Literature DB >> 26364291

Validation of an Ultrasound Cardiac Output Monitor as a Bedside Tool for Pediatric Patients.

Fernando Beltramo1, Jondavid Menteer1,2, Asma Razavi1, Robinder G Khemani1,2, Jacqueline Szmuszkovicz1,2, Christopher J L Newth1,2, Patrick A Ross3,4.   

Abstract

The aim of our study was to determine the validity of cardiac output (CO) measurements taken with the ultrasonic cardiac output monitor (USCOM) by comparing to CO measured by pulmonary arterial catheter (PAC) thermodilution during cardiac catheterization. We enrolled thirty-one children (<18 years) undergoing cardiac catheterization in this double-blinded, prospective, observational study. The median CO measured by USCOM was 4.37 L/min (IQR 3.73, 5.60 L/min) compared to 4.28 L/min (IQR 3.52, 5.26 L/min) by PAC thermodilution. The bias (mean difference) between the two methods was 0.2 L/min, and the 95% limits of agreement were -1.2 to 1.6 L/min. The mean percentage error of CO between USCOM and PAC thermodilution was 11%. When excluding a sole outlier, the bias between the two measures decreased to 0.1 L/min (95% limits of agreement -0.6 to 0.9 L/min), and the percentage error was reduced to 8%. The median SVRI measured by USCOM was 22.0 Wood Units (IQR 17.0, 26.8 Wood Units) compared to 22.1 Wood Units (IQR 17.6, 27.4 Wood Units) by PAC thermodilution. Bias (mean difference) between the two methods was -0.6 Wood Units, and the 95% limits of agreement were -8.2 to 6.9 Wood Units. We found that the estimation of CO and by extension SVRI with USCOM is reliable against pulmonary artery catheter thermodilution in children with normal cardiac anatomy. Given the noninvasive nature of USCOM, speed of measurement, and relative ease of use, it may be useful as a bedside tool for pediatric patients.

Entities:  

Keywords:  Cardiac catheterization; Cardiac output; Child; Pediatrics; Thermodilution; Ultrasonics

Mesh:

Year:  2015        PMID: 26364291     DOI: 10.1007/s00246-015-1261-y

Source DB:  PubMed          Journal:  Pediatr Cardiol        ISSN: 0172-0643            Impact factor:   1.655


  27 in total

Review 1.  Minimally invasive hemodynamic monitoring for the intensivist: current and emerging technology.

Authors:  John C Chaney; Stephen Derdak
Journal:  Crit Care Med       Date:  2002-10       Impact factor: 7.598

2.  Cardiac output estimation with a new Doppler device after off-pump coronary artery bypass surgery.

Authors:  Rajesh Chand; Yatin Mehta; Naresh Trehan
Journal:  J Cardiothorac Vasc Anesth       Date:  2006-01-18       Impact factor: 2.628

Review 3.  Cardiac output monitoring.

Authors:  Eric E C de Waal; Frank Wappler; Wolfgang F Buhre
Journal:  Curr Opin Anaesthesiol       Date:  2009-02       Impact factor: 2.706

4.  Clinical assessment of cardiac performance in infants and children following cardiac surgery.

Authors:  Jonathan R Egan; Marino Festa; Andrew D Cole; Graham R Nunn; Jonathan Gillis; David S Winlaw
Journal:  Intensive Care Med       Date:  2005-02-15       Impact factor: 17.440

Review 5.  Minimally invasive cardiac output monitoring.

Authors:  Patrick Morgan; Nawaf Al-Subaie; Andrew Rhodes
Journal:  Curr Opin Crit Care       Date:  2008-06       Impact factor: 3.687

6.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

7.  Clinicians' abilities to estimate cardiac index in ventilated children and infants.

Authors:  S M Tibby; M Hatherill; M J Marsh; I A Murdoch
Journal:  Arch Dis Child       Date:  1997-12       Impact factor: 3.791

8.  Assessment of portable continuous wave Doppler ultrasound (ultrasonic cardiac output monitor) for cardiac output measurements in neonates.

Authors:  Sascha Meyer; David Todd; Bruce Shadboldt
Journal:  J Paediatr Child Health       Date:  2009 Jul-Aug       Impact factor: 1.954

Review 9.  Monitoring cardiac function in intensive care.

Authors:  S M Tibby; I A Murdoch
Journal:  Arch Dis Child       Date:  2003-01       Impact factor: 3.791

10.  Urgent ultrasound guided hemodynamic assessments by a pediatric medical emergency team: a pilot study.

Authors:  David J Zorko; Karen Choong; Jonathan Gilleland; Barbara Agar; Shawn Baker; Cindy Brennan; Eleanor Pullenayegum
Journal:  PLoS One       Date:  2013-06-25       Impact factor: 3.240

View more
  10 in total

1.  Reliability of the Ultrasound Cardiac Output Monitor for Pediatric Patients.

Authors:  Joris Lemson; Shane M Tibby
Journal:  Pediatr Cardiol       Date:  2016-03-18       Impact factor: 1.655

2.  Ultrasound Cardiac Output Monitor (USCOM™) Measurements Prove Unreliable Compared to Cardiac Magnetic Resonance Imaging in Adolescents with Cardiac Disease.

Authors:  Avichai Weissbach; Amichay Rotstein; Yaniv Lakovsky; Eytan Kaplan; Gili Kadmon; Einat Birk; Elhanan Nahum
Journal:  Pediatr Cardiol       Date:  2021-01-04       Impact factor: 1.655

3.  Measurement of Cardiac Output Using an Ultrasonic Cardiac Output Monitor (USCOM) in Patients with Single-Ventricle Physiology.

Authors:  Wonshill Koh; Kristin A Schneider; Huaiyu Zang; Sarosh P Batlivala; Matthew P Monteleone; Alexis L Benscoter; Meghan M Chlebowski; Ilias D Iliopoulos; David S Cooper
Journal:  Pediatr Cardiol       Date:  2022-02-06       Impact factor: 1.838

4.  [Application of ultrasonic cardiac output monitor in evaluation of cardiac function in children with severe pneumonia].

Authors:  Juan Fan; Zhi-Ying Chen; Peng-Yuan Chen; Chang-Hui Chen
Journal:  Zhongguo Dang Dai Er Ke Za Zhi       Date:  2016-09

Review 5.  Noninvasive hemodynamic monitoring of septic shock in children.

Authors:  Emad Mohamed Fathi; Hassib Narchi; Fares Chedid
Journal:  World J Methodol       Date:  2018-06-28

6.  Comparison of three non-invasive hemodynamic monitoring methods in critically ill children.

Authors:  Chanapai Chaiyakulsil; Marut Chantra; Poomiporn Katanyuwong; Anant Khositseth; Nattachai Anantasit
Journal:  PLoS One       Date:  2018-06-18       Impact factor: 3.240

7.  Serum NT-Pro-BNP versus Noninvasive Bedside Inotropic Index in Paediatric Shock: A Contest of Myocardial Performance in Response to Fluid Loading.

Authors:  Antonius Hocky Pudjiadi; Tuty Rahayu; Stephanie Wijaya; Fatima Safira Alatas
Journal:  Crit Care Res Pract       Date:  2021-11-30

Review 8.  Hemodynamic monitoring and management of pediatric septic shock.

Authors:  En-Pei Lee; Han-Ping Wu; Oi-Wa Chan; Jainn-Jim Lin; Shao-Hsuan Hsia
Journal:  Biomed J       Date:  2021-10-12       Impact factor: 7.892

9.  Race predictors and hemodynamic alteration after an ultra-trail marathon race.

Authors:  Noppon Taksaudom; Natee Tongsiri; Amarit Potikul; Chawakorn Leampriboon; Apichat Tantraworasin; Anong Chaiyasri
Journal:  Open Access J Sports Med       Date:  2017-10-06

10.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.