Literature DB >> 18249300

Image quality produced by different cone-beam computed tomography settings.

Jeffrey C Kwong, J Martin Palomo, Michael A Landers, Alex Figueroa, Mark G Hans.   

Abstract

INTRODUCTION: The aim of this study was to evaluate image quality at different cone-beam computed tomography settings and 3 fields of view.
METHODS: A Hitachi CB MercuRay (Hitachi Medical Systems, Tokyo, Japan) was modified to allow different setting combinations. The variables consisted of 4 milliampere settings (2, 5, 10, and 15 mA), 2 kilovolt (peak) settings (100 and 120 kV[p]), presence or absence of a copper filter, and 3 fields of view (6, 9, and 12 in). Thirty-two scans were taken on a cadaver head and 16 scans on a dry skull. The groups were divided by field of view, and the images were ranked by at least 30 judges. Diagnostic quality was addressed in a questionnaire. Descriptive statistics and rankings were calculated with Excel 2003 (Microsoft, Redmond, Wash) and the Friedman and Wilcoxon signed rank tests with SPSS software (version 14.0.1; SPSS, Chicago, Ill).
RESULTS: The presence or absence of a filter showed significant differences (P <.006) in 2 pairs of the 9-in field of view. Variation in kilovolt (peak) settings showed significant differences (P <.006) in the 6-in 5-mA images with a filter. Altering the milliampere settings showed significant differences (P <.008) in the 6- and 12-in groups. The 9-in group showed significant differences between 2 mA and 10 and 15 mA. Overall, the 6-, 9-, and 12-in images had diagnostic quality 56%, 99%, and 99% of the time, respectively.
CONCLUSIONS: Presence or absence of a filter and the kilovolt (peak) setting did not affect overall image quality. Images taken at lower milliampere settings showed good diagnostic quality.

Entities:  

Mesh:

Year:  2008        PMID: 18249300     DOI: 10.1016/j.ajodo.2007.02.053

Source DB:  PubMed          Journal:  Am J Orthod Dentofacial Orthop        ISSN: 0889-5406            Impact factor:   2.650


  30 in total

1.  Effect of electric potential and current on mandibular linear measurements in cone beam CT.

Authors:  S Panmekiate; W Apinhasmit; A Petersson
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

Review 2.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

3.  Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children.

Authors:  Jose A Hidalgo Rivas; Keith Horner; Badri Thiruvenkatachari; Jonathan Davies; Chrysoula Theodorakou
Journal:  Br J Radiol       Date:  2015-08-17       Impact factor: 3.039

4.  Optimization of exposure parameters for cone beam computed tomography sialography.

Authors:  F M Jadu; M L Hill; M J Yaffe; Ernest W N Lam
Journal:  Dentomaxillofac Radiol       Date:  2011-09       Impact factor: 2.419

5.  Influence of CBCT parameters on the output of an automatic edge-detection-based endodontic segmentation.

Authors:  J Michetti; M Georgelin-Gurgel; J-P Mallet; F Diemer; K Boulanouar
Journal:  Dentomaxillofac Radiol       Date:  2015-06-02       Impact factor: 2.419

6.  Impact of prone, supine and oblique patient positioning on CBCT image quality, contrast-to-noise ratio and figure of merit value in the maxillofacial region.

Authors:  Juha Koivisto; Maureen van Eijnatten; Jorma Järnstedt; Kirsi Holli-Helenius; Prasun Dastidar; Jan Wolff
Journal:  Dentomaxillofac Radiol       Date:  2017-04-07       Impact factor: 2.419

7.  Image quality optimization using a narrow vertical detector dental cone-beam CT.

Authors:  Danieli Moura Brasil; Ruben Pauwels; Wim Coucke; Francisco Haiter-Neto; Reinhilde Jacobs
Journal:  Dentomaxillofac Radiol       Date:  2019-01-31       Impact factor: 2.419

8.  Short-term and long-term effects of rapid maxillary expansion on the nasal soft and hard tissue.

Authors:  Cassie T Truong; Hyeran H Jeon; Puttipong Sripinun; Ann Tierney; Normand S Boucher
Journal:  Angle Orthod       Date:  2021-01-01       Impact factor: 2.079

Review 9.  Dose optimization by altering the operating potential and tube current exposure time product in dental cone beam CT: a systematic review.

Authors:  Rebekah Goulston; Jonathan Davies; Keith Horner; Frederick Murphy
Journal:  Dentomaxillofac Radiol       Date:  2016-01-06       Impact factor: 2.419

10.  Influence of scanning and reconstruction parameters on quality of three-dimensional surface models of the dental arches from cone beam computed tomography.

Authors:  Bassam Hassan; Paulo Couto Souza; Reinhilde Jacobs; Soraya de Azambuja Berti; Paul van der Stelt
Journal:  Clin Oral Investig       Date:  2009-06-09       Impact factor: 3.573

View more