Abstract
OBJECTIVE: The aim of this study was to compare the accuracy of three electronic apex locators (EALs) (Propex Pixi, Mini Root ZX, Raypex 5) in determining working length (WL) under different environments (existence of blood-pulp/sodium hypochlorite in root canal space) using micro-computed tomography (micro-CT) measurements.Entities:
Year: 2020 PMID: 33353910 PMCID: PMC7881377 DOI: 10.14744/eej.2020.30633
Source DB: PubMed Journal: Eur Endod J ISSN: 2548-0839
Figure 1. 3D model of the root canal is shown at (a) The apical constriction point of root canal is shown at (b) with “a” and also coronal reference point is shown with yellow line. The horizontal micro-CT section of apical constriction is shown with smaller figure
Mean and Standard Deviation and error (mm) obtained for distances to AC in absolute value according to the tested EALs under different circumtances
| Groups | Mean | SD | SD error |
|---|---|---|---|
| Root ZX Blood | 0.385 | 0.299 | 0.0611 |
| Root ZX NaOCl | 0.359 | 0.266 | 0.0543 |
| Raypex Blood | 0.429 | 0.296 | 0.0605 |
| Raypex NaOCl | 0.47 | 0.274 | 0.056 |
| Pixi Blood | 0.36 | 0.259 | 0.0529 |
| Pixi NaOCl | 0.464 | 0.297 | 0.0607 |
SD: Standard deviation
Figure 2The horizontal line indicates the position of the AC found in Micro Ct measurements. Positive values (above the line) indicate measurements beyond the AC; negative values are short of the AC