| Literature DB >> 35958289 |
Daniel Reitz1, Johannes Muecke1, Vanessa da Silva Mendes1, Guillaume Landry1, Michael Reiner1, Maximilian Niyazi1, Claus Belka1, Philipp Freislederer1, Stefanie Corradini1.
Abstract
Background and purpose: Patients receiving cranial radiotherapy are immobilized with a thermoplastic mask to restrict patient motion. Depending on the target volume margins and treatment dose, different mask systems are used. Intrafractional movements can be monitored using stereoscopic X-ray imaging. The aim of the present work was to compare the magnitudes of intrafractional deviation for different mask systems. Material and methods: Four different head mask systems (open face mask, open mask, stereotactic mask, double mask) used in the treatment of 40 patients were investigated. In total 487 treatment fractions and 3708 X-ray images were collected. Deviations were calculated by comparison of the acquired X-ray images with digitally reconstructed radiographs. The results of intrafractional X-ray deviations for translational and rotational axes were compared between the different mask systems.Entities:
Year: 2022 PMID: 35958289 PMCID: PMC9361321 DOI: 10.1016/j.phro.2022.07.002
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Number of patients, treatment sessions and stereoscopic images as well as entities and kind of treatment per mask1.
| BLOF | BLSRS | ITVDM | ITVOM | Sum | ||
|---|---|---|---|---|---|---|
| Patients number | 10 | 9 | 12 | 9 | 40 | |
| Treatment sessions | 198 | 25 | 63 | 201 | 487 | |
| Stereoscopic images | 826 | 393 | 323 | 732 | 2274 | |
| 539 | 81 | 300 | 514 | 1434 | ||
| Images per session | 7 | 19 | 10 | 6 | ||
| Entity/Treatment | 4 | 1 | 1 | 4 | 10 | |
| 3 | 1 | 0 | 4 | 8 | ||
| 1 | 0 | 0 | 1 | 2 | ||
| 1 | 1 | 7 | 0 | 9 | ||
| 0 | 6 | 4 | 0 | 10 | ||
| 1 | 0 | 0 | 0 | 1 | ||
VS: vestibular schwannoma; WBRT: whole brain radiotherapy; FSRT: fractionated stereotactic radiotherapy; FRT: fractionated radiotherapy; SRS: stereotactic radiosurgery.
Fig. 1A: Box plots of intrafractional motion in 6 DOF and translational deviation vector for the four different mask systems; B-H: Empirical cumulative distribution functions of absolute 6 DOF and translational deviation vector for the four different mask systems; dashed horizontal black lines are showing the lower and upper limit of the 95% confidence interval (n = 40 patients, 3708 stereoscopic images, coplanar configuration).