Subjective image quality of solid-state and photostimulable phosphor systems for digital intra-oral radiography.

E Borg1, A Attaelmanan, H G Gröndahl.   

Abstract

OBJECTIVES: To compare subjectively the image quality of intra-oral radiographs from six digital systems.
METHODS: Two generations of two different solid-state detectors; Visualix-1 and -2 (Gendex Dental Systems, Milan, Italy), Computed Dental Radiography (CDR) and CDR Active Pixel Sensor (APS) (Schick Technologies Inc., Long Island City, NY, USA), and two photostimulable phosphor (PSP) systems; Digora (Soredex, Orion Corporation, Helsinki, Finland) and DenOptix (Gendex Dental Systems, Milan, Italy) were compared. Tooth-containing specimens from different areas of dried mandibles were radiographed at exposures between 91-9400 microGy. Images were transferred to a personal computer, displayed in random order and evaluated in their original form and after applying a histogram equalisation algorithm. Eight observers graded subjective image quality using a 5-point scale.
RESULTS: Both CDR systems scored highest for image quality but within the narrowest exposure range. The Visualix images received the lowest scores. The PSP systems produced acceptable image quality at both lower and higher exposures than the solid-state systems. Enhanced images were generally considered to be inferior to the original images, except for those produced by the four solid-state systems at very low exposures.
CONCLUSIONS: (i) the PSP systems provided a clinically acceptable image quality over a wide exposure range; (ii) the CDR systems had the best image quality but over the narrowest exposure ranges; (iii) the Visualix systems had the lowest image quality; and (iv) histogram equalisation did not generally improve image quality.

Entities:  

Mesh:

Year:  2000        PMID: 10808218     DOI: 10.1038/sj/dmfr/4600501

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  15 in total

1.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 1: the influence of X-ray generator on periodontal measurements using conventional and digital receptors.

Authors:  Bart Vandenberghe; Livia Corpas; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-05       Impact factor: 3.573

2.  A comprehensive in vitro study of image accuracy and quality for periodontal diagnosis. Part 2: the influence of intra-oral image receptor on periodontal measurements.

Authors:  Bart Vandenberghe; Hilde Bosmans; Jie Yang; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2010-05-12       Impact factor: 3.573

3.  The effect of delayed scanning of storage phosphor plates on occlusal caries detection.

Authors:  E Sogur; B G Baksi; A Mert
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

4.  Comparison of radiographic measurements obtained with conventional and indirect digital imaging during endodontic treatment.

Authors:  Ligia Buloto Schmitd; Tatiana de Castro Lima; Luiz Eduardo Montenegro Chinellato; Clóvis Monteiro Bramante; Roberto Brandão Garcia; Ivaldo Gomes de Moraes; Norberti Bernardineli
Journal:  J Appl Oral Sci       Date:  2008 Mar-Apr       Impact factor: 2.698

5.  Measurement accuracy and perceived quality of imaging systems for the evaluation of periodontal structures.

Authors:  B Güniz Baksi
Journal:  Odontology       Date:  2008-07-27       Impact factor: 2.634

6.  Effect of alternative photostimulable phosphor plates erasing times on subjective digital image quality.

Authors:  D P Melo; A Dos Anjos Pontual; S M de Almeida; P S F Campos; M C Alves; G M Tosoni
Journal:  Dentomaxillofac Radiol       Date:  2010-01       Impact factor: 2.419

7.  The quality of radiographs accompanying endodontic referrals to a health authority clinic.

Authors:  B S Chong; J Miller; S Sidhu
Journal:  Br Dent J       Date:  2015-07-24       Impact factor: 1.626

8.  The effect of anatomical noise on perception of low contrast in intra-oral radiographs: an in vitro study.

Authors:  Lars Olsson; Mats Nilsson; Björn Svenson; Kristina Hellén-Halme
Journal:  Dentomaxillofac Radiol       Date:  2016-02-19       Impact factor: 2.419

9.  Influence of image filters on the reproducibility measurements of alveolar bone loss.

Authors:  Wivian Eiras Garcia Wong Alves; Evelise Ono; Jefferson Luis Oshiro Tanaka; Titular Edmundo Medici Filho; Luiz César de Moraes; Mari Eli Leonelli de Moraes; Julio Cezar de Melo Castilho
Journal:  J Appl Oral Sci       Date:  2006-12       Impact factor: 2.698

10.  In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX.

Authors:  Fernando Accorsi Orosco; Norberti Bernardineli; Roberto Brandão Garcia; Clovis Monteiro Bramante; Marco Antonio Húngaro Duarte; Ivaldo Gomes de Moraes
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.