Image-guided stereotactic radiosurgery for cranial lesions: large margins compensate for reduced image guidance frequency.

Harun Badakhshi1, David Kaul, Peter Wust, Edzard Wiener, Volker Budach, Volker Buadch, Reinhold Graf.   

Abstract

BACKGROUND: We investigated patient positioning during radiosurgery of cranial lesions, and calculated clinical target volume (CTV) to planning target volume (PTV) margins using a modified common margin recipe. We simulated CTV-to-PTV margins for reduced image guidance, and repositioning for the first table angle only. PATIENTS AND METHODS: Patients were immobilized with a thermoplastic mask. Positioning was verified and corrected using the ExacTrac/Novalis Body. Each patient was repositioned before each beam. A common margin recipe was adapted for estimation of CTV-to-PTV margins. Necessary margins were estimated to correct positioning for the initial table angle only in comparison.
RESULTS: In total, 269 radiosurgery treatments with 967 different-angle setups (mean 3.6 different angles) were performed on 190 patients. Residual translational errors were (one standard deviation) 0.3 mm in left-right (LR), superior-inferior (SI), and anterior-posterior (AP) directions, with a mean three-dimensional vector of 0.5 mm. Margins for residual errors after correction were calculated in LR, SI, and AP directions as 0.8 mm. For simulated reduced frequency setup correction, we calculated CTV-to-PTV margins as 1.9, 1.9, and 1.6 mm, respectively.
CONCLUSIONS: The ExacTrac/Novalis Body system allows for accurate positioning of the patient with a residual error comparable to invasive mask fixation. If verification is only performed after initial positioning, adaption of CTV-to-PTV margins should be considered.

Entities:  

Keywords:  Margin; frameless stereotactic radiosurgery; image guidance; image-guided intervention

Mesh:

Year:  2013        PMID: 24123042

Source DB:  PubMed          Journal:  Anticancer Res        ISSN: 0250-7005            Impact factor:   2.480


  4 in total

1.  Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation.

Authors:  Barbara Zollner; Christian Heinz; Sabrina Pitzler; Farkhad Manapov; Steffi Kantz; Maya Christine Rottler; Maximilian Niyazi; Ute Ganswindt; Claus Belka; Hendrik Ballhausen
Journal:  Radiat Oncol       Date:  2016-12-07       Impact factor: 3.481

2.  Monitoring frequency of intra-fraction patient motion using the ExacTrac system for LINAC-based SRS treatments.

Authors:  Benjamin C Lewis; William J Snyder; Siyong Kim; Taeho Kim
Journal:  J Appl Clin Med Phys       Date:  2018-03-25       Impact factor: 2.102

3.  The potential of an optical surface tracking system in non-coplanar single isocenter treatments of multiple brain metastases.

Authors:  Ans C C Swinnen; Michel C Öllers; Chin Loon Ong; Frank Verhaegen
Journal:  J Appl Clin Med Phys       Date:  2020-04-01       Impact factor: 2.102

4.  A retrospective analysis of setup and intrafraction positional variation in stereotactic radiotherapy treatments.

Authors:  Micah Barnes; Adam Yeo; Kenton Thompson; Claire Phillips; Tomas Kron; Nicholas Hardcastle
Journal:  J Appl Clin Med Phys       Date:  2020-11-03       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.