Literature DB >> 24817801

Updates in the real-time Dose Tracking System (DTS) to improve the accuracy in calculating the radiation dose to the patients skin during fluoroscopic procedures.

Vijay K Rana1, Stephen Rudin2, Daniel R Bednarek2.   

Abstract

We have developed a dose-tracking system (DTS) to manage the risk of deterministic skin effects to the patient during fluoroscopic image-guided interventional cardiac procedures. The DTS calculates the radiation dose to the patient's skin in real-time by acquiring exposure parameters and imaging-system geometry from the digital bus on a Toshiba C-arm unit and displays the cumulative dose values as a color map on a 3D graphic of the patient for immediate feedback to the interventionalist. Several recent updates have been made to the software to improve its function and performance. Whereas the older system needed manual input of pulse rate for dose-rate calculation and used the CPU clock with its potential latency to monitor exposure duration, each x-ray pulse is now individually processed to determine the skin-dose increment and to automatically measure the pulse rate. We also added a correction for the table pad which was found to reduce the beam intensity to the patient for under-table projections by an additional 5-12% over that of the table alone at 80 kVp for the x-ray filters on the Toshiba system. Furthermore, mismatch between the DTS graphic and the patient skin can result in inaccuracies in dose calculation because of inaccurate inverse-square-distance calculation. Therefore, a means for quantitative adjustment of the patient-graphic-model position and a parameterized patient-graphic library have been developed to allow the graphic to more closely match the patient. These changes provide more accurate estimation of the skin-dose which is critical for managing patient radiation risk.

Entities:  

Keywords:  Interventional fluoroscopic procedures; dose mapping; dose tracking system; fluoroscopy; skin dose

Year:  2013        PMID: 24817801      PMCID: PMC4013095          DOI: 10.1117/12.2007706

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  6 in total

Review 1.  Skin injuries from fluoroscopically guided procedures: part 2, review of 73 cases and recommendations for minimizing dose delivered to patient.

Authors:  T R Koenig; F A Mettler; L K Wagner
Journal:  AJR Am J Roentgenol       Date:  2001-07       Impact factor: 3.959

Review 2.  Skin injuries from fluoroscopically guided procedures: part 1, characteristics of radiation injury.

Authors:  T R Koenig; D Wolff; F A Mettler; L K Wagner
Journal:  AJR Am J Roentgenol       Date:  2001-07       Impact factor: 3.959

3.  Verification of the performance accuracy of a real-time skin-dose tracking system for interventional fluoroscopic procedures.

Authors:  Daniel R Bednarek; Jeffery Barbarits; Vijay K Rana; Srikanta P Nagaraja; Madhur S Josan; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2011-02-13

Review 4.  Severe skin reactions from interventional fluoroscopy: case report and review of the literature.

Authors:  L K Wagner; M D McNeese; M V Marx; E L Siegel
Journal:  Radiology       Date:  1999-12       Impact factor: 11.105

5.  Radiation-induced skin injuries from fluoroscopy.

Authors:  T B Shope
Journal:  Radiographics       Date:  1996-09       Impact factor: 5.333

Review 6.  Potential biological effects following high X-ray dose interventional procedures.

Authors:  L K Wagner; P J Eifel; R A Geise
Journal:  J Vasc Interv Radiol       Date:  1994 Jan-Feb       Impact factor: 3.464

  6 in total
  5 in total

1.  A new method of real-time skin dose visualization. Clinical evaluation of fluoroscopically guided interventions.

Authors:  Fazel Boujan; Nicolas Clauss; Emilie Santos; Sjirk Boon; Gerard Schouten; Luc Mertz; Jean-Louis Dietemann
Journal:  Neuroradiology       Date:  2014-08-17       Impact factor: 2.804

2.  A tracking system to calculate patient skin dose in real-time during neurointerventional procedures using a biplane x-ray imaging system.

Authors:  V K Rana; S Rudin; D R Bednarek
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

3.  A system to track skin dose for neuro-interventional cone-beam computed tomography (CBCT).

Authors:  Sarath Vijayan; Zhenyu Xiong; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-29

4.  Incorporating Corrections for the Head-Holder and Compensation Filter when Calculating Skin Dose during Fluoroscopically-Guided Interventions.

Authors:  Sarath Vijayan; Vijay K Rana; Stephen Rudin; Daniel R Bednarek
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-03-18

5.  Vendor-independent skin dose mapping application for interventional radiology and cardiology.

Authors:  Marko Krajinović; Nikola Kržanović; Olivera Ciraj-Bjelac
Journal:  J Appl Clin Med Phys       Date:  2021-01-13       Impact factor: 2.102

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.