Iva Krolo1, Boze Mihaljevic1, Aida Kasumovic1, Freja Bagatin1, Maja Malenica Ravlic1, Jelena Skunca Herman1.
Abstract
INTRODUCTION: Goldmann applanation tonometry (GAT) is named as a gold standard for intraocular pressure (IOP) measurement. AIM: To assess the accuracy of intraocular pressure (IOP) measurements using rebound tonometry over hydrogel and silicone hydrogel contact lenses (CLs) of different powers.Entities:
Keywords: Contact Lens; Hydrogel; Icare; Rebound Tonometry; Silicone Hydrogel
Year: 2020 PMID: 33417647 PMCID: PMC7780781 DOI: 10.5455/aim.2020.28.185-189
Source DB: PubMed Journal: Acta Inform Med ISSN: 0353-8109
Figure 1.The distribution of CL materials included in study (%)
Figure 2.IOP values with and without CLs compared between three dioptic powers (minus, plus and cylindical dioptric power [dcyl]).
Figure 3.Differences of measured IOP with and without CLs in groups of different dioptric power.
Figure 4.Bland-Altmann plot where influence of CCT on difference between two IOP measurements (∆IOP), with and without CLs, is presented. Increase in CCT correlates positively with increase of measurement error of RT.
Two groups of CLs that comparing had largest differences in water content and modulus.
| Lens | Material | Water content (%) | Modulus (MPa) | Dk/t | Diameter (mm) | Central thickness (mm) |
|---|---|---|---|---|---|---|
| CL 1 | Methafilcon A | 55 | 0.50 | 23.5 | 14.2 | 0.15 |
| CL 2 | Balafilcon A | 36 | 1.50 | 112 | 14.0 | 0.09 |
Influence of soft lens material (rigiditiy) on difference between two IOP measurements (with and without CLs). Larger deviations in IOP measurements over materials with bigger modulus and smaller water content (CL 2).
| n | mean | SD | T | P | ||
|---|---|---|---|---|---|---|
| Icare WITH | CL 1 | 15 | 19.1 | ±2.3 | 3.42 | 0.002 |
| CL 2 | 15 | 26.1 | ±7.6 | |||
| Icare WITHOUT | CL 1 | 15 | 20.5 | ±6.4 | 0.21 | 0.827 |