Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.

Firdevs Çınar1, Yakup Üstün2.   

Abstract

OBJECTIVE: The aim of this study was to compare the accuracy of three electronic apex locators (EALs) (Propex Pixi, Mini Root ZX, Raypex 5) in determining working length (WL) under different environments (existence of blood-pulp/sodium hypochlorite in root canal space) using micro-computed tomography (micro-CT) measurements.
METHODS: Twenty-five single-rooted human teeth that were scheduled for extraction were selected for the study. Measurements were performed with the Propex Pixi, Mini Root ZX, Raypex 5 in vivo conditions in the presence of NaOCl irrigant solution or blood-pulp tissue. After that the teeth were extracted and scanned using micro-CT. The measurements of WL obtained with the different EALs in different environments were statistically compared. Significance was set at P<0.05.
RESULTS: There were no significant differences among the WL measurements performed with EALs or micro CT groups under different clinical enviroments. All EALs tested gave reliable results in respect to apical constriction.
CONCLUSION: The accuracy of the tested EALs is not affected by pulp tissues and blood or NaOCl.

Entities:  

Year:  2020        PMID: 33353910      PMCID: PMC7881377          DOI: 10.14744/eej.2020.30633

Source DB:  PubMed          Journal:  Eur Endod J        ISSN: 2548-0839


This study examined the accuracy of three electronic apex locators in determining the working length under different environments. The new generation EALs are not affected from NaOCl or blood/pulp tissue existence in the root canal.

INTRODUCTION

In endodontology, the working length (WL) is defined as the distance from a coronal reference point to the point at which canal preparation and root canal filling should terminate. However the end point of instrumentation is still debated (1). In the root canal anatomy, apical constriction (AC) is the narrowest diameter of the root canal system (2, 3). Several studies consider the AC of the root canal system as the end point of instrumentation or obturation processes (2, 3). However, the location of AC is difficult due to the anatomical variations that could vary depending on various factors (4-6). There are several methods for determination of the WL such as assessment of preoperative radiographs, tactile sensation and electronic apex locators (EALs) (7, 8). Electronic apex locators (EAL) provide reliable and accurate results in working length determinations (9). Propex Pixi (Dentsply Maillefer, Ballaigues, Switzerland) is a recently introduced multi-frequency EAL, Mini Root ZX (J Morita Corp., Tokyo, Japan) is another contemporary EAL that uses the ‘ratio method’. Raypex 5 (VDW, Munich, Germany) is a frequency-based EAL, and it measures the impedance at two different frequencies (0.4 kHz and 8 kHz), but uses only one frequency at each time interval and the measurements are based on mean square values of signals (10). There are several investigations about the accuracy of EALs in different environments (9, 11-14). However, the accuracy of EALs may be affected by some conditions such as periapical lesions, immature apex formation, apical foramen size, electrical conductivity of the intracanal fluid (such as, irrigants, pulp) (12, 14-17). The use of micro-computed tomography (micro-CT) gives opportunity to view detailed 3-dimensional (3D) analyses of extracted teeth’s internal anatomy with the high-resolution radiographic scannings of the specimen (18-21). With the aid of micro-CT, the AC can be determined accurately without destruction of the tooth structure (4). The AC can be determined by evaluating the canal cross-sectional area and not by locating the minor diameter in one longitudinal section (22). In one micro-CT study, ElAyouti et al. (4) determined the location and dimension of the AC in molar teeth. Piasecki et al. (5) investigated the accuracy of two apex locators in different anatomic variations using micro-CT. However, according to our knowledge there is no wrıtten data comparing the in vivo WL measurements of different EALs under different circumstances using micro-CT. The aim of this study was to compare the performances of different EALs in determining WL-AC under different circumstances (existence of blood, pulp/sodium hypochlorite in root canal space) with micro-CT measurements.

MATERIALS AND METHODS

Erciyes University Medical Sciences ethical board (Kayseri, Turkey) approved this investigation (Ethics Approval Number: 2016-459). All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The present investigation included 18 patients (11 male, 7 female) who were consulted to the Erciyes University Faculty of Dentistry, Department of Oral and Maxillofacial Surgery for extraction of their teeth with single canal and single root (16 incisors, 9 premolars). Inclusion criteria for patients were as following: Patients aged 20-60 years who do not have general health problems that would prevent endodontic therapy Teeth diagnosed for extraction due to orthodontic or periodontal treatment plannings Teeth without periapical pathology visible on radiography. Teeth with single root and canal with completely formed apices and teeth with no root resorption (internal or external) Teeth with vital pulp were included. Teeth that respond to cold test (Coltene/Whaledent Inc, Cuyahoga, OH) were included. Anesthesia (Ultracain DS forte® Sanofi-Aventis, Frankfurt Main, Germany) was performed, and a dental rubber dam was placed. If a flat reference point was not available, the cusp tip or incisal edge was flattened by using a diamond bur (Henry Schein®, Melville, NY, USA) in a high-speed handpiece under water cooling. Access cavity was opened by using a #4 carbide bur. Restrictive cervical bulk dentin was removed by using Gates-Glidden (VDW, Munich, Germany) drills with a low speed handpiece. Apical patency was checked with #8 K-File (Dentsply Maillefer, Ballaigues, Switzerland). Then, WL measurements were performed with EALs by an experienced endodontist. The WL measurements were performed under two different circumstances (in existence of blood and pulp tissue and in existence of NaOCl in the root canals).

WL measurements in presence of blood and pulp tissue in the root canals

A sterile paper point was inserted into the root canal and checked if it was wet by blood in order to confirm the existence of blood and pulp tissue in the root canal space. The measurements were recorded as WLBlood for each EAL. A #10 size K-file (Dentsply Maillefer) attached to the EAL device was then introduced into the root canal, and the lip hook of each EAL device was put onto the lip of the patient.

WL measurements in presence of 2.5 % sodium hypochlorite in the root canals

Pulp tissue was removed with a sterile barbed broach and root canal was irrigated with copious 2.5% NaOCl by a 27-gauge needle. The measurements were recorded as WLNaOCl for each EAL. A #10 size K-file attached to the EAL device was then introduced into the root canal, and the lip hook of each EAL device was put onto the lip of the patient.

Measurements with Mini Root ZX

The Mini Root ZX (J Morita Corp., Tokyo, Japan) was used in accordance with the manufacturer’s instructions. A size 10 K-file attached to EAL was placed into the root canal and advanced down the canal without force. A rubber stop on the file, which was in contact with the flat, horizontal surface of the crown, was fixed to the file with a flowable light-curing resin (Nova Compo HF Flow; Imicryl, Konya, Turkey) at the position when the flashing bars of the EAL reached the level that presented ‘00’ on its LCD screen.

Measurements with RayPex 5

The Raypex 5 (VDW, Munich, Germany) was used in accordance with the manufacturer’s instructions. A #10 K-file was attached to EAL and advanced within the root canal to the major foramen (red light) and retracted until three green and four yellow bars appeared. Then, rubber stop on the file, which was in contact with the flat, horizontal surface of the crown, was fixed to the file with a flowable light-curing resin.

Measurements with Propex Pixi

The Propex Pixi (Dentsply Maillefer, Ballaigues, Switzerland) was used in accordance with the manufacturer’s instructions. When the 0.0 signal was seen by the operator, the file was gently moved beyond the apex until the ‘‘beyond apex’’ signal appeared, at which point the file was quickly withdrawn and stopped at the 0.0 point. A rubber stop on the file, which was in contact with the flat, horizontal surface of the crown, was fixed to the file with a flowable light-curing resin. All measurements were considered to be valid if the reading/signal on screen remained stable for at least 5 seconds. The distances between the stop and the file tip was measured with digital calipers (Insize Mini digital caliper; Istanbul, Turkey) (at 0.01 mm accuracy) and recorded as the initial length of the root. The WL measurements were repeated 3 times with each EAL, and the mean value was recorded. Finally, 0.5 mm was subtracted from this measurement and recorded for each EAL. After the measurements completed with each EAL for two circumtances, the teeth were extracted in a gently manner without causing any fracture.

Micro-CT Scan

Each sample was scanned in the micro-CT SkyScan 1272 (Bruker microCT, Kontich, Belgium) at a 70 KV, 142 µA and 12 µm resolution. Images were reconstructed using NRecon V1.6.9.3 software (Bruker microCT), and analyzed using CTAn v1.11.10 software (Bruker microCT ). All scans were visualized by 1 calibrated operator. If the AF was deviated from the long axis of the root, the images were reoriented to show the exact inclination of the AF. The slices were moved coronally from AF to AC and the narrowest diameter was detected by measuring canal lumens in buccolingual and mesiodistal directions (Fig. 1).
Figure 1

. 3D model of the root canal is shown at (a) The apical constriction point of root canal is shown at (b) with “a” and also coronal reference point is shown with yellow line. The horizontal micro-CT section of apical constriction is shown with smaller figure

. 3D model of the root canal is shown at (a) The apical constriction point of root canal is shown at (b) with “a” and also coronal reference point is shown with yellow line. The horizontal micro-CT section of apical constriction is shown with smaller figure The length of the canal was acquired from the coronal reference point(at the most incisal or occlusal plane of teeth with a flat and constant reference point also the same reference point for all EALs in clinical applications) up to the most coronal slice showing the AC (WL) at 0.025 mm precision and recorded (Fig 1). Also distance from AC to major foramen was calculated and recorded. All measurements were completed through the center of the root canals by the same operator.

Statistical analyses

All statistical analyses were performed with Sigmastat 3.1 (SigmaStat; Systat, San Jose, CA). The normality test of Shapiro-Wilk was applied to the data (a=0.05). The sample failed the normality test (P<0.05); consequently, Friedman Repeated Measures Analysis of Variance on Ranks test was used (a=0.05).The EAL measurements were evaluated within the ±0.50-mm range from the reference landmarks.

RESULTS

According to the measurements mean distance from AC to major foramen was calculated as 0.416 mm. There were no significant differences among the WL measurements performed with EALs or micro CT groups (P>0.05) (Table 1, Fig. 2). Neither the presence of blood-pulp tissue nor 2.5 % NaOCl solution in the root canal did significantly affect the WL measurements.
TABLE 1

Mean and Standard Deviation and error (mm) obtained for distances to AC in absolute value according to the tested EALs under different circumtances

GroupsMeanSDSD error
Root ZX Blood0.3850.2990.0611
Root ZX NaOCl0.3590.2660.0543
Raypex Blood0.4290.2960.0605
Raypex NaOCl0.470.2740.056
Pixi Blood0.360.2590.0529
Pixi NaOCl0.4640.2970.0607

SD: Standard deviation

Figure 2

The horizontal line indicates the position of the AC found in Micro Ct measurements. Positive values (above the line) indicate measurements beyond the AC; negative values are short of the AC

Mean and Standard Deviation and error (mm) obtained for distances to AC in absolute value according to the tested EALs under different circumtances SD: Standard deviation The horizontal line indicates the position of the AC found in Micro Ct measurements. Positive values (above the line) indicate measurements beyond the AC; negative values are short of the AC

DISCUSSION

In the present study, accuracies of 3 different EALs used in different clinical conditions to determine AC was compared by micro-CT. In order to obtain a standardized reference point incisal surfaces of teeth were flattened. WL measurements were done from this flattened surface to AC as determined by micro-CT images or EALs. In order to avoid sequence bias, apex locators were used in a randomized order. Before in vivo measurements, all EALs were always fully charged and always used as described in the manufacturer’s manual. De Camargo et al. (23) found that preflaring of the root canal orifices give more accurate results in EALs. Therefore all the teeth were preflared before the EAL measurements. In previous studies, EAL readings with an error range of ±0.5 mm considered as accurate (5, 16, 24, 25). Therefore, in the present study, EAL readings with an error range of ±0.5 mm were accepted as accurate. In sample selection, only teeth with single root, straight and single canal and single foramen were used, because all WL measurements were done from the center point of root canal space from apical reference point (apical constriction) to coronal reference point. By this way, all WL measurements were done in a straight line through multiple reference points. Micro-CT is a nondestructive method for evaluation anatomical landmarks in tooth samples. Therefore, we preferred to use micro-CT analyzing tool for WL measurements. Previous studies determined the location of the apical constriction with longitudinal slices (2, 26), however, Schell et al. (22) showed that this method is unsuitable for correct determination of AC. With the aid of micro-CT technology, it is possible to determine the narrowest apical part of the entire root canal. The distance between AC-AF was investigated by some authors (2, 4). Kuttler (2) stated the distance from AC to AF as 0.52 mm to 0.66 mm. Elayouti et al. (4) showed the AC-AF distance as 0.2 mm. In the present study, we found the mean distance between AC-AF as 0.416 mm. However, in the present study we have evaluated 25 teeth (16 incisors, 9 premolars). Our results could be related to our sample size and different teeth types. Further investigations with larger sample sizes are needed. In several studies it was reported that endodontic irrigants could affect the accuracy of EALs (27, 28). However, some investigators found that EALs gave accurate results irrespective of the used irrigant (29, 30). Besides, in endodontic treatment process of vital teeth bleeding is inevitable. Since blood is an electrolyte, it may also influence the accuracy of EALs (14, 31). Saatchi et al.(14) evaluated in vivo the accuracy of the Raypex 5 and Root ZX EALs in the presence of blood in the root canal space. They found that presence of blood in the root canal space did not influence the accuracy of tested EALs. Our results were in accordance with the study of Saatchi et al. (14). According to our results, in existence of blood-pulp tissue in the root canal, there were no statistically significant differences among the groups (P>0.05). However, Propex Pixi and Root ZX mini showed more similar results when compared to micro-CT measurements, and Raypex 5 gave slightly different results than Propex Pixi and Root ZX mini (P>0.05).This could be related to their new generation technology. In several studies, EAL measurements performed in existence of NaOCl irrigation solution in the root canal gave inaccurate results (27, 28, 32, 33). In the present study, we found that NaOCl did not affect the accuracy of EALs. None of these studies evaluated the accuracies of EALs with micro-CT. Differences in results may be attributed to different methodologies. On the other side, Ebrahim et al. (31) found that Root ZX EAL gave accurate results in existence of NaOCl in the root canals. They related this result to NaOCl’s highly electro-conductivity and infiltration property into dentinal tubules which resulted in reduction of electrical impedance of the root canal wall. In addition, Gomes et al. (34) showed that NaOCl solution did not affect the in vivo accuracy of Raypex 5. According to the results, we also found that NaOCl did not affect the accuracy of EALs. This could be related to the high electro-conductivity and infiltration property of NaOCl.

CONCLUSION

Within the limitations of this study, it can be concluded that the accuracy of Raypex 5, Root ZX mini and Propex Pixi EALs are not affected by existence of blood-pulp tissue or NaOCl.
  34 in total

1.  Micro-Computed Tomography Analysis of the Root Canal Morphology of Palatal Roots of Maxillary First Molars.

Authors:  Marília Marceliano-Alves; Flávio Rodrigues Ferreira Alves; Daniel de Melo Mendes; José Claudio Provenzano
Journal:  J Endod       Date:  2015-11-26       Impact factor: 4.171

2.  Apical limit of root canal instrumentation and obturation, part 2. A histological study.

Authors:  D Ricucci; K Langeland
Journal:  Int Endod J       Date:  1998-11       Impact factor: 5.264

3.  Evaluation of the accuracy of three electronic apex locators using glass tubules.

Authors:  W Fan; B Fan; J L Gutmann; Z Bian; M W Fan
Journal:  Int Endod J       Date:  2006-02       Impact factor: 5.264

4.  Influence of preflaring on the accuracy of length determination with four electronic apex locators.

Authors:  Ericson Janolio de Camargo; Ronald Ordinola Zapata; Paulo Leal Medeiros; Clovis Monteiro Bramante; Norberti Bernardineli; Roberto Brandão Garcia; Ivaldo Gomes de Moraes; Marco Antônio Húngaro Duarte
Journal:  J Endod       Date:  2009-09       Impact factor: 4.171

5.  Apical constriction: location and dimensions in molars-a micro-computed tomography study.

Authors:  Ashraf ElAyouti; Maren Hülber-J; Martin S Judenhofer; Thomas Connert; Julia G Mannheim; Claus Löst; Bernd J Pichler; Christiane von Ohle
Journal:  J Endod       Date:  2014-01-17       Impact factor: 4.171

Review 6.  Clinical efficacy of electronic apex locators: systematic review.

Authors:  Jorge N R Martins; Duarte Marques; António Mata; João Caramês
Journal:  J Endod       Date:  2014-05-02       Impact factor: 4.171

7.  Evaluation of the Reliability of Cone-beam Computed Tomography Scanning and Electronic Apex Locator Measurements in Working Length Determination of Teeth with Large Periapical Lesions.

Authors:  Yakup Üstün; Tuğrul Aslan; Ahmet Ercan Şekerci; Burak Sağsen
Journal:  J Endod       Date:  2016-07-30       Impact factor: 4.171

8.  An in vivo evaluation of Endex electronic apex locator.

Authors:  A L Frank; M Torabinejad
Journal:  J Endod       Date:  1993-04       Impact factor: 4.171

9.  In vivo measurement accuracy in vital and necrotic canals with the Endex apex locator.

Authors:  D L Mayeda; J H Simon; D F Aimar; K Finley
Journal:  J Endod       Date:  1993-11       Impact factor: 4.171

10.  The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device.

Authors:  A Cemal Tinaz; L Sibel Sevimli; Güliz Görgül; Emin G Türköz
Journal:  J Endod       Date:  2002-03       Impact factor: 4.171

View more
  1 in total

1.  Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Authors:  Franck Diemer; Emma Plews; Marie Georgelin-Gurgel; Lora Mishra; Hyeon-Cheol Kim
Journal:  Materials (Basel)       Date:  2022-01-23       Impact factor: 3.623

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.