Applanation tonometry in silicone hydrogel contact lens wearers.

R J Allen1, A Dev Borman, G M Saleh.   

Abstract

INTRODUCTION: Previous studies have investigated intraocular pressure (IOP) measurements through conventional soft (hydrogel) therapeutic contact lenses, and have found that an accurate IOP can be recorded in normal eyes, and in eyes with abnormal anterior segments. The IOP measurement through soft contact lenses may be affected by the water content and centre thickness of the lens. Silicone hydrogel contact lenses are now being used as therapeutic contact lenses due to their high oxygen permeability. The purpose of this study is to investigate if IOP can be accurately measured in a subject wearing a silicone hydrogel contact lens.
METHODS: In a cohort study, the IOP was measured with a Goldmann applanation tonometer without a contact lens and then repeated with a hydrogel contact lens in situ.
RESULTS: The IOP of 20 eyes of 10 volunteers with no ocular pathology was measured. The mean difference (+/-S.D.) found between IOP measurement with (mean 15.55+/-1.70 mmHg) and without (mean 16.05+/-1.90 mmHg) contact lens was found to be -0.5+/-0.89 mmHg. Statistical analysis was performed which revealed a correlation coefficient of 0.89. No significant statistical difference was found between the two groups with paired t-test (p=0.19).
CONCLUSION: Accurate measurement of IOP by applanation tonometry can be achieved through a silicone hydrogel contact lens.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17382579     DOI: 10.1016/j.clae.2007.02.008

Source DB:  PubMed          Journal:  Cont Lens Anterior Eye        ISSN: 1367-0484            Impact factor:   3.077


  7 in total

1.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

2.  Comparison of intraocular pressure measurement with Scheimpflug-based noncontact tonometer with and without hydrogel contact lenses.

Authors:  Mukesh Kumar; Rohit Shetty; Chaitra Jayadev; Debarun Dutta; Maneck D Nicolsan; Sriharsha Nagaraj; Rajesh S Kumar
Journal:  Indian J Ophthalmol       Date:  2015-04       Impact factor: 1.848

3.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

4.  Study of the Efficacy and Safety of Contact Lens Used in Trabeculectomy.

Authors:  Bin Li; Miaomiao Zhang; Zhen Yang
Journal:  J Ophthalmol       Date:  2019-07-01       Impact factor: 1.909

5.  Rebound Tonometry over Soft Contact Lenses.

Authors:  Iva Krolo; Boze Mihaljevic; Aida Kasumovic; Freja Bagatin; Maja Malenica Ravlic; Jelena Skunca Herman
Journal:  Acta Inform Med       Date:  2020-09

6.  Intraocular pressure readings obtained through soft contact lenses using four types of tonometer.

Authors:  Joji Takenaka; Eriko Kunihara; Ulfah Rimayanti; Junko Tanaka; Makoto Kaneko; Yoshiaki Kiuchi
Journal:  Clin Ophthalmol       Date:  2015-10-05

7.  The measurement of intraocular pressure over positive soft contact lenses by rebound tonometry.

Authors:  Fabrizio Zeri; Mario De Cusatis; Luigi Lupelli; Peter Graham Swann
Journal:  J Optom       Date:  2015-10-21
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.