynamic equilibrium. Any change in the GLY
concentration in the sample will change the current’s equilibrium
value, which is proportional to the related concentration change over
a certain range, thus allowing the quantitative determination of the
GLY concentration in the sample. The overall concept was recently
reported by our group, providing the first demonstration of an electrochemical
biosensing approach to measuring the GLY concentration in urine, sweat,
and serum, thanks to the specificity of the GLY-Ox, which was successfully
encapsulated in the sensor core.[17] Following,
we translate the GLY biosensing concept for the unprecedented transdermal
analysis of GLY in ISF.
Tailoring of the Components of the MN Biosensor for GLY Detection
Figure displays
the amperometric dynamic response of the optimized MN biosensor at
increasing GLY concentrations in a PBS background. The response time
(t90) was <7 s across the entire concentration
range. The linear range of response (LRR) was from 25 to 600 μM,
with a sensitivity (slope) of −0.0880 ± 0.001 nA μM–1 (1.5% of variation coefficient within three consecutive
calibrations of the same MN biosensor) with an intercept of −51.2
± 1.1 nA (3.6%). The amperometric response was rather reproducible,
even between different MN biosensors (n = 7), producing
variations of <5.8% for the slope and <10.4% for the intercept
(see Figure S2).
Figure 2
Dynamic amperometric
response to increasing H2O2 concentrations.
Inset: average calibration graph observed
for three consecutive responses.
Dynamic amperometric
response to increasing H2O2 concentrations.
Inset: average calibration graph observed
for three consecutive responses.After reaching a GLY concentration of 600 μM
in solution,
the calibration graph begun to deviate from its linearity. There are
two possible reasons for this behavior: (i) the enzyme reaches its
saturation state, i.e., the limiting factor of the enzymatic reaction
rate is the concentration of the enzyme; thus, adding a greater substrate
(GLY) concentration does not affect the overall reaction or (ii) the
PB layer does not respond to higher H2O2 concentrations
in accordance with its upper limit of detection (upper LOD). The first
explanation seems to be the more plausible of the two because the
PB-MN electrode presents a wide range of responses toward H2O2 (see Figure S3) via the
PW/PB interconversion. Using a signal-to-noise ratio of 3 in the calibration
graph, the LOD was calculated to be 7.9 μM GLY concentration.The LRR of the MN biosensor included GLY levels within healthy
conditions (400–600 μM)[21] and
most of the unhealthy levels,[3,21] although appropriate
clinical data are difficult to locate, likely due to the absence of
appropriate analytical tools for ISF measurements. Regarding the expected
daily variations, and assuming that blood and ISF GLY present close
profiles, fluctuations of more than ±100 μM have been reported
in volunteers under different diet conditions.[22] On the other hand, the administration of a dietary supplementation
of GLY has demonstrated to be effective in preventing/treating many
disorders, as well as enhancing the sleep quality and neurological
operations.[11] Accordingly, the employment
of the developed GLY MN biosensor could be used for achieving personalized
therapies via GLY profile monitoring. Both, punctual and continuous
GLY measurements are thus valuable toward the replacement of recurrent
lab-based daily measurements, additionally pursuing unprecedented
clinical data.Although the achieved LRR seems to be sufficient
to assess the
just inspected clinical examples, in any specific case where the LRR
needs to be expanded toward higher concentrations (such as for cancer-related
applications),[23,24] an additional external polymeric
layer could be added to the MN, or the thickness of the Nafion layer
could be increased. Using one of these strategies, and acknowledging
that a decrease in sensitivity will occur, the number of substrate
molecules reaching the enzyme will be restricted, as demonstrated
by our group for lactate.[17] For very low
GLY concentrations (<25 μM), the MN biosensor can act as
an alarm rather than for quantification.To obtain the described
analytical characteristics, the MN biosensor
structure was investigated as follows. First, we evaluated the implementation
of a PB layer on the stainless-steel MN to detect H2O2. We attempted to deposit the PB directly onto the bare MN;
however, the formed film was not mechanically stable and was found
to be removed during the characterization experiments. This occurred
regardless of whether the PB film was electrodeposited or created
via drop-casting. We opted for the initial addition of a C layer to
the bare MN to facilitate the stronger attachment of the PB film due
to the fact that a passive film on the stainless-steel surface (essentially
formed by the oxide/hydroxide of chromium and iron) may prevent the
direct bonding of the PB.[25]We explored
the addition of a PB layer to C-coated MNs via either
a chemical reaction or electrodeposition. The chemical reaction was
carried out by drop-casting 8 μL of PB precursor solution (0.1
M of FeCl3, K3[Fe(CN)6], KCl, and
HCl) onto the MN tip, then allowing 20 min for a reaction at room
temperature under dark conditions. Next, the solution was removed,
and the MN tip was cleaned in 0.1 M HCl followed by curing for 1 h
at 100 °C in an oven. The electrodeposition was performed in
a solution containing 2.5 mM FeCl3, 2.5 mM K3[Fe(CN)6], 100 mM KCl, and 100 mM HCl. Cyclic voltammetry
was run from −0.5 to 0.6 V at 50 mV s–1 and
with an increasing number of scans (5, 10, and 30). Figure a depicts the cyclic voltammograms
of the PB-MNs in PBS solution. As observed, all the signals presented
an oxidation peak at ca. 150 mV and a reduction peak at ca. 0 mV.
Previous observations reported in the literature related these peaks
to the PW oxidation to PB and PB reduction to PW, respectively.[26]
Figure 3
(a) Cyclic voltammograms observed in PBS with differently
prepared
PB-C-MNs. (b) Amperometric responses observed with Nafion-Enzyme-PB-C-MNs
prepared with increasing volumes of the enzyme. Inset: corresponding
calibration graphs.
(a) Cyclic voltammograms observed in PBS with differently
prepared
PB-C-MNs. (b) Amperometric responses observed with Nafion-Enzyme-PB-C-MNs
prepared with increasing volumes of the enzyme. Inset: corresponding
calibration graphs.The redox behavior of the PB-MN manifested in higher
peak currents
when the PB was prepared via the electrodeposition method (no matter
what the number of scans) as opposed to the drop-casting chemical
reaction approach (labeled as chemical in Figure a). Increasing the volume of the precursor
solution used for the drop-casting resulted in much more mechanically
unstable films; hence, we continued with the electrodeposition method
rather than the drop-casting one. When the electrodeposition approach
was applied with five cyclic voltammetric scans (from −0.5
to 0.6 V), the peak current and the reproducibility between subsequent
cycles were slightly lower than with the 10- or 30-scan regimes. This
behavior translated into a higher sensitivity for those MNs prepared
with electrodeposited PB with 10 and 30 scans when the PB-MNs were
tested regarding increasing H2O2 concentrations
in amperometric mode (Figure S4). Considering
the results in Figure a, an applied potential of −50 mV was selected for the reduction
of PB to PW, thus generating the mechanism proposed in Figure d. No significant differences
were found between the PB-MNs prepared with 10 and 30 scans; thus,
the preparation with 10 scans was selected for further studies.We then studied the effect of the amount of GLY-Ox deposited on
the PB-based MN. Increasing volumes of the GLY-Ox/CHI mixture (2,
4, 6 and 8 μL) were deposited via drop-casting on top of the
PB-MN. Figure b shows
the amperometric responses to increasing GLY concentrations in the
PBS background. As observed, a higher amperometric response, and therefore
a higher slope, was displayed when the deposited volume of the enzyme
solution was increased from 2 to 4 μL (sensitivities of −0.0449
nA μM–1 and −0.122 nA μM–1). This increase was not as drastic for higher volumes
(slopes of −0.153 nA μM–1 and −0.130
nA μM–1 for 6 and 8 μL, respectively).
In essence, the saturation of the enzyme is reached from a certain
deposited volume/amount, because the available active sites will be
the same. We decided to use a volume of 6 μL because it presented
the maximum sensitivity and was enough to comfortably perform the
drop-casting deposition in the modified MN.Regarding the final
external layer of Nafion, we confirmed that
this was necessary to avoid any interference from ascorbic acid while
preserving the response time and calibration graph of the sensor.
This compound is of special concern because it can be directly oxidized
in the PW/PB lattice, resulting in a less negative value for the current
level provided by any GLY concentration.[27] It was found that the addition of an external Nafion layer produced
a slightly less noisy response at increasing GLY concentrations than
a MN biosensor without that layer. Thus, the response time regarding
increasing GLY concentrations was found to increase with the thickness
of this layer (8 s versus 30 s), i.e., increasing the volume deposited
on the MN, namely, 4 instead of 2 μL (Figure S5a). On the other hand, the ascorbic acid response was totally
suppressed in the MN prepared with a volume of 2 μL, whereas
without any Nafion layer, the MN biosensor responded to ascorbic acid
(Figure S5b).
Investigation of the Main Analytical Characteristics of the
MN Biosensor for GLY Detection
The reversibility of the MN
biosensor was demonstrated by measuring solutions with low and high
GLY concentrations within the LRR in the following order: 50, 500,
50, 500, 50, and 500 μM (Figure S6). Almost negligible variations were found in the steady state currents
(<2%), resulting in an averaged calibration graph with a sensitivity
of −0.107 ± 0.001 nA μM–1 and
an intercept of −76.7 ± 2.9 nA. The medium-term drifts
observed in GLY concentrations of 50 and 500 μM were acceptably
low (0.06 and 0.30 nA h–1, Figure S7), permitting the detection of these GLY concentrations with
errors of <1% over 1 h. In addition, long-term drift experiments
revealed variations of 0.12 and 1.28 nA h–1 for
a 12-h period, and 0.23 and 1.38 for a 24-h period, for 50 and 500
μM (Figure S7). The lifetime of the
MN biosensor was tested with daily calibrations: a decrease of 4%
and ca. 40% of the sensitivity was found in the days 4th and 10th,
while random changes appeared for the intercept. In view of these
results, the MN biosensor is suitable to follow random increases and
decreases in GLY concentration, with a rather reproducible response
that is acceptably stable (and thus precise) over 12 h. For a longer
continuous monitoring of GLY in ISF, it is advisable to either re-condition
and re-calibrate the sensor or substitute it by a new patch. The MNs
can be used after ca. 5 days of being prepared.Glucose, lactate,
pyruvate, urea, and ascorbic acid were investigated as potential interferences.
The effects of concentrations higher than those expected in the ISF
on the amperometric response were examined (Figure S8). It was confirmed that the biosensor qualifies as interference-free
in the detection of GLY because none of the tested compounds influenced
the GLY response. Notably, major ions, metabolites, and AAs commonly
present in biological fluids, together with GLY derivatives, are not
expected to influence the amperometric response, as previously demonstrated
with a similarly structured biosensor and traditional specificity
studies on the GLY-Ox enzyme.[17,28]Calibration graphs
were obtained at different pH in the background
solution, while keeping the temperature (T) constant
and vice versa, to study the influences of the pH and T on the response of the MN biosensor (Figure S9). These two factors may significantly alter the activity
of the GLY-Ox enzyme. Under healthy conditions, the ISF pH and T are known to be independent of ambient fluctuations and
subjects. However, the pH could slightly decrease and the T decrease/increase in patients with certain diseases.[29,30] Accordingly, we selected a pH range from 6.5 to 7.5, and a T range from 25 to 40 °C in this study. It was observed
that the dynamic responses, and consequently the calibration graphs,
were almost identical regardless of pH and T values
(see Table S1). The main differences appeared
at the highest GLY concentration assayed (500 μM), although
the most marked variations occurred because two different MN biosensors
were used for the pH and T studies (2.9% of variation
in the slope and ca. 30% in the intercept across all results). Hence,
the developed MN biosensor can be calibrated at room temperature and
at a pH of 7.5 for any posterior skin test.Figure a depicts
the dynamic responses at increasing GLY concentrations accomplished
in either PBS (pH = 7.4) or artificial ISF (AISF, pH = 7.4) using
the same MN biosensor. As observed, similar responses were obtained;
however, there was a slight shift in the provided current (i.e., off-set).
Thus, the calibration parameters were very similar, except for the
intercept, representing excellent repeatability (three subsequent
calibrations) independent of the background medium used for the calibration:
LRRs for GLY concentrations of 25 to 600 μM; sensitivities (slopes)
of −0.0880 ± 0.001 nA μM–1 (1.5%
variation) and −0.0780 ± 0.004 nA μM–1 (5.3% variation) for PBS and AISF, respectively; and intercepts
of −51.2 ± 1.9 nA (3.6% variation) and −36.0 ±
4.1 nA (11.4% variation) for PBS and AISF, respectively. Due to the
observed changes in the intercepts of the linear graphs, it was most
convenient to perform the calibration in the AISF medium, as this
best mimics the real ISF where on-body measurements will be performed
with the MN biosensor, thus minimizing any possible matrix effect.
Figure 4
Dynamic
amperometric responses and the corresponding calibration
graphs for the MN biosensor at increasing GLY concentrations under
different conditions: (a) PBS and AISF as the background electrolytes
(the calibration graphs are the average of three sequential calibrations);
(b) using commercial RE and CE, commercial C/RE, RE MN, and commercial
CE and C/RE MN; (c) before and after insertions into rat skin.
Dynamic
amperometric responses and the corresponding calibration
graphs for the MN biosensor at increasing GLY concentrations under
different conditions: (a) PBS and AISF as the background electrolytes
(the calibration graphs are the average of three sequential calibrations);
(b) using commercial RE and CE, commercial C/RE, RE MN, and commercial
CE and C/RE MN; (c) before and after insertions into rat skin.Next, we investigated the possibility of using
the MN GLY biosensor
in conjunction with a C/RE MN (based on an Ag/AgCl element) to avoid
implementing a three-electrode system in the final patch design, thus
reducing the complexity and number of MNs necessary. Accordingly,
the C/RE MN was expected to mediate the current flow through the WE
while maintaining a constant potential.[31] Notably, the current passing through the C/RE was relatively low
(always lower than 0.1 μA); therefore, the risk of current induced
changes in the Ag/AgCl-based MN leading to a change in the provided
potential is expected to be negligible. Figure b displays the dynamic responses of the MN
biosensor at increasing GLY concentrations in the AISF, together with
the corresponding calibration graphs, when using different RE and
CE systems: (i) commercial Ag/AgCl RE, and commercial Pt CE (COMM-RE);
(ii) commercial Ag/AgCl RE acting as a C/RE; (iii) Ag/AgCl MN as the
RE and commercial Pt as the CE; and (iv) the Ag/AgCl MN as the C/RE.
It was evident that the LRR and slope were maintained independent
of the CE and RE used (Table S2): the variations
in the slope were always within the range displayed in the repeatability
and reproducibility studies (ca. 13%). On the other hand, a higher
variation was found for the intercept because of the different RE
and CE pairs and different MN GLY biosensors used in the systems (23.3%).
As a result, a two-electrode system composed of the MN GLY biosensor
and the C/RE MN can be used without affecting the results. Therefore,
this was the configuration used.Finally, it was confirmed that
changes in the amperometric response
of the two-electrode MN system were not induced by skin insertion.
Moreover, the external modification made to the MNs to introduce the
sensing elements did not detach and/or deteriorate. Calibration graphs
in the AISF background were produced before and after one and five
insertions in a piece of rat skin (Figure c), which was fixed in a holder to mimic
on-body conditions (see Figure a for the in vitro setup). The results indicated that the
calibration graph remained nearly invariable, with an LRR from 25
to 500 μM, an average slope of −0.0875 ± 0.002 nA
μM–1 (2.7% variation), and an intercept of
−31.0 ± 1.0 nA (3.1% variation). Effectively, the low
variations indicated that the MN GLY biosensor can be initially calibrated
and then used for successive skin insertions for the intradermal detection
of GLY in the ISF without loss of integrity or biofouling. Moreover,
SEM images of the MN biosensor and the C/RE MN revealed no significant
changes or particle adherence on the MN’s surface (see Figure S1). Nevertheless, postcalibration graphs
should be generated after skin insertions (in in vitro, ex vivo, and
in vivo measurements) to confirm that the MN biosensor’s performance
has not deteriorated.
Figure 5
(a) Image of the in vitro setup, where a piece of rat
skin is fixed
in a holder to be analyzed with the MN biosensor. (b) Image of an
on-body test using the MN GLY biosensor on a euthanized rat. (c) Image
of the electronic board used for amperometric measurements, plus the
casing used to introduce it and run the wireless data acquisition
(KERIC). (d) Image of an in vivo test using the MN GLY biosensor in
an anesthetized rat (UUBF).
(a) Image of the in vitro setup, where a piece of rat
skin is fixed
in a holder to be analyzed with the MN biosensor. (b) Image of an
on-body test using the MN GLY biosensor on a euthanized rat. (c) Image
of the electronic board used for amperometric measurements, plus the
casing used to introduce it and run the wireless data acquisition
(KERIC). (d) Image of an in vivo test using the MN GLY biosensor in
an anesthetized rat (UUBF).
In Vitro Tests in Pieces of Rat Skin
Three pieces of
rat skin (approx. 2 × 2 cm each) were conditioned for 24 h at
4 °C in the fridge in solutions containing GLY concentrations
of 200, 300, or 400 μM in the PBS. Thereafter, each piece of
skin was fixed in a 3D-printed holder (Figure a), allowed to dry at room temperature and
squeezed with paper tissue to remove the remaining moisture. A patch
containing the MN biosensor and the C/RE MN was implemented into the
skin by carefully inserting the MNs, and the dynamic amperometric
signal was recorded. For each skin, measurements were conducted through
two insertions in different parts of the skin piece.Figure exemplifies the
amperometric response for a 3-point calibration graph accomplished
previously to the in vitro experiment, together with the results of
the two insertions in the skin conditioned in 400 μM GLY. The
calibration was run in PBS background, the same as in the solution
used for the skin conditioning. As observed, the recorded current
in each insertion was fairly constant and the value was within the
calibration graph. Thus, the calibration graph was employed to calculate
the GLY concentration in each piece of skin and the results are collected
in Table as the average
of two insertions.
Figure 6
Left: Responses of the MN GLY biosensor
at increasing
GLY concentrations, which were used to obtain a calibration graph
prior to the in vitro experiments. Right: Dynamic
current registered during two consecutive insertions with the same
MN patch (GLY biosensor and the C/RE MN) in a piece of rat skin conditioned
in a 400 μM GLY solution.
Table 1
GLY Levels Detected in Different Samples
and Rats by Means of Both the Developed MN GLY Biosensor and the Fluorescence
Kit
sample/specimen
[GLY] average ± SD (μM)
% diff.
MNd
fluorescence
rat skin 1 (200 μM)a
72.5 ± 4.3
88.6 ± 0.6
18
rat skin 2 (300 μM)a
142.5 ±
3.5
111.6 ±
1.4
27
rat skin 3 (400 μM)a
216.7 ± 13.8
196.3 ± 11.9
10
rat
1 (euthanized)b
189.1 ± 7.4
193.6e
2
rat 2 (euthanized)b
103.0 ± 4.7
111.1e
7
rat 3 (euthanized)b
205.2 ± 25.7
202.1 ± 17.5
2
rat
4 (anesthetized)c
290.1 ± 14.4
232.6 ± 7.2
24
In vitro experiments with rat skins
conditioned in different GLY concentrations.
On-body tests with euthanized rats
at KERIC.
In vivo tests
with anesthetized
rats at UUBF.
The measurements
averaged 4 to 6,
as some outliers were identified.
Only one measurement was possible
due to insufficient volumes of collected ISF.
Left: Responses of the MN GLY biosensor
at increasing
GLY concentrations, which were used to obtain a calibration graph
prior to the in vitro experiments. Right: Dynamic
current registered during two consecutive insertions with the same
MN patch (GLY biosensor and the C/RE MN) in a piece of rat skin conditioned
in a 400 μM GLY solution.In vitro experiments with rat skins
conditioned in different GLY concentrations.On-body tests with euthanized rats
at KERIC.In vivo tests
with anesthetized
rats at UUBF.The measurements
averaged 4 to 6,
as some outliers were identified.Only one measurement was possible
due to insufficient volumes of collected ISF.In addition, the ISF inside the skin was collected
by a homemade
peristaltic pump with a hollow MN hub (more details in the Supporting Information). In all cases, a volume
of 1–3 μL was obtained. GLY contents in the collected
ISF samples were analyzed with the commercial fluorescence kit for
GLY detection in biological samples, and the results are provided
in Table , along with
the estimated results obtained with the MNs. Similar values were provided
with both techniques, indicating the accuracy of the results obtained
with the MN GLY biosensor. Notably, in all cases, the detected GLY
concentration was lower than that in the corresponding conditioning
solution. This behavior was previously noted in skin pieces conditioned
in different ions (e.g., potassium) and biomolecules (such as urea)
and attributed to specific partition and diffusion conditions.[19,32]
Ex Vivo On-Body Measurements in Euthanized Rats (KERIC Facilities)
The potential of the MN patch (MN biosensor and C/RE MN) to provide
intradermal measurements of GLY was investigated in three euthanized
rats. Figure b displays
an image of one of the rats undergoing MN measurements on its shaved
back region. Three different MN patches were inserted twice in different
(but proximal) positions of the shaved back to collect current readings
and transform them into GLY concentrations according to the calibration
previously performed in AISF. Figure a presents all the GLY measurements for rat 1 as an
example. As observed, similar measurements were obtained for the two
insertions with each patch (variations of <5.5%) but also between
the three MN patches (variations of <10%). For each rat, the GLY
concentration was calculated as the average of four to six measurements,
depending on the identification of any outliers due to malfunctions
of the patch and/or electrical connections. The results are shown
in Table .
Figure 7
(a) Dynamic
current registered in the ex vivo on-body test of (euthanized)
rat 1 using three different MN patches, each one inserted twice in
the shaved back of the rat. (b) GLY concentration registered during
the in-vivo on-body test of (anesthetized) rat 4 using two different
MN patches, each one inserted twice in the shaved back of the rat.
Dashed lines indicate the levels of GLY found in collected ISF with
the fluorescence assay kit.
(a) Dynamic
current registered in the ex vivo on-body test of (euthanized)
rat 1 using three different MN patches, each one inserted twice in
the shaved back of the rat. (b) GLY concentration registered during
the in-vivo on-body test of (anesthetized) rat 4 using two different
MN patches, each one inserted twice in the shaved back of the rat.
Dashed lines indicate the levels of GLY found in collected ISF with
the fluorescence assay kit.In addition, ISF samples were collected from each
rat, and the
GLY content was analyzed by means of the fluorescence assay kit (Table ). Interestingly,
we found that the amount of ISF extracted (and collected) was related
to age. Seemingly, younger rats have more ISF; thus, we were able
to obtain 20–30 μL from 2-month-old rats, likely due
to both their smoother skin and higher hydration levels compared to
older rats.[33] On the contrary, we could
only obtain a few microliters from older rats, which sometimes made
it difficult to validate the on-body measurements. Inspecting the
estimated GLY levels, the MN and fluorescence methods produced very
similar results, with differences of <7% between them, pointing
out once more the accuracy of the MN measurements (Table ).
In Vivo On-Body Measurements of Anesthetized Rats (UUBF Facilities)
The developed MN patch was additionally used for intradermal measurements
of GLY in three anesthetized rats (in accordance with the Uppsala
Committee on Ethics of Animal, Dnr 5.8.18-18873/2018, DOUU-2020-025).
In this case, the MNs (GLY biosensor and C/RE) were integrated into
a miniaturized wireless potentiostatic board (Figure c). The board was placed inside a casing
for protection and to facilitate portability during the in vivo on-body
measurements (Figure c). Figure d shows
one of the rats connected to the isoflurane flow (for anesthesia,
details in the Supporting Information)
while the on-body GLY measurements with the MN patch were performed.
Two different MN patches were tested on the rats via two insertions
in different (but proximal) positions on their shaved backs. Unfortunately,
conclusive measurements were only possible for one of the rats (labeled
as rat 4), with the absence of reproducible data for the other two
due mainly to problems with the MN connections to the electronic board. Figure b displays all the
GLY measurements acquired from rat 4. While very similar data were
obtained in the two insertions of the first patch and the first insertion
of the second patch, the last measurement rather deviated from the
rest and was not considered in the final calculation of the GLY concentration.
The results are also found in Table . As observed, a higher difference between the MN readings
and fluorescence kit results was found for the ISF GLY (>20%),
compared
to the case for euthanized rats. The results chiefly demonstrated
the possibility that rather accurate intradermal measurements of GLY
can be made.
Conclusions
The very first MN biosensor for transdermal
GLY analysis in ISF
has been herein presented. The solid stainless steel MNs have been
externally tailored with a series of elements for the specific and
selective determination of GLY. The performance of the MN GLY biosensor
has been deeply investigated, and it has been found to have a fast
response time; good reversibility and reproducibility; a negligible
response to pH, T, and the main interfering agents
present in ISF; and an LRR suitable for GLY analysis in ISF. Validated
in vitro assays using pieces of rat skin, together with ex vivo and
in vivo on-body tests in rats, have successfully demonstrated the
capability of the developed MN biosensor in terms of providing accurate
intradermal GLY measurements in dermal ISF. Remarkably, to the best
of our knowledge, these are the very first on-body transdermal measurements
in ISF that are reported in rats. The developed experimental path
may serve as a practical guide for the appropriate analytical characterization
of any newly developed MN (bio)sensor. Furthermore, the results herein
constitute meaningful advances in the development and application
of wearable eHealth devices based on minimally invasive MNs. For GLY
detection, it is expected that the developed device contributes to
generate unprecedented data concerning the connection of GLY levels
in ISF with its use as a biomarker and/or in disorders’ treatments.
Authors: Mohit Jain; Roland Nilsson; Sonia Sharma; Nikhil Madhusudhan; Toshimori Kitami; Amanda L Souza; Ran Kafri; Marc W Kirschner; Clary B Clish; Vamsi K Mootha Journal: Science Date: 2012-05-25 Impact factor: 47.728
Authors: Pradnya P Samant; Megan M Niedzwiecki; Nicholas Raviele; Vilinh Tran; Juan Mena-Lapaix; Douglas I Walker; Eric I Felner; Dean P Jones; Gary W Miller; Mark R Prausnitz Journal: Sci Transl Med Date: 2020-11-25 Impact factor: 17.956
Authors: Marc Parrilla; María Cuartero; Sara Padrell Sánchez; Mina Rajabi; Niclas Roxhed; Frank Niklaus; Gastón A Crespo Journal: Anal Chem Date: 2019-01-04 Impact factor: 6.986
Authors: Farshad Tehrani; Hazhir Teymourian; Brian Wuerstle; Jonathan Kavner; Ravi Patel; Allison Furmidge; Reza Aghavali; Hamed Hosseini-Toudeshki; Christopher Brown; Fangyu Zhang; Kuldeep Mahato; Zhengxing Li; Abbas Barfidokht; Lu Yin; Paul Warren; Nickey Huang; Zina Patel; Patrick P Mercier; Joseph Wang Journal: Nat Biomed Eng Date: 2022-05-09 Impact factor: 25.671