Comparison of the ICare rebound tonometer with the Goldmann tonometer in a normal population.

P Fernandes1, J A Díaz-Rey, A Queirós, J M Gonzalez-Meijome, J Jorge.   

Abstract

The aim of this study was to evaluate the accuracy of measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare) in comparison with the Goldmann applanation tonometer (AT). The left eyes of 46 university students were assessed with the two tonometers, with induction tonometry being performed first. The ICare was handled by an optometrist and the Goldmann tonometer by an ophthalmologist. In this study, statistically significant differences were found when comparing the ICare rebound tonometer with applanation tonometry (AT) (p < 0.05). The mean difference between the two tonometers was 1.34 +/- 2.03 mmHg (mean +/- S.D.) and the 95% limits of agreement were +/-3.98 mmHg. A frequency distribution of the differences demonstrated that in more than 80% of cases the IOP readings differed by <3 mmHg between the ICare and the AT. In the present population the ICare overestimates the IOP value by 1.34 mmHg on average when compared with Goldmann tonometer. Nevertheless, the ICare tonometer may be helpful as a screening tool when Goldmann applanation tonometry is not applicable or not recommended, as it is able to estimate IOP within a range of +/-3.00 mmHg in more than 80% of the population.

Mesh:

Year:  2005        PMID: 16101950     DOI: 10.1111/j.1475-1313.2005.00327.x

Source DB:  PubMed          Journal:  Ophthalmic Physiol Opt        ISSN: 0275-5408            Impact factor:   3.117


  33 in total

1.  Age variations in intraocular pressure in a cohort of healthy Austrian school children.

Authors:  W A Dusek; B K Pierscionek; J F McClelland
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

2.  Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties.

Authors:  J Moreno-Montañés; I Gosende; J Caire; M García-Granero
Journal:  Eye (Lond)       Date:  2010-10-01       Impact factor: 3.775

Review 3.  Rebound tonometry: new opportunities and limitations of non-invasive determination of intraocular pressure.

Authors:  A Cervino
Journal:  Br J Ophthalmol       Date:  2006-12       Impact factor: 4.638

4.  Evaluation of the impedance tonometers TGDc-01 and iCare according to the international ocular tonometer standards ISO 8612.

Authors:  Peter Christian Ruokonen; Thomas Schwenteck; Jörg Draeger
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-02-21       Impact factor: 3.117

5.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

6.  The influence of refractive errors on IOP measurement by rebound tonometry (ICare) and Goldmann applanation tonometry.

Authors:  Teresio Avitabile; Antonio Longo; Daniela Rocca; Roberta Amato; Caterina Gagliano; Marine Castaing
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-29       Impact factor: 3.117

7.  Intraocular pressure measurement over soft contact lens by rebound tonometer: a comparative study.

Authors:  Senay Asik Nacaroglu; Emine Seker Un; Mehmet Giray Ersoz; Yelda Tasci
Journal:  Int J Ophthalmol       Date:  2015-06-18       Impact factor: 1.779

8.  Comparison of ICare and IOPen vs Goldmann applanation tonometry according to international standards 8612 in glaucoma patients.

Authors:  Milena Pahlitzsch; Jeanette Brünner; Johannes Gonnermann; Anna-Karina B Maier; Necip Torun; Eckart Bertelmann; Matthias Kj Klamann
Journal:  Int J Ophthalmol       Date:  2016-11-18       Impact factor: 1.779

Review 9.  Tonometers-which one should I use?

Authors:  Kanza Aziz; David S Friedman
Journal:  Eye (Lond)       Date:  2018-02-19       Impact factor: 3.775

10.  Comparison of IOPen rebound tonometer with Goldmann applanation tonometer at different IOP levels.

Authors:  Fereydoun Farrahi; Farideh Sharifipour; Mohammad Malekahmadi; Bahman Cheraghian
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.