Ultrasound Cardiac Output Monitor (USCOM™) Measurements Prove Unreliable Compared to Cardiac Magnetic Resonance Imaging in Adolescents with Cardiac Disease.

Avichai Weissbach1,2, Amichay Rotstein3,4, Yaniv Lakovsky5,4, Eytan Kaplan6,4, Gili Kadmon6,4, Einat Birk3,4, Elhanan Nahum6,4.   

Abstract

The purpose of this stuy is to prospectively assess the reliability of the ultrasound cardiac output monitor (USCOM™) for measuring stroke volume index and predicting left ventricular outflow tract diameter in adolescents with heart disease. Sixty consecutive adolescents with heart disease attending a tertiary medical center underwent USCOM™ assessment immediately after cardiac magnetic resonance imaging. USCOM™ measured stroke volume index and predicted left ventricular outflow tract diameter were compared to cardiac magnetic resonance imaging-derived values using Bland-Altman analysis. Ten patients with an abnormal left ventricular outflow tract were excluded from the analysis. An adequate USCOM™ signal was obtained in 49/50 patients. Mean stroke volume index was 46.1 ml/m2 by the USCOM™ (range 22-66.9 ml/m2) and 42.9 ml/m2 by cardiac magnetic resonance imaging (range 24.7-59.9 ml/m2). The bias (mean difference) was 3.2 ml/m2; precision (± 2SD of differences), 17 ml/m2; and mean percentage error, 38%. The mean (± 2SD) left ventricular outflow tract diameter was 0.445 ± 0.536 cm smaller by the USCOM™ algorithm prediction than by cardiac magnetic resonance imaging. Attempted adjustment of USCOM™ stroke volume index using cardiac magnetic resonance imaging left ventricular outflow tract diameter failed to improve agreement between the two modalities (bias 28.4 ml/m2, precision 44.1 ml/m2, percentage error 77.3%). Our study raises concerns regarding the reliability of USCOM™ for stroke volume index measurement in adolescents with cardiac disease, which did not improve even after adjusting for its inaccurate left ventricular outflow tract diameter prediction.

Entities:  

Keywords:  Cardiology; Congenital heart disease; Hemodynamic monitoring; Pediatric intensive care; Pediatrics

Mesh:

Year:  2021        PMID: 33394105     DOI: 10.1007/s00246-020-02531-8

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  13 in total

Review 1.  Cardiovascular magnetic resonance.

Authors:  D Pennell
Journal:  Heart       Date:  2001-05       Impact factor: 5.994

Review 2.  Clinical use of cardiovascular magnetic resonance.

Authors:  Gerald M Pohost; Lynne Hung; Mark Doyle
Journal:  Circulation       Date:  2003-08-12       Impact factor: 29.690

3.  Cardiac output estimation with a new Doppler device after off-pump coronary artery bypass surgery.

Authors:  Rajesh Chand; Yatin Mehta; Naresh Trehan
Journal:  J Cardiothorac Vasc Anesth       Date:  2006-01-18       Impact factor: 2.628

4.  Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit.

Authors:  H L Tan; M Pinder; R Parsons; B Roberts; P V van Heerden
Journal:  Br J Anaesth       Date:  2005-01-14       Impact factor: 9.166

5.  Non-invasive cardiac output determination by two-dimensional independent Doppler during and after cardiac surgery.

Authors:  Karsten Knobloch; Artur Lichtenberg; Michael Winterhalter; Dagmar Rossner; Max Pichlmaier; Rob Phillips
Journal:  Ann Thorac Surg       Date:  2005-10       Impact factor: 4.330

6.  Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs.

Authors:  Lester A Critchley; Zhi Y Peng; Benny S Fok; Anna Lee; Robert A Phillips
Journal:  Anesth Analg       Date:  2005-03       Impact factor: 5.108

7.  Cardiac index measurements by transcutaneous Doppler ultrasound and transthoracic echocardiography in adult and pediatric emergency patients.

Authors:  H Bryant Nguyen; Daryl P Banta; Gail Stewart; Tommy Kim; Ramesh Bansal; James Anholm; William A Wittlake; Stephen W Corbett
Journal:  J Clin Monit Comput       Date:  2010-06-20       Impact factor: 2.502

8.  Intra- and inter-observer reliability using a noninvasive ultrasound cardiac output monitor in healthy anesthetized children.

Authors:  Sonny Dhanani; Nick J Barrowman; Roxanne E Ward; Kimmo T Murto
Journal:  Paediatr Anaesth       Date:  2010-12-15       Impact factor: 2.556

9.  Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement.

Authors:  Walter Knirsch; Oliver Kretschmar; Maren Tomaske; Kathrina Stutz; Nicole Nagdyman; Christian Balmer; Achim Schmitz; Dominique Béttex; Felix Berger; Urs Bauersfeld; Markus Weiss
Journal:  Intensive Care Med       Date:  2008-02-23       Impact factor: 17.440

10.  Validation of an Ultrasound Cardiac Output Monitor as a Bedside Tool for Pediatric Patients.

Authors:  Fernando Beltramo; Jondavid Menteer; Asma Razavi; Robinder G Khemani; Jacqueline Szmuszkovicz; Christopher J L Newth; Patrick A Ross
Journal:  Pediatr Cardiol       Date:  2015-09-12       Impact factor: 1.655

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.