Literature DB >> 27484249

Evaluation of the Reliability of Cone-beam Computed Tomography Scanning and Electronic Apex Locator Measurements in Working Length Determination of Teeth with Large Periapical Lesions.

Yakup Üstün1, Tuğrul Aslan2, Ahmet Ercan Şekerci3, Burak Sağsen2.   

Abstract

INTRODUCTION: This study evaluates the endodontic working-length measurements in teeth with large periapical lesions and persistent intracanal exudate by using preexisting cone-beam computed tomography (CBCT). It compares the measurements with clinical root canal lengths determined by using 2 electronic apex locators.
METHODS: All patients had undergone a CBCT scan independent of the present study and needed root canal treatment of at least 1 tooth visible in the field of view. Seventy-three teeth with single roots and canals were studied. An endodontist measured each root canal length with 2 different electronic apex locators. The measurements were repeated 3 times by using a digital caliper, and the mean was recorded. This mean was compared with the root canal length as measured on CBCT sections by an oral radiologist not involved in the endodontic treatment. The CBCT measurements were taken twice for analysis of intrarater reliability. The data were statistically analyzed.
RESULTS: There were no significant measurement differences between the methods used (P > .05).
CONCLUSIONS: In teeth with large periapical lesions and persistent intracanal exudate, measurement of the root canal length by using CBCT was as reliable as measurements that used apex locators.
Copyright © 2016 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Apex locator; Propex Pixi; Raypex 6; cone-beam computed tomography; large periapical lesion; root canal length measurement

Mesh:

Year:  2016        PMID: 27484249     DOI: 10.1016/j.joen.2016.06.010

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  6 in total

Review 1.  Use of preoperative cone-beam computed tomography to aid in establishment of endodontic working length: A systematic review and meta-analysis.

Authors:  Andrew Paterson; Vittorio Franco; Shanon Patel; Federico Foschi
Journal:  Imaging Sci Dent       Date:  2020-09-16

2.  Accuracy of Software-Based Three-Dimensional Root Canal Length Measurements Using Cone-Beam Computed Tomography.

Authors:  J P Tchorz; K T Wrbas; C Von See; K Vach; Sbm Patzelt
Journal:  Eur Endod J       Date:  2018-11-13

3.  The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures.

Authors:  Inês Ferreira; Ana Cristina Braga; Irene Pina-Vaz
Journal:  Eur Endod J       Date:  2019-07-10

4.  Ex Vivo Evaluation of the Accuracy of 3 Electronic Apex Locators in Different Environments: A Micro-Computed Tomography Study.

Authors:  Firdevs Çınar; Yakup Üstün
Journal:  Eur Endod J       Date:  2020-12

5.  Root canal length measurement of molar teeth using conebeam computed tomography (CBCT): comparison of two dimensional versus three-dimensional methods.

Authors:  Selen Nihal Sisli; Orhan Gulen
Journal:  Eur Oral Res       Date:  2021-05-04

6.  In vitro Evaluation of the Accuracy of the Root Zx in the Presence of Naocl 2.5% and Chlorhexidine 0.2.

Authors:  Marjan Bolbolian; Siamak Golchin; Seyedmatin Faegh
Journal:  J Clin Exp Dent       Date:  2018-11-01
  6 in total