Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro.

Jaideep Sur1, Kenji Seki, Hiroshi Koizumi, Koh Nakajima, Tomohiro Okano.   

Abstract

OBJECTIVE: The purpose of this study was to evaluate the effect of tube current reduction on the quality of cone-beam computerized tomography (CBCT) images of the maxilla and mandible for presurgical implant planning. STUDY
DESIGN: Six cadaver heads were imaged with a CBCT (3D Accuitomo) operated at 80 kVp and variable tube current (8, 4, 2, or 1 mA) in full-scan mode. For the posterior region of maxilla and mandible, half-scans were also performed at similar settings. The images were evaluated independently by 5 oral radiologists for 15 anatomic landmarks regarding presurgical implant planning. The quality of images were compared with their respective reference images at 8 mA and ranked on a 4-point rating scale as excellent, good, fair, or nondiagnostic. The scores of all observers were averaged for each landmark at every exposure condition, and Bonferroni test (P<.05) was performed.
RESULTS: The 4 mA images at full-scan mode could visualize each landmark of maxilla and mandible and were evaluated to be the same or almost equivalent in quality as the 8 mA images. Even 2 mA images in full-scan mode and 4 mA in half-scan mode could be used for implant planning. The 1 mA images were unacceptable owing to the substantial degradation in image quality.
CONCLUSION: Significant dose reduction can be achieved by reducing tube current without substantial loss of image quality for presurgical implant planning in CBCT. Crown Copyright (c) 2010. Published by Mosby, Inc. All rights reserved.

Mesh:

Year:  2010        PMID: 20598589     DOI: 10.1016/j.tripleo.2010.03.041

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  13 in total

1.  Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site.

Authors:  A Dawood; J Brown; V Sauret-Jackson; S Purkayastha
Journal:  Dentomaxillofac Radiol       Date:  2011-12-19       Impact factor: 2.419

2.  Effect of electric potential and current on mandibular linear measurements in cone beam CT.

Authors:  S Panmekiate; W Apinhasmit; A Petersson
Journal:  Dentomaxillofac Radiol       Date:  2012-04-12       Impact factor: 2.419

3.  Development of a low-dose protocol for cone beam CT examinations of the anterior maxilla in children.

Authors:  Jose A Hidalgo Rivas; Keith Horner; Badri Thiruvenkatachari; Jonathan Davies; Chrysoula Theodorakou
Journal:  Br J Radiol       Date:  2015-08-17       Impact factor: 3.039

4.  Analysis of the priority of anatomic structures according to the diagnostic task in cone-beam computed tomographic images.

Authors:  Jin-Woo Choi
Journal:  Imaging Sci Dent       Date:  2016-12-20

5.  Effects of exposure parameters and slice thickness on detecting clear and unclear mandibular canals using cone beam CT.

Authors:  Gainer R Jasa; Mayumi Shimizu; Kazutoshi Okamura; Kenji Tokumori; Yohei Takeshita; Warangkana Weerawanich; Kazunori Yoshiura
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

Review 6.  Dose optimization by altering the operating potential and tube current exposure time product in dental cone beam CT: a systematic review.

Authors:  Rebekah Goulston; Jonathan Davies; Keith Horner; Frederick Murphy
Journal:  Dentomaxillofac Radiol       Date:  2016-01-06       Impact factor: 2.419

7.  A pragmatic approach to determine the optimal kVp in cone beam CT: balancing contrast-to-noise ratio and radiation dose.

Authors:  R Pauwels; O Silkosessak; R Jacobs; R Bogaerts; H Bosmans; S Panmekiate
Journal:  Dentomaxillofac Radiol       Date:  2014-04-08       Impact factor: 2.419

8.  Effect of Milliamperage Reduction on Pre-surgical Implant Planning Using Cone Beam Computed Tomography by Surgeons of Varying Experience.

Authors:  N El Sahili; S David-Tchouda; S Thoret; I Nasseh; A Berberi; T Fortin
Journal:  J Maxillofac Oral Surg       Date:  2017-12-20

9.  Artefacts at different distances from titanium and zirconia implants in cone-beam computed tomography: effect of tube current and metal artefact reduction.

Authors:  Arthur Xavier Maseti Mancini; Matheus Urias Cruz Santos; Hugo Gaêta-Araujo; Camila Tirapelli; Ruben Pauwels; Christiano Oliveira-Santos
Journal:  Clin Oral Investig       Date:  2021-02-05       Impact factor: 3.573

10.  Impact of Cone Beam Computed Tomography Dose in Pre-Surgical Implant Analysis.

Authors:  Nabil El Sahili; Ibrahim Nasseh; Antoine Berberi; Sandra David-Tchouda; Sophie Thoret; Thomas Fortin
Journal:  Open Dent J       Date:  2018-01-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.