Abstract
PURPOSE: To predict the anterior chamber volume (ACV) after implantable collamer lens (ICL) implantation based on ICL size and parameters of anterior segment optical coherence tomography (AS-OCT).Entities:
Mesh:
Year: 2020 PMID: 33196664 PMCID: PMC7668562 DOI: 10.1371/journal.pone.0242434
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Patient demographic information.
| Prediction group | Verification group | |
|---|---|---|
| No. of patients | 148 | 74 |
| Sex (male, female) | 64, 84 | 36, 38 |
| Age (years) | 32.11 ± 8.04 | 33.03 ± 6.74 |
| Manifest refractive sphere (D) | −7.34 ± 2.44 | −7.98 ± 2.48 |
| Preoperative IOP (mmHg) | 13.7 ± 3.00 | 14.1 ± 2.66 |
IOP, intraocular pressure; D, diopter.
Values were presented as mean ± standard deviation
The sex was verified through health insurance cards.
AS-OCT parameters and results of single linear regression analysis with postoperative ACV.
| Mean ± SD | Range (Min, Max) | Adjusted R2 | P | |
|---|---|---|---|---|
| ACV (mm3) | 190.03 ± 27.40 | 131.53, 251.23 | 0.762 | <0.001 |
| ICL size (mm) | 12.80 ± 0.34 | 12.1, 13.2 | 0.069 | 0.202 |
| ACW (mm) | 11.84 ± 0.41 | 10.82, 12.95 | 0.564 | <0.001 |
| ACD (mm) | 3.312 ± 0.25 | 2.77, 3.98 | 0.652 | <0.001 |
| LV (mm) | −0.15 ± 0.19 | -0.53, 0.27 | −0.378 | <0.001 |
| AOD250 (mm) | 0.40 ± 0.16 | 0.081, 0.85 | 0.314 | <0.001 |
| AOD500 (mm) | 0.62 ± 0.21 | 0.22, 1.18 | 0.413 | <0.001 |
| AOD750 (mm) | 0.87 ± 0.26 | 0.35, 1.6 | 0.465 | <0.001 |
| ARA250 (mm2) | 0.09 ± 0.051 | 0.003, 0.28 | 0.243 | 0.001 |
| ARA500 (mm2) | 0.23 ± 0.094 | 0.036, 0.53 | 0.314 | <0.001 |
| ARA750 (mm2) | 0.42 ± 0.15 | 0.14, 0.86 | 0.377 | <0.001 |
| TISA250 (mm2) | 0.08 ± 0.035 | 0.003, 0.19 | 0.285 | <0.001 |
| TISA500 (mm2) | 0.21 ± 0.80 | 0.036, 0.43 | 0.340 | <0.001 |
| TISA750 (mm2) | 0.40 ± 0.13 | 0.14, 0.76 | 0.396 | <0.001 |
| TIA250 (deg) | 54.97 ± 14.30 | 17.9, 119.3 | 0.353 | <0.001 |
| TIA500 (deg) | 49.81 ± 11.48 | 23.9, 98.1 | 0.472 | <0.001 |
| TIA750 (deg) | 48.05 ± 9.44 | 26.1, 87.1 | 0.520 | <0.001 |
| I-Curv (mm) | -0.01 ± 0.12 | -0.27, 0.21 | -0.371 | <0.001 |
Values are presented as mean ± standard deviation.
ACV, anterior chamber volume; ACW, anterior chamber width; ACD, anterior chamber depth; LV, lens vault; AOD, angle open distance; ARA, angle recess area; TISA, trabecular iris space area: TIA, trabecular iris angle
Stepwise multiple regression analysis of factors influencing postoperative ACV in patients with ICL.
| Independent Variables | Unstandardized Regression Coefficient | Standardized Partial Regression Coefficient | 95% CI | VIF | P |
|---|---|---|---|---|---|
| Constant | 87.655 | −28.268, 203.579 | 0.137 | ||
| Preoperative ACV | 0.466 | 0.470 | 0.271, 0.661 | 5.359 | <0.001 |
| ICL size | −43.036 | −0.532 | −52.589, −33.483 | 1.938 | <0.001 |
| ACW | 40.152 | 0.602 | 27.747, 52.558 | 4.807 | <0.001 |
| AOD750 | 17.003 | 0.161 | 2.180, 31.825 | 2.736 | 0.025 |
Values are presented as mean ± standard deviation.
ICL, implantable collamer lens; CI, confidential interval; VIF, variance inflation factor; ACV, Anterior Chamber Volume; ACW, Anterior Chamber Width; AOD, Angle Open Distance
Fig 1Correlation between the actual and the predictive postoperative anterior chamber volume (ACV).
There was a significant correlation between the postoperative and the predicted ACV values (R2 = 0.6996, P < 0.001).
Fig 2Agreement between the actual postoperative and the predictive anterior chamber volume (ACV).
Bland-Altman plot analysis showed no addition and proportional errors between the actual and predicted ACV values.
Fig 3Case presentation showing narrow angle after implantable collamer lens (ICL) implantation.
Anterior segment optical computed tomography images at the horizontal meridian before and after ICL implantation are shown (A, before; B, 3 months after surgery). In this case, the ICL size (12.6 mm) was calculated using the vault calculation equation, and a 12.6-mm ICL was implanted. As a result, angle narrowing occurred after surgery. The predictive postoperative ACV value (C) after the 12.6-mm ICL implantation was apparently low, whereas implantation of the 12.1-mm ICL was more suitable according to the postoperative ACV prediction. The generated ACV prediction equation may help determine the appropriate ICL size for such cases.