Goldmann applanation tonometry over daily disposable contact lens: accuracy and safety of procedure.

Fabrizio Zeri1, Luigi Lupelli, Paolo Formichella, Carlo Masci, Robert Fletcher.   

Abstract

PURPOSE: To study accuracy and safety, related to sensation (discomfort) and trauma, when using Goldmann applanation tonometry (GAT) on eyes wearing daily disposable soft contact lenses.
METHODS: The intra-ocular pressure (IOP) of 136 normal eyes of 68 subjects was measured by Goldmann tonometer. Measurements were made in one eye with a contact lens (hilafilcon A) without anaesthetic drops and then without the contact lens using one drop of 0.4% oxybuprocaine hydrochloride. Each contact lens used was identical as to back optic zone, back vertex power. Standard Goldmann procedure only was used for the fellow eye of each subject. Subjective sensation (discomfort) responses to both procedures were studied in a subgroup (66 eyes) using a scale of discomfort, from 1 (no sensation) to 5 (highest sensation). Epithelial staining after tonometry was evaluated for this subgroup.
RESULTS: No significant differences were found for the IOP with and without contact lens (t<1; p=0.63) for the IOP range studied. There was a good correlation between the two procedures (r=0.81; p<0.05). Lowest sensation was found with tonometry on the anaesthetized cornea; this condition was significantly different from other conditions (p<0.005). No difference was found among the other conditions (contact lens insertion, tonometry on contact lens and application of topical anaesthetic). Corneal epithelial staining following the standard tonometry procedure was significantly higher than following the procedure with a contact lens (p<0.00005).
CONCLUSIONS: Measurement of IOP by GAT over a daily disposable soft contact lens is accurate, compared to the standard procedure and within the IOP's normal range studied here. Also using a contact lens results in less trauma whilst discomfort is similar.

Entities:  

Mesh:

Year:  2007        PMID: 17693127     DOI: 10.1016/j.clae.2007.06.001

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  6 in total

1.  The influence of soft contact lenses on the intraocular pressure measurement.

Authors:  P G Firat; C Cankaya; S Doganay; M Cavdar; S Duman; E Ozsoy; B Koc
Journal:  Eye (Lond)       Date:  2011-11-11       Impact factor: 3.775

2.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

3.  Study of the Efficacy and Safety of Contact Lens Used in Trabeculectomy.

Authors:  Bin Li; Miaomiao Zhang; Zhen Yang
Journal:  J Ophthalmol       Date:  2019-07-01       Impact factor: 1.909

4.  Rebound Tonometry over Soft Contact Lenses.

Authors:  Iva Krolo; Boze Mihaljevic; Aida Kasumovic; Freja Bagatin; Maja Malenica Ravlic; Jelena Skunca Herman
Journal:  Acta Inform Med       Date:  2020-09

5.  Intraocular pressure readings obtained through soft contact lenses using four types of tonometer.

Authors:  Joji Takenaka; Eriko Kunihara; Ulfah Rimayanti; Junko Tanaka; Makoto Kaneko; Yoshiaki Kiuchi
Journal:  Clin Ophthalmol       Date:  2015-10-05

6.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.