Influence of preflaring on the accuracy of length determination with four electronic apex locators.

Ericson Janolio de Camargo1, Ronald Ordinola Zapata, Paulo Leal Medeiros, Clovis Monteiro Bramante, Norberti Bernardineli, Roberto Brandão Garcia, Ivaldo Gomes de Moraes, Marco Antônio Húngaro Duarte.   

Abstract

INTRODUCTION: The aim of this study was to compare the influence of preflaring on the accuracy of 4 electronic apex locators (EALs): Root ZX, Elements Diagnostic Unit and Apex Locator, Mini Apex Locator, and Apex DSP.
METHODS: Forty extracted teeth were preflared by using S1 and SX ProTaper instruments. The working length was established by reducing 1mm from the total length (TL). The ability of the EALs to detect precise (-1mm from TL) and acceptable (-1+/-0.5 mm from TL) measurements in unflared and preflared canals was determined.
RESULTS: The precise and acceptable (P/A) readings in unflared canals for Root ZX, Elements Diagnostic Unit and Apex Locator, Mini Apex, and Apex DSP were 50%/97.5%, 47.5%/95%, 50%/97.5%, and 45%/ 67.5%, respectively. For preflared canals, the readings were 75%/97.5%, 55%/95%, 75%/97.5%, and 60%/87.5%, respectively. For precise criteria, the preflared procedure increased the percentage of accurate electronic readings for the Root ZX and the Mini Apex Locator (P < .05). For acceptable criteria, no differences were found among Root ZX, Elements Diagnostic Unit and Apex Locator, and Mini Apex Locator (P > .05). Fisher test indicated the lower accuracy for Apex DSP (P < .05).
CONCLUSIONS: The Root ZX and the Mini Apex Locator devices increased significantly the precision to determine the real working length after the preflaring procedure. All the EALs showed an acceptable determination of the working length between the ranges of+/-0.5mm except for the Apex DSP device, which had the lowest accuracy.

Mesh:

Year:  2009        PMID: 19720236     DOI: 10.1016/j.joen.2009.05.030

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  14 in total

1.  The use of cone beam computed tomography to predetermine root canal lengths in molar teeth: a comparison between two-dimensional and three-dimensional measurements.

Authors:  J P Tchorz; P J Poxleitner; S Stampf; S B M Patzelt; D Rottke; E Hellwig; M J Altenburger
Journal:  Clin Oral Investig       Date:  2013-07-30       Impact factor: 3.573

2.  Accuracy of five electronic foramen locators with different operating systems: an ex vivo study.

Authors:  Bruno Carvalho de Vasconcelos; Michelli de Medeiros Bueno; Suyane Maria Luna-Cruz; Marco Antonio Hungaro Duarte; Carlos Augusto de Oliveira Fernandes
Journal:  J Appl Oral Sci       Date:  2013 Mar-Apr       Impact factor: 2.698

3.  In vivo accuracy of conventional and digital radiographic methods in confirming root canal working length determination by Root ZX.

Authors:  Fernando Accorsi Orosco; Norberti Bernardineli; Roberto Brandão Garcia; Clovis Monteiro Bramante; Marco Antonio Húngaro Duarte; Ivaldo Gomes de Moraes
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

4.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

5.  Comparison of accuracy of two electronic apex locators in the presence of various irrigants: An in vitro study.

Authors:  J Paras Mull; Vinutha Manjunath; Mk Manjunath
Journal:  J Conserv Dent       Date:  2012-04

6.  Calcium hydroxide dressing residues after different removal techniques affect the accuracy of Root-ZX apex locator.

Authors:  Emel Uzunoglu; Ayhan Eymirli; Mehmet Özgür Uyanik; Semra Çalt; Emre Nagas
Journal:  Restor Dent Endod       Date:  2014-11-05

7.  Root ZX Electronic Foramen Locator: An Ex Vivo Study of Its Three Models' Precision and Reproducibility.

Authors:  Bernardo Almeida Aguiar; Rafael Santos Reinaldo; Luciana Maria Arcanjo Frota; Mônica Sampaio do Vale; Bruno Carvalho de Vasconcelos
Journal:  Int J Dent       Date:  2017-03-07

8.  Influence of calcium hydroxide residues after using different irrigants on the accuracy of two electronic apex locators: An in vitro study.

Authors:  Nooshin Sadat Shojaee; Zahra Zaeri; Mohammad Mehdi Shokouhi; Fereshteh Sobhnamayan; Alireza Adl
Journal:  Dent Res J (Isfahan)       Date:  2020-01-21

9.  Comparative evaluation of the accuracy of two electronic apex locators in determining the working length in teeth with simulated apical root resorption: An in vitro study.

Authors:  Vidya Saraswathi; Archit Kedia; Tina Puthen Purayil; Vasudev Ballal; Aakriti Saini
Journal:  J Conserv Dent       Date:  2016 Sep-Oct

10.  Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study.

Authors:  Masoud Saatchi; Shiva Iravani; Mehrdad Akhavan Khaleghi; Amin Mortaheb
Journal:  Iran Endod J       Date:  2017
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.