Alexander Wiorek1, Marc Parrilla1, María Cuartero1, Gastón A Crespo1. 1. Department of Chemistry, School of Engineering Sciences in Chemistry, Biotechnology and Health, KTH Royal Institute of Technology, Teknikringen 30, SE-100 44 Stockholm, Sweden.
Abstract
We present an epidermal patch for glucose analysis in sweat incorporating for the first time pH and temperature correction according to local dynamic fluctuations in sweat during on-body tests. This sort of correction is indeed the main novelty of the paper, being crucial toward reliable measurements in every sensor based on an enzymatic element whose activity strongly depends on pH and temperature. The results herein reported for corrected glucose detection during on-body measurements are supported by a two-step validation protocol: with the biosensor operating off- and on-bodily, correlating the results with UV-vis spectrometry and/or ion chromatography. Importantly, the wearable device is a flexible skin patch that comprises a microfluidic cell designed with a sweat collection zone coupled to a fluidic channel in where the needed electrodes are placed: glucose biosensor, pH potentiometric electrode and a temperature sensor. The glucose biosensor presents a linear range of response within the expected physiological levels of glucose in sweat (10-200 μM), and the calibration parameters are dynamically adjusted to any change in pH and temperature during the sport practice by means of a new "correction approach". In addition, the sensor displays a fast response time, appropriate selectivity, and excellent reversibility. A total of 9 validated on-body tests are presented: the outcomes revealed a great potential of the wearable glucose sensor toward the provision of reliable physiological data linked to individuals during sport activity. In particular, the developed "correction approach" is expected to impact into the next generation of wearable devices that digitalize physiological activities through chemical information in a trustable manner for both sport and healthcare applications.
We present an epidermal patch for glucose analysis in sweat incorporating for the first time pH and temperature correction according to local dynamic fluctuations in sweat during on-body tests. This sort of correction is indeed the main novelty of the paper, being crucial toward reliable measurements in every sensor based on an enzymatic element whose activity strongly depends on pH and temperature. The results herein reported for corrected glucose detection during on-body measurements are supported by a two-step validation protocol: with the biosensor operating off- and on-bodily, correlating the results with UV-vis spectrometry and/or ion chromatography. Importantly, the wearable device is a flexible skin patch that comprises a microfluidic cell designed with a sweat collection zone coupled to a fluidic channel in where the needed electrodes are placed: glucose biosensor, pH potentiometric electrode and a temperature sensor. The glucose biosensor presents a linear range of response within the expected physiological levels of glucose in sweat (10-200 μM), and the calibration parameters are dynamically adjusted to any change in pH and temperature during the sport practice by means of a new "correction approach". In addition, the sensor displays a fast response time, appropriate selectivity, and excellent reversibility. A total of 9 validated on-body tests are presented: the outcomes revealed a great potential of the wearable glucose sensor toward the provision of reliable physiological data linked to individuals during sport activity. In particular, the developed "correction approach" is expected to impact into the next generation of wearable devices that digitalize physiological activities through chemical information in a trustable manner for both sport and healthcare applications.
Epidermal
wearable chemical
sensing technology allows for the continuous profiling of distinct
biomarkers (i.e., ions and molecules) in sweat during perspiration,
and this is highly significant to understand the physiology of individuals
under certain activities and ultimately to provide a personalized
training routine.[1] Noteworthy, wearable
sensor patches are not only restricted to sport-physiology, and in
fact, the science behind its development envisions to promote the
true digital transformation of the body-status in the broadest sense:[2−4] this includes clinical diagnosis supported by artificial intelligence
to shed light on prediction patterns.Most of the papers about
epidermal sensing patches focuses the
discussion and purpose of the investigation on sport-physiology, but
rarely, strict validations using a gold-reference and pilot tests
are depicted. This may likely be one of the reasons to understand
the lack of these devices in the commercializing pipeline for sweat
sensing, whereas the interest of general people and market is undoubtedly
high.[5] In this regard, we have recently
reported on a rather detailed guideline for sweat sampling based on
the regional absorbent patch method, aiming at the suppression of
potential bias in the final correlation of on-body data once the wearable
patch is cross-validated based on sweat sampling.[6]Importantly, epidermal wearable ion-sensing has been
exhaustively
covered by the Javey group with elegant descriptions of the circuitry,
stretchable materials, apps and spiral-patterned microfluidic channel,
among others.[7−9] Recently, our group has gone one step forward in
terms of simplicity in microfluidic integration, cytotoxicity, strictness
of the analytical validation, representative number of on-body tests,
and as worth mentioning, the verification of genuine and accurate
chemical information.[6,10] Apart from ions (i.e., electrolytes),
there are other relevant biomolecules that captivate the attention
of physiologist and clinician from the body-status point of view such
as glucose, lactate, and ethanol.[11,12] In particular,
while the importance of glucose sensing is well-known in terms of
diabetes monitoring and therapy, there are a number of open questions
about physiological relevance in sport, the correlation between blood/sweat
in such aerobic/anaerobic conditions, and, finally, the partition
mechanism from blood to sweat. This is the reason for new glucose
(bio)sensors arising daily in the literature.[13]The group of Kim has been working on graphene-based electrochemical
devices for glucose sensing in sweat.[14] Although, the stretchable device is rather focused on diabetes therapy
(i.e., it delivers a drug from loaded-microneedles once the level
of glucose is high [hyperglycemia]), there are some insightful aspects
related to the analytical performance of the glucose sensor. Indeed,
Kim et al. suggested the necessity of pH and temperature (T) correction for enzyme-based sensors, an important point
seldom mentioned in glucose wearable sensor assessments.[14] Perspiration is a dynamic process that involves
fluctuations in pH and T during sport practice, being also levels
and trends completely different comparing several subjects (approximately
2 units in pH and a range of ca. 20–40 °C for the T). These fluctuations are expected to influence the biosensor
response as a result of changes in the activity of the enzyme on which
the sensor is based.[15] This puts in evidence
the necessity to evaluate those effects off-site, off-body, and on-body
to achieve an accurate monitoring of glucose. In this sense, Javey
and co-workers reported on T compensation[4] while Kim et al. reported on both T and pH correction of the glucose profile for short analysis time
(less than 10 min).[14]In this rich
and exciting context about epidermal sweat sensing
progress, in particular for glucose detection, we understand that
there is a lack of a systematic study proving the importance of pH
and T correction while sweat is flowing through a
microfluidic cell for an extensive period of time (1 h), which is
representative of sport training. In contrast to the current state
of the art, we feel the necessity to ensure the quality (i.e., accuracy)
of the provided data by using the sampling method published recently
by our group.[6] Overall, we propose herein
an analytical protocol to correct the data provided by the glucose
epidermal patch based on pH and T measurements in
sweat (i.e., inside the microfluidic cell) as well as a validation
based on sweat collection. Indeed, we have developed a novel sensing
and fluidics design for glucose, pH, and T monitoring
in sweat and use it as proof-of-concept during cycling. Finally, the
validated on-body data fully demonstrates the relevance of the epidermal
patch with the glucose biosensor to monitor a continuous profile in
sweat of different subjects during physical activity in a total of
9 on-body tests at different conditions (T1–9).
Experimental
Section
Fabrication of the Sensor Array
Figure a–c illustrates the glucose, pH, and T sensors fabricated on a flexible polyester sheet (0.125
mm thick, RS components, Sweden) as the substrate. The glucose sensor
consists of a three-electrode system, each of them prepared on individual
circular patterns (2 mm of diameter and separated in 2 mm between
them) that were connected to a straight path (8 mm long) that is screen-printed
using Ag/AgCl ink (C2131007D3, Gwent group, U.K.) and serving for
further electrical connections. The circular patterns for the working
and counter electrodes (WE1 and CE) were prepared with
carbon ink (C2030519P4, Gwent group, U.K.), while the reference electrode
(RE1) consisted of just the straight track made of Ag/AgCl.
Subsequently, the WE1 surface was modified with Prussian
blue mediator (PB), an enzymatic layer (i.e., chitosan and glucose
oxidase, CHI + GOx), and external polymeric layer (i.e., Nafion).
The pH sensor was fabricated using the same conductive pattern as
for the glucose biosensor, this time being composed of a working electrode
(WE2) and a reference electrode (RE2) for the
potentiometric readout. Subsequently, the WE2 and RE2 were modified first with a layer of multiwalled carbon nanotubes
(MWCNTs) and thereafter with the pH selective membrane (pHSM) and
the reference membrane (RM) on top, respectively (see the Supporting Information for the membrane compositions).
For the T sensor, two electrodes based on straight
Ag/AgCl ink were used. These electrodes were electrically connected
by drop casting MWCNTs dispersed in THF. For further details on the
manufacturing of each electrode, the reader is referred to the Supporting Information.
Figure 1
(a) Glucose biosensor.
In yellow, the WE1 modified with
the PB, the enzyme-based layer (GOx + CHI) and Nafion. In black, the
carbon-based CE. The RE1 is an Ag/AgCl rectangular path.
The electrode patterns are protected by a layer of adhesive tape.
(b) pH potentiometric sensor. In green, the WE2 modified
with the pH selective membrane (pHSM). In blue, the RE2 modified with the reference membrane (RM). (c) Temperature sensor
comprising two Ag/AgCl electrodes connected by a layer of MWCNTs.
(d) Image of the sensor array. (e) Images of the microfluidic cell
(containing a hole as the collection zone and the microfluidic channel)
and the skin patch (the electrodes are placed in the microfluidic
channel). (f) Illustration of the patch attached to the skin. Arrows
indicate the sweat flow direction. 1: Adhesive transfer tape. 2: Sweat
collection zone. 3: Sensor platform. 4: Microfluidic cell.
(a) Glucose biosensor.
In yellow, the WE1 modified with
the PB, the enzyme-based layer (GOx + CHI) and Nafion. In black, the
carbon-based CE. The RE1 is an Ag/AgCl rectangular path.
The electrode patterns are protected by a layer of adhesive tape.
(b) pH potentiometric sensor. In green, the WE2 modified
with the pH selective membrane (pHSM). In blue, the RE2 modified with the reference membrane (RM). (c) Temperature sensor
comprising two Ag/AgCl electrodes connected by a layer of MWCNTs.
(d) Image of the sensor array. (e) Images of the microfluidic cell
(containing a hole as the collection zone and the microfluidic channel)
and the skin patch (the electrodes are placed in the microfluidic
channel). (f) Illustration of the patch attached to the skin. Arrows
indicate the sweat flow direction. 1: Adhesive transfer tape. 2: Sweat
collection zone. 3: Sensor platform. 4: Microfluidic cell.
Implementation of the Sensors into the Microfluidic Cell for
On-Body Measurements
The sensor array (Figure d) was then attached to a 3D-printed microfluidic
cell (Figure e) by
using adhesive transfer tape (3 M 9471LE), so that the electrodes
were placed coinciding with the microfluidic channel (see the Supporting Information for specific details).
For on-body measurements, the device was attached to the individual
skin on the forehead by adhesive transfer tape (Figure S1 in the Supporting Information), thereby avoiding
any possible leaking of sweat and allowing for an adequate pressure
from the eccrine glands to provide continued and passive sweat flow
during perspiration.[16] A circular collection
zone (diameter of 0.1 mm, thickness of 1 mm) was used to sample the
sweat, being connected to a fluidic channel where the electrodes are
embedded (Figure e,f).
As the total area of the zone just before the electrodes is 133 mm2 with an approximate volume of 6.7 μL, the sweat will
start flowing through the surface of the first electrode after ca.
5 min of the perspiration starting. The fluidic channel contains an
internal volume of approximately 4 μL. Thus, considering the
initial time needed to fill in the collection area and a typical perspiration
rate of 1 μL cm–2 min–1 for
a midintensity, it will take ca. 10–13 min to display a reliable
sweat profile after the sweat reaches the surface of all of the electrodes
in the array.
Results and Discussion
The analytical
performance of the glucose biosensor was characterized
prior to its implementation in the microfluidic cell. The biosensor
presented a linear range of response from 10 to 1500 μM, with
a variation coefficient of less than 4% for the calibration parameters
(slope and intercept) for six consecutive calibration graphs using
the same electrode (Figure S2a,b in the
Supporting Information). Acceptable between-electrode reproducibility
(n = 6 different electrodes) was found, in all the
cases showing the same linear range of response (Figure S3a in the Supporting Information), which is important
toward on-body measurements. Furthermore, very low midterm drift was
displayed by the biosensor (100 μM glucose solution: 7.7 ±
1.1 nA h–1 over 2 h). The response time was calculated
to be in the range of 4–7 s, which reflects rather fast response.
Excellent reversibility was also obtained, with deviations of less
than 5% for the amperometric signal (Figure S2c,d in the Supporting Information). The glucose biosensor did not present
any response toward lactate, pyruvate, uric acid, and ascorbic acid
at the expected levels in sweat, and therefore, no matrix effect is
expected in real sweat measurements (see Figure S3b in the Supporting Information). Because pH and T changes in the sample are expected to influence the performance
of any enzymatic biosensor, the evaluation of the glucose biosensor
calibration under different pH and T conditions was
accomplished (Figure ). For example, the T in sweat may range from 26
to 38 °C depending on the body region, and pH varies from 4 to
7.5 between subjects.[3,7,17−19]Figure a,b show the dynamic response and corresponding calibration curves
for the biosensor after changing the pH of the background solution
(i.e., standard additions of glucose to either acetate of phosphate
buffers of different pH).
Figure 2
(a) Dynamic response of the biosensor toward
increasing glucose
concentrations in background of different pH. The final concentration
of glucose is indicated in each signal change. All the experiments
were accomplished using the same electrode. (b) Calibration graphs
corresponding to pH 7 and 4. (c) Dynamic response of the biosensor
at increasing glucose concentration in 10 mM phosphate buffer solution
(100 mM NaCl) at pH 6.6 and changing the T of the solution. All of
the experiments were accomplished using the same electrode. (d) Corresponding
calibration graphs. The biosensor was interrogated by applying a constant
potential of −0.05 V versus the Ag/AgCl reference electrode
and under stirred conditions.
(a) Dynamic response of the biosensor toward
increasing glucose
concentrations in background of different pH. The final concentration
of glucose is indicated in each signal change. All the experiments
were accomplished using the same electrode. (b) Calibration graphs
corresponding to pH 7 and 4. (c) Dynamic response of the biosensor
at increasing glucose concentration in 10 mM phosphate buffer solution
(100 mM NaCl) at pH 6.6 and changing the T of the solution. All of
the experiments were accomplished using the same electrode. (d) Corresponding
calibration graphs. The biosensor was interrogated by applying a constant
potential of −0.05 V versus the Ag/AgCl reference electrode
and under stirred conditions.There is a clear dependence of the amperometric signal with the
pH: the slope decreased and the first glucose concentration that the
biosensor was able to detect increased with more acidic pH background;
that is, the lower the pH, the more negative the baseline and this
creates that lower glucose concentration are not detected by the biosensor.
In principle, this is a direct consequence of the reduction of the
enzyme activity with the pH.[15]Figure c depicts
the dynamic response of the biosensor at increasing T in the sample. As observed, the amperometric signal was enhanced
with increasing T, which is in principle expected
owing to the increment of the enzymatic activity with T. Furthermore, we confirmed that drastic changes in either pH or T does not deteriorate the biosensor response and therefore,
any current change is reversible. For this purpose, we evaluated the
response at a fixed glucose concentration before and after changing
the pH from 7 to 4 and T from 22 to 37 °C (Figure S4 in the Supporting Information). Despite
the current changing in more than 100 nA for drastic changes in pH
and T, the initial response of the biosensor did
not deteriorate and was rather reversible, thus recovering the initial
current value after drastic changes in pH and T (change
in the initial current lower than 1% was observed).
In Vitro Evaluation of
the Analytical Performance of the Glucose
Biosensor after Implementation in the Microfluidic Cell
To
characterize the analytical performance of the glucose biosensor once
implemented in the microfluidic cell (i.e., flow mode), a flow rate
of 4 μL min–1 (achieved by means of a peristaltic
pump) was selected. This flow rate corresponds to a sweat rate of
3 μL cm–2 min–1, which is
close to reported values of 2.4 μL cm–2 min–1 achieved under mild activity conditions during on-body
tests accomplished on the forehead.[20]Figure a presents four successive
dynamic responses (replicates) at increasing glucose concentrations
in artificial sweat background, and Figure b shows the corresponding calibration graphs
with the error bars displaying the standard deviation for the measurements.
The repeatability of the calibration parameters was rather good (less
than 3% of variation, slope of 1.14 ± 0.03 nA μM–1 and intercept of −50.7 ± 1.1 nA), while keeping the
linear range of response within the expected physiological range in
sweat: from 10 to 200 μM.[21,22] Other reported glucose
biosensors, slightly exceed the lower concentration of this range.[23] The response time of the biosensor in flow mode
was lower than 20 s along the linear range of response, considering
all the time traces obtained along this work. This is a fast response
and indeed very convenient for the acquisition of real time data during
on-body tests.
Figure 3
(a) Successive dynamic responses (replicates, n = 4) of the glucose biosensor under flow mode (4 μL
min–1) in artificial sweat background (pH 6.6).
The final
concentration of glucose is indicated in each signal change. (b) Average
calibration curve with error bars showing the standard deviation (slope
of −1.14 ± 0.03 nA μM–1 and intercept
of −50.7 ± 1.1 nA). (b) Dynamic response of the glucose
biosensor at increasing flow rates in artificial sweat background
(pH 6.6). (c) Average calibration curve observed at increasing flow
rates (slope of −1.36 ± 0.13 nA μM–1 and intercept of −49.4 ± 1.0 nA). The final concentration
of glucose is indicated in each signal change. (d) Dynamic response
and average calibration graph observed in the reversibility test (4
μL min–1) in artificial sweat background (pH
6.6): Slope of −1.45 ± 0.01 nA μM–1 and intercept of −43.9 ± 1.8 nA. (e) Dynamic response
and average calibration graph observed before and after torsion strain
to the microfluidic cell containing the glucose biosensor (4 μL
min–1) in artificial sweat background (pH 6.6):
Slope of −1.49 ± 0.13 nA μM–1 and
intercept of −23.2 ± 1.4 nA. The final concentration of
glucose is indicated in each signal change. (f) Dynamic responses
at decreasing pH. (g) Corresponding calibration graphs. (h) Dynamic
response at different T in artificial sweat background (pH 6.6). (i)
Corresponding calibration graphs.
(a) Successive dynamic responses (replicates, n = 4) of the glucose biosensor under flow mode (4 μL
min–1) in artificial sweat background (pH 6.6).
The final
concentration of glucose is indicated in each signal change. (b) Average
calibration curve with error bars showing the standard deviation (slope
of −1.14 ± 0.03 nA μM–1 and intercept
of −50.7 ± 1.1 nA). (b) Dynamic response of the glucose
biosensor at increasing flow rates in artificial sweat background
(pH 6.6). (c) Average calibration curve observed at increasing flow
rates (slope of −1.36 ± 0.13 nA μM–1 and intercept of −49.4 ± 1.0 nA). The final concentration
of glucose is indicated in each signal change. (d) Dynamic response
and average calibration graph observed in the reversibility test (4
μL min–1) in artificial sweat background (pH
6.6): Slope of −1.45 ± 0.01 nA μM–1 and intercept of −43.9 ± 1.8 nA. (e) Dynamic response
and average calibration graph observed before and after torsion strain
to the microfluidic cell containing the glucose biosensor (4 μL
min–1) in artificial sweat background (pH 6.6):
Slope of −1.49 ± 0.13 nA μM–1 and
intercept of −23.2 ± 1.4 nA. The final concentration of
glucose is indicated in each signal change. (f) Dynamic responses
at decreasing pH. (g) Corresponding calibration graphs. (h) Dynamic
response at different T in artificial sweat background (pH 6.6). (i)
Corresponding calibration graphs.During sport practice (or workout), slight changes in the sweating
rate of the subject are expected.[24] Therefore,
the response of the biosensor was evaluated at three different flow
rates (4, 6, and 8 μL min–1, selected as per
mimicking real sweat rates in the body during sport practice[24]). Figure c depicts the dynamic responses at the different flow rates,
and Figure d shows
the average calibration graph. As observed, the response of the biosensor
increased with the flow rate and this is specially noticed at higher
glucose concentrations (from 100 μM). According to the standard
deviations observed in the calibration parameters of the average calibration
graph (slope of 1.36 ± 0.13 nA μM–1 and
intercept of −49.4 ± 1.1 nA), it was calculated that,
the most drastic change in the sweat rate may induce an error of less
than the 10% in the calculation of the glucose concentration, which
is an acceptable level in precision considering further physiological
measurements.[25] This error will specially
affect higher concentrations (>100 μM), which indeed were
never
observed in any of the on-body tests developed in the present work.
The glucose conversion rate at the enzyme layer is connected to the
concentration gradient of glucose in the external membrane phase (Nafion),
and it is expected that at higher flow rates the glucose concentration
gradient is more established and sustained given the faster replenishment
of glucose once the aqueous diffusion layer is reduced.Beyond
any change in the flow rate, the between-electrode reproducibility
(using different twin electrodes) was found to be rather good, with
less than 1.2% of change in the calibration parameters (see Figure S5 in the Supporting Information). Then,
a reversibility test was performed by changing the concentration of
glucose from 25 to 150 μM and from 150 to 25 μM along
three cycles (see Figure e). The biosensor exhibited a very reversible response with
a change in the calibration parameters of less than 4%: average calibration
graph with slope of 1.45 ± 0.01 nA μM–1 and intercept of 43.9 ± 1.7 nA. In addition, the biosensor
presented excellent drift in flow mode at 4 μL min–1 (1.3 nA h–1 over 2.5 h, even lower value than
in batch mode). Advantageously, the analytical response of the biosensor
is maintained after applying a strong torsion strain. Thus, the calibration
graph was found to be maintained in the accomplished resilience test
(see Figure S6 in the Supporting Information):
less than 5% of change in the calibration parameters.Similar
as in batch mode, the influence of the pH and T in
the response under flow conditions was investigated (Figures f–i). In
both cases, if the pH or the T decreases significantly
in sweat, the biosensor presents a loss of sensitivity together with
a shift of the intercept (or background signal) to more negative currents.
Importantly, the trends found for both pH and T influences
will be subsequently used to elaborate a correction of the amperometric
response of the glucose biosensor.
pH- and Temperature Correction
for the Amperometric Response
of the Glucose Biosensor
Sensors for monitoring pH and T fluctuations in the sweat analyzed by the glucose biosensor
were implemented in the microfluidic cell (i.e., the entire epidermal
patch). The details for the handmade fabrication are provided in the Supporting Information. Briefly, the pH sensor
consists of a potentiometric one based on a WE2 and a RE2 prepared as reported elsewhere[6] but on the same polyester substrate used to fabricate the biosensor.
The temperature sensor is based on a chemoresistor principle, where
the resistance of carbon nanonotubes linearly depends on the temperature.[26,27]The pH sensor displayed a reproducible near-Nernstian slope
of −53.8 ± 0.1 mV and intercept of 343.5 ± 16.3 mV
(Figures a and S7a in the Supporting Information, n = 3), with excellent reversibility of the calibration parameters
(slope of −52.9 ± 0.6 mV and intercept of 325.0 ±
4.2 mV, Figure S7b in the Supporting Information)
within a linear range of response from pH 4.5 to 8.1, which contains
the physiological range expected for pH in sweat (pH 4.5–7.5).[4] Good performance was also observed for the T sensor: slope of −4.4 ± 0.1 kΩ °C–1 and intercept of 1063.4 ± 0.2 kΩ for three
successive calibrations (Figure b), within a linear range from 19 to 43 °C and
showing excellent reversibility (less than 3.5% of variation in the
calibration parameters, see Figure S8 in
the Supporting Information). Both sensors presented neglectable influence
of torsion application on the calibration parameters (applying the
same resilience test as for the glucose biosensor, see Figure S6 in the Supporting Information, calibration
parameters changed <1.2% in both cases).
Figure 4
(a) Calibration graph
of the pH sensor (T = 20
°C). (b) Calibration graph of the T sensor in
artificial sweat background (pH 6.6). (c) Correction factors (f) as a function of sweat pH for the calibration parameters
of the glucose biosensor. (d) Correction factors (g) as a function of sweat T for the calibration parameters
of the glucose biosensor.
(a) Calibration graph
of the pH sensor (T = 20
°C). (b) Calibration graph of the T sensor in
artificial sweat background (pH 6.6). (c) Correction factors (f) as a function of sweat pH for the calibration parameters
of the glucose biosensor. (d) Correction factors (g) as a function of sweat T for the calibration parameters
of the glucose biosensor.Next, we elaborate on the pH and T correction
of the glucose concentration dynamic assuming that the effects of
pH and T on the biosensor (i.e., the enzyme) are
independent from each other and that every single glucose biosensor
presents the same kind of influence from pH and T variations. Accordingly, it is possible to universally tabulate
the individual effect of pH and T in the slope and
intercept of every glucose biosensor by calculating a factor of variation
(fslope and fintercept for the pH and gslope and gintercept for the T) at each condition
with respect to the reference conditions of pH 6.6 and T of 20 °C,
using the following equations:Figure c,d
shows
the correction factors for the slope and intercept calculated at different
pH and T in sweat, from the data collected in Figure f–i and using eq –4. Notably, the errors bars refer to the measurements of three
different electrodes. As observed, the pH has a larger effect than
the T on the biosensor response, specifically in
the intercept, and this is manifested in higher correction factors.Considering now pH and T effects in accumulative way, it is possible
to correct the entire response of the biosensor on the basis of recalculating
the slope and intercept (k and m respectively) of the initial off-body calibration graph in artificial
sweat (at pH 6.6 and 20 °C for reference). Thus, by applying
a first-order correction algorithm, we dynamically calculate the corrected
slopes k′ and intercepts m’ that were finally used to obtain the dynamic glucose concentration
in sweat over the on-body test, using the following equations:Notably,
we selected to accomplish the off-body
calibration of the sensors before the on-body tests to check its proper
functioning. Indeed, the calibration graph was demonstrated to be
maintained after the assessment of the on-body test, with a variation
of less than the 3% of the slope and intercept (see Figure S9 in the Supporting Information as an example). Then,
for the dynamic correction, the calculation of the f and g factors at every pH and T was performed by considering linearity between the two experimental
factors that are the closest to the pH and T range measured in the
corresponding on-body tests (i.e., assuming linearity between two
close points and accomplishing extra or interpolation). For example,
if the pH range is found to be 5.7–5.9 in the on-body test,
we took the correction factors calculated at 5.5 and 6.6 for the slope
and the intercept (Figure c) and interpolate the dynamic correction in the fitting line
between these two points. Whether the T is measured
to be in the range 32–34 °C, we took the correction factors
calculated at 30 and 33 °C for the slope and the intercept (Figure d) and extrapolate
the dynamic correction using the fitting line between these two points.
In addition, for clarification, the reader can find the entire calculation
process for one particular glucose concentration in the Supporting Information. The same calculations
are applied to every single point of each dynamic on-body profile
with the help of a mathematical code in Matlab.In an ideal
situation, it would be necessary to fit the data to
a more complex algorithm that considers join influences of pH and T, to better cover the effect in the enzyme activity.[25,26] For this purpose, a bigger data bank comprising a larger number
of experiments at different pH and T would be necessary.
In addition, it would be important to include a correction based on
the influence of the perspiration rate in the correction factors (by
adding an extra sensor into the epidermal patch to quantify the sweat
flow rate on the sensors’ surface), while for resilience effect
seems not necessary to be incorporated in the final correction (small
changes in the calibration parameters were detected for drastic torsion
deformations, exaggerating those movement occurring in the forehead
of the athlete). Overall, the aim of the present work is to demonstrate
that pH and T corrections are needed to approximate
the results observed in on-body measurements to those obtained with
the gold standard techniques.
Validation of the Glucose
Biosensor Operating off- and on-Bodily
The developed epidermal
patch for glucose detection was first validated
under off-body operation, i.e., by collecting sweat samples and analyzing
them with the epidermal patch operating by means of the peristaltic
pump and with a gold standard technique (UV–vis[28] and/or IC,[9] as detailed
in the Supporting Information). Notably,
other authors have already reported on this approach for the glucose
biosensors validation, but with the absence of pure on-body validation.[7] In a second step, the epidermal patch on-bodily
operating in 9 tests was accomplished. The workout program in the
bike (T1-T9) was as follows: (i) a 5 min warm up, (ii) 35 min low-
to midintensity workup (65–75% of maximum heart rate), and
(iii) a cool down step. The heartrate of the subject was monitored
during the entire exercise program.For the sweat collection,
a modification of the regional absorbent patch method was used as
reported elsewhere.[6] Briefly, an absorbent
patch (absorbent material mounted in Hydrofilm tape) was placed on
the forehead of the subject, while he/she was practicing exercise
in a static bike, and was replaced every 10 min. Then, after detachment,
the absorbent material was squeezed with a syringe into vial to extract
the sample, which was briefly stored at 4 °C until analysis.
More details about the sweat collection method are provided in the Supporting Information. Before starting the sport
practice (time 0 min), because the subject was not sweating yet, iontophoresis
was applied to the arm of the individual to collect sweat. Blood tests
were accomplished at time 0 min and also during the sport practice
without stopping the exercise.Figure a displays
the correlation between the values found for glucose content in 16
different samples (collected during T1–T4 based on subject
1 and 2) analyzed with the epidermal patch (glucose biosensor) operating
off-site the body (using the peristaltic pump) and the UV–vis
method. Rather good correlation was obtained, as pointed out by Pearson
correlation coefficient of 0.90. Table S1 in the Supporting Information displays the obtained values for glucose
concentration and the percent difference between both techniques,
being on average lower than the 10% and therefore confirming the observed
good correlation, while raw (not-corrected) data are further from
the gold standard technique.
Figure 5
(a) Correlation in the off-body validation (n =
16). (b) Correlation in the on-body validation (n = 43). Profiles observed in the on-body test of subject 1 (T1):
(c) Heart rate of the subject during the workout, which correlates
the practice with the physical level (I, warm-up; II, low-intensity;
III, midintensity; and IV, cool-down). (d) Sweat temperature. (e)
Sweat pH. (f) Sweat glucose. Environmental conditions of 20 °C,
relative humidity of 70%, 50 min cycling.
(a) Correlation in the off-body validation (n =
16). (b) Correlation in the on-body validation (n = 43). Profiles observed in the on-body test of subject 1 (T1):
(c) Heart rate of the subject during the workout, which correlates
the practice with the physical level (I, warm-up; II, low-intensity;
III, midintensity; and IV, cool-down). (d) Sweat temperature. (e)
Sweat pH. (f) Sweat glucose. Environmental conditions of 20 °C,
relative humidity of 70%, 50 min cycling.A total number of 9 on-body tests (T1-T9) were accomplished using
the biosensor attached to the forehead of the subject together with
the sampling pad. Before every test, all of the sensors were externally
calibrated (see the Supporting Information). Figure b depicts
the correlation of the glucose concentration found with the biosensor
and the collected sweat samples analyzed by the gold standard technique
(UV–vis or IC), corresponding this to the validation of real
on-body measurements. Rather good correlation was obtained, as pointed
out by Pearson correlation coefficient of 0.85 calculated for 43 samples
(see Table S2 in the Supporting Information
for the entire list of the raw data).Studying now each on-body
test (T1-T9) more in detail, Figure c depicts the heartrate
monitored over T1, which may be used as a measure of the level of
exercise, whereas Figure d–f shows the dynamic profiles for sweat temperature,
pH, and glucose concentration, respectively. In addition, the glucose
profile before and after the correction approach that considers dynamic
pH and T fluctuations in sweat, together with sample
analysis with UV–vis are provided in Figure f. As observed, the sweat temperature remained
relatively constant over the entire on-body test (28.5 °C in
average), whereas the pH experienced a decrease coinciding with the
cool-dawn part of the workout (approximately from 6.8 to 6.4). Regarding
the glucose profile, the concentration is higher at the beginning
of the test (65 μM in average), followed by a decrease until
a rather constant value after 20 min of the workout (25 μM in
average), coinciding with the low/and mid/intensity activity, and
further showing only minor fluctuations over the rest of the exercise.
This trend is indeed expected, owing to the dilution of glucose in
the eccrine glands of the skin while physical exercise continues.[16]A comparison of the raw dynamic profile
for glucose and that calculated
after pH and T corrections with the measurements
obtained with UV–vis (Figure f) revealed that, indeed, the corrected profile lies
closer to the measurements provided by the standard method (less than
8% in percent difference, see Table S2 in
the Supporting Information). Specifically, without pH and T correction, the percent difference with the standard technique
range in the order of 20–30%. This trend was also observed
for all of the 9 on-body tests accomplished in the present work even
founding higher percentage differences (see Table S2 in the Supporting Information), therefore confirming the
need for the developed “correction approach” for the
response of the glucose biosensor. Any deviations from the values
obtained with the gold standard technique may come from the further
need of a joined pH–T correction together
with any change in the perspiration rate, while contributing this
latter in a lower extent to the final reliability of the outcomes
(see above). Correction on the slope of the pH-selective electrode
according to the T could be also considered toward
an improved version of the presented correction, while adding complexity
to the final algorithm.Figure presents
the glucose dynamic profiles in sweat after pH and T correction that
were observed in 8 different on-body tests (T2-T9) at different conditions. Figure a–c show the
glucose profiles for the same subject at different times after food
intake: 4, 3, and 1 h after lunch (T2-T4). A similar trend as that
described in T1 was observed: initially, there is a higher glucose
level, followed by a decrease during work-out, with the exception
of the final point in T4, where an increase is observed during the
cool down step. With no dependence on the time selected to assess
the on-body test, the glucose level in the subject was always between
10 and 30 μM. Notably, on-body observations rather agree with
UV–vis and the biosensor (in the epidermal patch) operating
off-body, especially for longer workout times (see Table S1 in the Supporting Information).
Figure 6
Dynamic glucose profiles
observed during different on-body tests.
Environmental conditions of 20 °C and relative humidity of 70%.
Panels a–c correspond to the same subject practicing sport
after different times of having lunch; panels d–f correspond
to different subjects in the morning; and panels g and h correspond
to other two subjects in the morning, with the difference that they
took a sugary drink close to the 20 min of the sport practice.
Dynamic glucose profiles
observed during different on-body tests.
Environmental conditions of 20 °C and relative humidity of 70%.
Panels a–c correspond to the same subject practicing sport
after different times of having lunch; panels d–f correspond
to different subjects in the morning; and panels g and h correspond
to other two subjects in the morning, with the difference that they
took a sugary drink close to the 20 min of the sport practice.Different glucose levels were found for each subject
(Figures and 6), and always within the expected physiological
range:[21,22] subject 1 (T1) from 20 to 30 μM, subject
2 (T2-T4) from 10
to 30 μM, subject 3 (T5) from 10 to 25 μM, subject 4 (T6)
from 10 to 20 μM, subject 5 (T7) constant level of approximately
30 μM, subject 6 (T8) constant level of approximately 20 μM,
and subject 7 (T9) from 25 to 50 constant level of approximately 30
μM. Interestingly, all of the subjects follow the same trend
of presenting a higher concentration at the start of the workout,
followed by a stabilization at a lower glucose concentration. Regarding
the intake of sugary drinks during the workout (Figure g,h), whereas subject 6 (T8) did not manifest
a significant increase in glucose levels after the intake, subject
7 (T9) presented a gradual increase of almost double in the glucose
concentration. Both subjects took 50 cL of the same commercially available
isotonic drink. Some results previously pointed out that glucose taken
during sport practice is quickly burnt, and therefore, it is not reflected
in sweat glucose changes. Indeed, only a high consumption of sugar
may lead to measurable changes in sweat glucose.[29] So far, our results likely reflect the different metabolism
in each subject.Variations found in pH and T between the different
subjects as
well as over the exercise practice of one subject rather justify the
use of the developed correction approach. For example, in subjects
3–7 (Figure S11 in the Supporting
Information), we found a pH range from 5.5 to 8.3 and a T range from 31 to 36 °C, with variations close to 0.8 pH units
and 2 °C in the most extreme changes during the workout. Notably,
it would be convenient to widen the calibration graph for pH up to
8.5 to encompas the entire pH range in further on-body experiments.In particular, the highest deviation from the values obtained with
the standard methods are observed in T9 from 30 min of the exercise
(see Figure h), where
both pH is low and the temperature high. Most likely, this points
out the need for an algorithm contemplating the joint effect of pH
and T for the correction of the biosensor response.
Indeed, the consideration of the change in kinetics of both the enzyme[30] and the mediator[31,32] activity/behavior
could be essential in such extreme case.In all of the on-body
tests, blood glucose was also analyzed at
the beginning, middle, and end of the exercise time by means of a
commercial glucometer (see the Supporting Information). In addition, glucose level in sweat before the workout was measured
in sweat samples collected by iontophoresis using a commercially available
device with clinical use approval.[33] The
iontophoresis device was placed in the arm of the individual following
the instructions of the manufacturer. Then, the sweat sample was introduced
in the epidermal patch by means of the peristaltic pump and measure
with the glucose biosensor.Figure a displays
the glucose values measured in each test, whereas Figure b presents the molar ratio
between the glucose concentration in sweat and blood. Several trends
are observed. First, the glucose amount in blood was always higher
than in sweat, presenting molar ratios between 7 and 27 (×103) in the initial measurements. Subsequently, the same trend
as for the sweat glucose in the on-body experiments was found: a decrease
and a maintenance of the glucose concentration, somehow qualitatively
confirming the outcomes of the wearable sensor during the 9 on-body
tests.
Figure 7
(a) Glucose concentration in blood measured with the glucometer
at different times over the T1-T9. (b) Sweat/blood glucose concentration
ratio observed at different times over T1-T9. (c) Correlation between
blood and sweat concentrations (n = 26). (d) Correlation
between blood and sweat concentrations observed at time 0 min for
every subject (n = 6) and for only one subject over
time (n = 5). Notably, the samples corresponding
to time 0 min were collected by iontophoresis.
(a) Glucose concentration in blood measured with the glucometer
at different times over the T1-T9. (b) Sweat/blood glucose concentration
ratio observed at different times over T1-T9. (c) Correlation between
blood and sweat concentrations (n = 26). (d) Correlation
between blood and sweat concentrations observed at time 0 min for
every subject (n = 6) and for only one subject over
time (n = 5). Notably, the samples corresponding
to time 0 min were collected by iontophoresis.However, no clear correlation was found between sweat and blood
glucose concentration further than the same qualitative trend. When
all the values were correlated (n = 26, Figure c), a Pearson coefficient
of 0.55 was obtained, which does not confirm any clear correlation
for the glucose measured in the two biological fluids. Indeed, if
the correlation is restricted to only the initial measurements at t = 0 min (n = 7) or one subject (n = 5, T8), as shown in Figure d, the existence of a poor correlation was
manifested with Person coefficients of 0.27 and 0.22, respectively.
We are not the first to point out a poor correlation between glucose
concentration in blood and sweat in individuals practicing sport.[7] On the other hand, there are some reports on
existing glucose sweat and blood correlation (in absence of sport
practice) claiming that any absence of correlation is due to improper
sweat sampling.[34,35] Nevertheless, with the epidermal
patch, no sampling is needed. Overall, massive studies are mandatory
to clarify such correlation. Nevertheless, the main aim of the present
work is not on the physiological side but the analytical development
and evaluation of more reliable glucose measurements in sweat.
Conclusions
More reliable results are observed when the amperometric response
of a wearable glucose biosensor is corrected by real-time fluctuations
of pH and T in sweat during on-body measurements.
Accordingly, a new epidermal patch containing a glucose biosensor
as well as pH and T sensors implemented in a flexible
microfluidic cell is herein proposed as a source for reliable physiological
outcomes during sport practice. Our results are supported by a double
validation accomplished with the sensing device operating off- and
on-body. A number of 9 on-body tests involving different subjects
under distinct conditions demonstrated rather good correlation with
gold standard techniques. The developed methodology is expected to
impact the design of the next generation of wearable chemical sensors,
in particular those based on enzymatic sensors, aimed at more trustworthy
data for sport and healthcare applications. To date, the success of
wearable devices based on chemical sensors is based on the handling
by the same researchers who invented the device. Importantly, pH and T correction will help these devices to work more independently
while providing accurate outcomes.
Authors: James Moyer; Donald Wilson; Irina Finkelshtein; Bruce Wong; Russell Potts Journal: Diabetes Technol Ther Date: 2012-02-29 Impact factor: 6.118
Authors: Lindsay B Baker; Corey T Ungaro; Bridget C Sopeña; Ryan P Nuccio; Adam J Reimel; James M Carter; John R Stofan; Kelly A Barnes Journal: J Appl Physiol (1985) Date: 2018-02-08
Authors: Murat A Yokus; Tanner Songkakul; Vladimir A Pozdin; Alper Bozkurt; Michael A Daniele Journal: Biosens Bioelectron Date: 2020-01-18 Impact factor: 10.618
Authors: Hnin Yin Yin Nyein; Mallika Bariya; Liisa Kivimäki; Sanna Uusitalo; Tiffany Sun Liaw; Elina Jansson; Christine Heera Ahn; John A Hangasky; Jiangqi Zhao; Yuanjing Lin; Tuomas Happonen; Minghan Chao; Christina Liedert; Yingbo Zhao; Li-Chia Tai; Jussi Hiltunen; Ali Javey Journal: Sci Adv Date: 2019-08-16 Impact factor: 14.136
Authors: Farshad Tehrani; Hazhir Teymourian; Brian Wuerstle; Jonathan Kavner; Ravi Patel; Allison Furmidge; Reza Aghavali; Hamed Hosseini-Toudeshki; Christopher Brown; Fangyu Zhang; Kuldeep Mahato; Zhengxing Li; Abbas Barfidokht; Lu Yin; Paul Warren; Nickey Huang; Zina Patel; Patrick P Mercier; Joseph Wang Journal: Nat Biomed Eng Date: 2022-05-09 Impact factor: 25.671
Authors: Roozbeh Ghaffari; Da Som Yang; Joohee Kim; Amer Mansour; John A Wright; Jeffrey B Model; Donald E Wright; John A Rogers; Tyler R Ray Journal: ACS Sens Date: 2021-08-05 Impact factor: 9.618
Authors: Tiziana Fuoco; Maria Cuartero; Marc Parrilla; Juan José García-Guzmán; Gaston A Crespo; Anna Finne-Wistrand Journal: Biomacromolecules Date: 2021-01-27 Impact factor: 6.988
Authors: Le Minh Tu Phan; Thuy Anh Thu Vo; Thi Xoan Hoang; Sathish Panneer Selvam; Hoang Lan Pham; Jae Young Kim; Sungbo Cho Journal: Adv Mater Technol Date: 2021-05-06