Literature DB >> 33440056

Vendor-independent skin dose mapping application for interventional radiology and cardiology.

Marko Krajinović1,2, Nikola Kržanović1,2, Olivera Ciraj-Bjelac1,2.   

Abstract

PURPOSE: The purpose of this paper is to present and validate an originally developed application SkinCare used for skin dose mapping in interventional procedures, which are associated with relatively high radiation doses to the patient's skin and possible skin reactions.
METHODS: SkinCare is an application tool for generating skin dose maps following interventional radiology and cardiology procedures using the realistic 3D patient models. Skin dose is calculated using data from Digital Imaging and Communications in Medicine (DICOM) Radiation Dose Structured Reports (RDSRs). SkinCare validation was performed by using the data from the Siemens Artis Zee Biplane fluoroscopy system and conducting "Acceptance and quality control protocols for skin dose calculating software solutions in interventional cardiology" developed and tested in the frame of the VERIDIC project. XR-RV3 Gafchromic films were used as dosimeters to compare peak skin doses (PSDs) and dose maps obtained through measurements and calculations. DICOM RDSRs from four fluoroscopy systems of different vendors (Canon, GE, Philips, and Siemens) were used for the development of the SkinCare and for the comparison of skin dose maps generated using SkinCare to skin dose maps generated by different commercial software tools (Dose Tracking System (DTS) from Canon, RadimetricsTM from Bayer and RDM from MEDSQUARE). The same RDSRs generated during a cardiology clinical procedure (percutaneous coronary intervention-PCI) were used for comparison.
RESULTS: Validation performed using VERIDIC's protocols for skin dose calculation software showed that PSD calculated by SkinCare is within 17% and 16% accuracy compared to measurements using XR-RV3 Gafchromic films for fundamental irradiation setups and simplified clinical procedures, respectively. Good visual agreement between dose maps generated by SkinCare and DTS, RadimetricsTM and RDM was obtained.
CONCLUSIONS: SkinCare is proved to be very convenient solution that can be used for monitoring delivered dose following interventional procedures.
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  DICOM RDSR; SkinCare application; XR-RV3 GafChromic films; interventional procedures; skin dose assessment; software

Mesh:

Year:  2021        PMID: 33440056      PMCID: PMC7882120          DOI: 10.1002/acm2.13167

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  25 in total

1.  Accuracy and calibration of integrated radiation output indicators in diagnostic radiology: A report of the AAPM Imaging Physics Committee Task Group 190.

Authors:  Pei-Jan P Lin; Beth A Schueler; Stephen Balter; Keith J Strauss; Kevin A Wunderle; M Terry LaFrance; Don-Soo Kim; Richard H Behrman; S Jeff Shepard; Ishtiaq H Bercha
Journal:  Med Phys       Date:  2015-12       Impact factor: 4.071

2.  [Medical electrical equipment - part 2-43: particular requirements for the basic safety and essential performance of X-ray equipment for interventional procedures].

Authors:  Hiroyuki Miyake Et Al
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2011

3.  Measurement of maximum skin dose in interventional radiology and cardiology and challenges in the set-up of European alert thresholds.

Authors:  J Farah; A Trianni; E Carinou; O Ciraj-Bjelac; I Clairand; J Dabin; C De Angelis; J Domienik; H Jarvinen; R Kopec; M Majer; F Malchair; A Negri; L Novák; T Siiskonen; F Vanhavere; Ž Knežević
Journal:  Radiat Prot Dosimetry       Date:  2014-10-14       Impact factor: 0.972

4.  Benchmarking the DACS-integrated Radiation Dose Monitor® skin dose mapping software using XR-RV3 Gafchromic® films.

Authors:  Bouchra Habib Geryes; Lama Hadid-Beurrier; Marie-Joséphine Waryn; Antonella Jean-Pierre; Jad Farah
Journal:  Med Phys       Date:  2018-08-31       Impact factor: 4.071

5.  Validation of a dose tracking software for skin dose map calculation in interventional radiology.

Authors:  Paola Enrica Colombo; Federica Rottoli; Marco Felisi; Cristina De Mattia; Stefano Riga; Marina Sutto; Colby Dillion; Steve Massey; Alberto Torresin
Journal:  Phys Med       Date:  2020-04-03       Impact factor: 2.685

6.  Experimental evaluation of a radiation dose management system-integrated 3D skin dose map by comparison with XR-RV3 Gafchromic® films.

Authors:  Joël Greffier; Nicolas Grussenmeyer-Mary; Ahmed Larbi; Jean Goupil; Guillaume Cayla; Bertrand Ledermann; Jean Paul Beregi; Julien Frandon
Journal:  Phys Med       Date:  2019-09-27       Impact factor: 2.685

7.  Updates in the real-time Dose Tracking System (DTS) to improve the accuracy in calculating the radiation dose to the patients skin during fluoroscopic procedures.

Authors:  Vijay K Rana; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2013-03-06

8.  An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan.

Authors:  David Lewis; Andre Micke; Xiang Yu; Maria F Chan
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.506

9.  Effect of fluoroscopic X-ray beam spectrum on air-kerma measurement accuracy: implications for establishing correction coefficients on interventional fluoroscopes with KAP meters.

Authors:  Kevin A Wunderle; Joseph T Rakowski; Frank F Dong
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

10.  Analysis of a multicentre cloud-based CT dosimetric database: preliminary results.

Authors:  Francesca Calderoni; Federica Campanaro; Paola Enrica Colombo; Mauro Campoleoni; Cristina De Mattia; Federica Rottoli; Giannicola Galetta; Fabio Zucconi; Andrea Pola; Andrea Righini; Fabio Triulzi; Angelo Vanzulli; Alberto Torresin
Journal:  Eur Radiol Exp       Date:  2019-07-16
View more
  1 in total

1.  Statement of the Italian Association of Medical Physics (AIFM) task group on radiation dose monitoring systems.

Authors:  Francesco Ria; Loredana D'Ercole; Daniela Origgi; Nicoletta Paruccini; Luisa Pierotti; Osvaldo Rampado; Veronica Rossetti; Sabina Strocchi; Alberto Torresin
Journal:  Insights Imaging       Date:  2022-02-05
  1 in total