| Literature DB >> 33173072 |
T G Wolf1,2, F Fischer3,4, R K W Schulze4.
Abstract
To investigate potential correlations between objective CBCT image parameters and accuracy in endodontic working length determination ex vivo. Contrast-to-noise ratio (CNR) and spatial resolution (SR) as fundamental objective image parameters were examined using specific phantoms in seven different CBCT machines. Seven experienced observers were instructed and calibrated. The order of the CBCTs was randomized for each observer and observation. To assess intra-operator reproducibility, the procedure was repeated within six weeks with a randomized order of CBCT images. Multivariate analysis (MANOVA) did not reveal any influence of the combined image quality factors CNR and SR on measurement accuracy. Inter-operator reproducibility as assessed between the two observations was poor, with a mean intra-class correlation (ICC) of 0.48 (95%-CI 0.38, 0.59) for observation No. 1. and 0.40 (95%-CI 0.30, 0.51) for observation No. 2. Intra-operator reproducibility pooled over all observers between both observations was only moderate, with a mean ICC of 0.58 (95%-CI 0.52 to 0.64). Within the limitations of the study, objective image quality measures and exposure parameters seem not to have a significant influence on accuracy in determining endodontic root canal lengths in CBCT scans. The main factor of variance is the observer.Entities:
Mesh:
Year: 2020 PMID: 33173072 PMCID: PMC7655868 DOI: 10.1038/s41598-020-76424-4
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
CBCT devices and parameters used for the evaluation. FOV: field of view. Voxel size provided as stated by the manufacturer.
| CBCT device | kV | mA | Scan-time (s) | FOV (mm) | Voxel size (mm) |
|---|---|---|---|---|---|
| 3D Accuitomo 80 (J Morita Corp, Kyoto, Japan) | 90 | 8 | 17.5 | 80 × 80 | 0.160 |
| 3D eXam (KaVo Dental, Biberach, Germany) | 120 | 5 | 14.7 | 160 × 80 | 0.20 |
| Veraviewepocs 3D R100 (J Morita Corp, Kyoto, Japan) | 90 | 8 | 9.4 | 100 × 80 | 0.125 |
| PaX-Duo3D (Vatech, Gyeonggi-do, Korea) | 90 | 8 | 24 | 85 × 85 | 0.2 |
| Sanora 3Dx (Sorodex, Tuusula, Finland) | 90 | 8 | 20 | 100 × 80 | 0.15 |
| ProMax 3D Mid (Planmeca Oy, Helsinki, Finland) | 90 | 8 | 12 | 80 × 80 | 0.14 |
| Orthophos SL (Dentsply Sirona, Bensheim, Germany) | 85 | 7 | 14.2 | 80 × 80 | 0.160 |
Figure 1Resulting (exemplary) CBCT scans with the polygon measurement tool.
Objective image quality parameters. CNR: contrast to noise ratio (Eq. 1), MTF 10%: approximated modulation transfer function at 10% modulation.
| CBCT device | CNR (air–water) | MTF 10% (Lp/mm) |
|---|---|---|
| 3D Accuitomo 80 (J Morita Corp, Kyoto, Japan) | 0.67 | 2.4 |
| 3D eXam (KaVo Dental, Biberach, Germany) | 1.94 | 1.8 |
| Veraviewepocs 3D R100 (J Morita Corp, Kyoto, Japan) | 1.03 | 2.1 |
| PaX-Duo3D (Vatech, Gyeonggi-do, Korea) | 0.18 | 2.3 |
| Scanora 3Dx (Sorodex, Tuusula, Finland) | 0.91 | 2.3 |
| ProMax 3D Mid (Planmeca Oy, Helsinki, Finland) | 0.85 | 2.5 |
| Orthophos SL (Dentsply Sirona, Bensheim, Germany) | 0.79 | 2.8 |
Figure 2Box plots of the absolute errors pooled over the seven observers for the first and second observation. Boxes represent the data between the 25% and the 75%-quartiles, while whiskers represent the highest and lowest values except for outliers. The latter are those values lying outside 1.5 times the interquartile range above the upper quartile and below the lower quartile.
Figure 3Box plots of the absolute errors separated for the seven observers, yet pooled over the two observations. Boxes represent the data between the 25% and the 75%-quartiles, while whiskers represent the highest and lowest values except for outliers. The latter are those values lying outside 1.5 times the interquartile range above the upper quartile and below the lower quartile.