Validation of a dose tracking software for skin dose map calculation in interventional radiology.

Paola Enrica Colombo1, Federica Rottoli2, Marco Felisi2, Cristina De Mattia2, Stefano Riga2, Marina Sutto2, Colby Dillion3, Steve Massey3, Alberto Torresin2.   

Abstract

PURPOSE: Validate the skin dose software within the radiation dose index monitoring system NEXO[DOSE]® (Bracco Injeneering S.A., Lausanne, Switzerland). It provides the skin dose distribution in interventional radiology (IR) procedures.
METHODS: To determine the skin dose distribution and the Peak Skin Dose (PSD) in IR procedures, the software uses exposure and geometrical parameters taken from the radiation dose structured report and additional information specific to each angiographic system. To test the accuracy of the software, GafChromic® XR-RV3 films, wrapped under a cylindrical PMMA phantom, were irradiated with different setups. Calculations and films results are compared in terms of absolute dose and geometric accuracy, using two angiographic systems (Philips Integris Allura FD20, Siemens AXIOM-ArtisZeego).
RESULTS: Calculated and film measured PSD values agree with an average difference of 7% ± 5%. The discrepancies in dose evaluation increase up to 33% in lower dose regions, because the algorithm does not consider the out-of-field scatter contribution of the neighboring fields, which is more significant in these areas. Regarding the geometric accuracy, the differences between the simulated dose spatial distributions and the measured ones are<3 mm (4%) in simple tests and 5 mm (5%) in setups closer to clinical practice. Moreover, similar results are obtained for the two studied angiographic system vendors.
CONCLUSIONS: NEXO[DOSE]® provides an accurate skin dose distribution and PSD estimate. It will allow faster and more accurate monitoring of patient follow-up in the future.
Copyright © 2020 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Interventional Radiology; Peak Skin Dose; Radiation Dose Structured Report; Radiochromic film; Skin dose map

Mesh:

Year:  2020        PMID: 32251850     DOI: 10.1016/j.ejmp.2020.03.008

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  3 in total

1.  Vendor-independent skin dose mapping application for interventional radiology and cardiology.

Authors:  Marko Krajinović; Nikola Kržanović; Olivera Ciraj-Bjelac
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

2.  Statement of the Italian Association of Medical Physics (AIFM) task group on radiation dose monitoring systems.

Authors:  Francesco Ria; Loredana D'Ercole; Daniela Origgi; Nicoletta Paruccini; Luisa Pierotti; Osvaldo Rampado; Veronica Rossetti; Sabina Strocchi; Alberto Torresin
Journal:  Insights Imaging       Date:  2022-02-05

Review 3.  Notifications and alerts in patient dose values for computed tomography and fluoroscopy-guided interventional procedures.

Authors:  Eliseo Vano; Reinhard Loose; Guy Frija; Graciano Paulo; Efstathios Efstathopoulos; Claudio Granata; Riccardo Corridori; Alberto Torresin; Jonas S Andersson; Virginia Tsapaki; Josefin Ammon; Christoph Hoeschen
Journal:  Eur Radiol       Date:  2022-03-16       Impact factor: 7.034

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.