Literature DB >> 35160807

Effect of Sodium Hypochlorite Concentration on Electronic Apex Locator Reliability.

Franck Diemer1, Emma Plews1, Marie Georgelin-Gurgel1, Lora Mishra2, Hyeon-Cheol Kim3.   

Abstract

This ex vivo study aimed to measure the performance of an electronic apex locator (EAL) in the presence of sodium hypochlorite irrigants with different concentrations. Two EALs (Root ZX Mini and Locapex 6) were used to locate the apical foramen in 10 extracted single-rooted teeth in the presence of 0.5%, 2.5% and 5% sodium hypochlorite. Visual working lengths were also determined using #10 K-file under a microscope before the electronic measurements were made. The performance of both EALs was compared for the electronic working lengths determined under the different concentrations of sodium hypochlorite. A multiple-way ANOVA and PLSD Fisher's test with an α risk fixed at 5% were conducted. There were no statistical differences in the working lengths determined by both EALs between the three groups with different concentrations of sodium hypochlorite and their visual control measurements. When a ± 0.5 mm margin was applied, the Root ZX Mini and the Locapex 6 presented 88% and 83% accuracy, respectively. Sodium hypochlorite concentration in irrigants does not affect the accuracy and reliability of either the Root ZX Mini or the Locapex 6. Electronic apex locators are reliable with any concentration of sodium hypochlorite irrigants.

Entities:  

Keywords:  concentration; electronic apex locators; sodium hypochlorite; working length

Year:  2022        PMID: 35160807      PMCID: PMC8836685          DOI: 10.3390/ma15030863

Source DB:  PubMed          Journal:  Materials (Basel)        ISSN: 1996-1944            Impact factor:   3.623


  35 in total

1.  Accuracy of three electronic apex locators in the presence of different irrigating solutions.

Authors:  Ana Laura Pion Carvalho; Cacio Moura-Netto; Abilio Albuquerque Maranhão de Moura; Márcia Martins Marques; Harry Davidowicz
Journal:  Braz Oral Res       Date:  2010 Oct-Dec

Review 2.  The Precision of Electronic Apex Locators in Working Length Determination: A Systematic Review and Meta-analysis of the Literature.

Authors:  Igor Tsesis; Tamar Blazer; Gil Ben-Izhack; Silvio Taschieri; Massimo Del Fabbro; Stefano Corbella; Eyal Rosen
Journal:  J Endod       Date:  2015-09-26       Impact factor: 4.171

3.  Evaluation of the accuracy of three electronic apex locators using glass tubules.

Authors:  W Fan; B Fan; J L Gutmann; Z Bian; M W Fan
Journal:  Int Endod J       Date:  2006-02       Impact factor: 5.264

4.  Critical diameter of apical foramen and of file size using the Root ZX apex locator: an in vitro study.

Authors:  Manuela Herrera; Camilo Ábalos; Cristina Lucena; Amparo Jiménez-Planas; Rafael Llamas
Journal:  J Endod       Date:  2011-07-16       Impact factor: 4.171

5.  Ex vivo accuracy of Root ZX II, Root ZX Mini and RomiApex A-15 apex locators in extracted vital pulp teeth.

Authors:  Thaís M da Silva; Flávio R F Alves
Journal:  J Contemp Dent Pract       Date:  2014-05-01

Review 6.  Electronic canal length measurement.

Authors:  C Kobayashi
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  1995-02

7.  The Influence of Various Irrigants on the Accuracy of 2 Electronic Apex Locators in Locating Simulated Root Perforations.

Authors:  Demet Altunbaş; Alper Kuştarcı; Mustafa Toyoğlu
Journal:  J Endod       Date:  2017-01-16       Impact factor: 4.171

8.  The effects of sodium hypochlorite concentrations on the accuracy of an apex locating device.

Authors:  A Cemal Tinaz; L Sibel Sevimli; Güliz Görgül; Emin G Türköz
Journal:  J Endod       Date:  2002-03       Impact factor: 4.171

9.  The effects of file size, sodium hypochlorite and blood on the accuracy of Root ZX apex locator in enlarged root canals: an in vitro study.

Authors:  A K Ebrahim; T Yoshioka; C Kobayashi; H Suda
Journal:  Aust Dent J       Date:  2006-06       Impact factor: 2.291

10.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02
View more