Literature DB >> 33417507

Diagnostic image quality of hand-held and wall-mounted X-ray devices in bitewing radiography: a non-inferiority clinical trial.

Reinier Cornelis Hoogeveen1, Siham Ouchene1, Wer Berkhout1.   

Abstract

OBJECTIVES: The present clinical trial was intended to clarify whether subjective assessments of diagnostic X-ray image quality achieved via hand-held (HH) Nomad Pro 2 (KaVo Kerr, Brea, CA, USA) X-ray device is non-inferior that of the wall-mounted (WM) KaVo Focus (KaVo Dental, Bieberich, Germany).
METHODS: A prospective, cross-over, and in vivo non-inferiority clinical trial was conducted to compare these two diagnostic modalities. Based on sampling calculations, 205 patients were selected for study, generating 410 paired bitewing radiographs in randomized sequence. The films were assessed independently, engaging three observers blinded to modality for random, side-by-side-comparisons. Diagnostic image quality was rated as follows: no preference, HH preference, or WM preference. Observer judgements were combined accordingly to reach a majority.
RESULTS: Collective observer ratings indicated no preference for diagnostic image quality in 63.9% of cases, with WM preference at 16.6% and HH preference at 19.5%. The difference in HH and WM preferences (19.5%-16.6% = 2.9%) was within the expected 95% confidence interval. Majority agreement was reached in 82.7%.
CONCLUSIONS: Subjectively assessed diagnostic image quality in bitewing radiographs acquired by HH and WM devices did not differ significantly. The hand-held device is thus non-inferior to the WM in this regard. Our data set of paired bitewing radiographs may subsequently aid in future research.

Entities:  

Keywords:  Dental equipment; Dental radiography; Diagnostic equipment; Diagnostic imaging

Mesh:

Year:  2021        PMID: 33417507      PMCID: PMC8231675          DOI: 10.1259/dmfr.20200471

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   3.525


  17 in total

1.  Subjective image quality of solid-state and photostimulable phosphor systems for digital intra-oral radiography.

Authors:  E Borg; A Attaelmanan; H G Gröndahl
Journal:  Dentomaxillofac Radiol       Date:  2000-03       Impact factor: 2.419

2.  The dynamic range of digital radiographic systems: dose reduction or risk of overexposure?

Authors:  W E R Berkhout; D A Beuger; G C H Sanderink; P F van der Stelt
Journal:  Dentomaxillofac Radiol       Date:  2004-01       Impact factor: 2.419

3.  Dental staff doses with handheld dental intraoral x-ray units.

Authors:  Joel E Gray; Edgar D Bailey; John B Ludlow
Journal:  Health Phys       Date:  2012-02       Impact factor: 1.316

4.  Maintaining radiation exposures as low as reasonably achievable (ALARA) for dental personnel operating portable hand-held x-ray equipment.

Authors:  Thomas J McGiff; Robert A Danforth; Edward E Herschaft
Journal:  Health Phys       Date:  2012-08       Impact factor: 1.316

5.  Forensic oral imaging quality of hand-held dental X-ray devices: comparison of two image receptors and two devices.

Authors:  P Pittayapat; P Thevissen; S Fieuws; R Jacobs; G Willems
Journal:  Forensic Sci Int       Date:  2009-11-12       Impact factor: 2.395

6.  Estimated operator exposure for hand holding portable X-ray units during imaging of the equine distal extremity.

Authors:  Reid Tyson; Douglas C Smiley; Robert S Pleasant; Gregory B Daniel
Journal:  Vet Radiol Ultrasound       Date:  2010-10-07       Impact factor: 1.363

7.  Comparison of the performance of intraoral X-ray sensors using objective image quality assessment.

Authors:  Kristina Hellén-Halme; Curt Johansson; Mats Nilsson
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2016-02-13

8.  Using hand-held dental x-ray devices: ensuring safety for patients and operators.

Authors:  Mina Mahdian; Andrew J Pakchoian; Didem Dagdeviren; Adel Alzahrani; Elnaz Jalali; Aditya Tadinada; Alan Lurie
Journal:  J Am Dent Assoc       Date:  2014-11       Impact factor: 3.634

9.  Justification and good practice in using handheld portable dental X-ray equipment: a position paper prepared by the European Academy of DentoMaxilloFacial Radiology (EADMFR).

Authors:  W E R Berkhout; A Suomalainen; D Brüllmann; R Jacobs; K Horner; H C Stamatakis
Journal:  Dentomaxillofac Radiol       Date:  2015-02-24       Impact factor: 2.419

10.  A clinical trial of the Nomad portable X-ray unit.

Authors:  Sharon L Brooks; Wallin E McMinn; Eriha Benavides
Journal:  J Mich Dent Assoc       Date:  2009-02
View more