Mats Geijer1,1,1, Sverrir Kiernan2, Martin Sundberg2, Gunnar Flivik2. 1. Department of Radiology, University of Gothenburg, Gothenburg, Sweden. 2. Department of Radiology, Sahlgrenska University Hospital, Gothenburg, Sweden.
In total hip arthroplasty (THA), restoration of anatomy and seeking for a correct
biomechanical situation is crucial for function.[1-4] Global offset (GO; the sum of
acetabular offset (AO) and femoral offset (FO)) and femoral neck anteversion (FNA)
are measurements that may be used to evaluate postoperative outcome regarding
implant position and degree of restored anatomy in THA. Imbalance in GO may lead to
limping. Reduced FO may increase acetabular polyethylene wear.[5] FNA below 10° appears deleterious to the long-term outcome for cemented stems.[6]To make as good anatomical restoration and biomechanical situation as possible during
THA, it is common to make a preoperative plan on conventional radiographs, so-called
templating, to know what implant to use and in what position to insert it. If the
affected side, planned for surgery, is too deformed, the planning can be made on the
contralateral, hopefully, more normal side. The goal of surgery is to deepen the
acetabular socket and reduce the AO to make room for the acetabular component and,
at the same time, increase FO to keep GO constant. Asymmetry of the different
elements or measurements of the hip joints may, however, render the contralateral
hip unsuitable to use for templating, and several articles have reported significant
asymmetry of different measurements.[7-9] Contrarily, a high degree of hip
joint symmetry has also been reported.[10],[11]FO is traditionally measured on anteroposterior (AP) pelvic radiographs,[12] where several methods of measurement exist.[12],[13] It has been shown that measurements on AP pelvic radiographs (here called
functional FO) underestimate the true FO[14],[15] (Fig. 1) and that it
is influenced by hip rotation[16] and flexion.[17] The need for more exact measurement techniques has been suggested for more
exact templating methods and navigational assistance[18] at surgery. 3D measurements using CT have provided measurements with high reproducibility.[19]
Fig. 1.
Illustration of the true femoral offset (TFO), which is the true
perpendicular distance from the proximal femoral long axis and the
rotational center of the femoral head, and the functional femoral offset
(FFO), which is the horizontal distance from the proximal femoral long axis
and the rotational center of the femoral head.
FO: femoral offset.
Illustration of the true femoral offset (TFO), which is the true
perpendicular distance from the proximal femoral long axis and the
rotational center of the femoral head, and the functional femoral offset
(FFO), which is the horizontal distance from the proximal femoral long axis
and the rotational center of the femoral head.FO: femoral offset.Measurements of FNA have traditionally been done using conventional radiographs with
various methods,[20-22] which has been
seen as sufficiently exact to provide measurements for manual templating and manual
orientation of the implants. Calculations in three dimensions (3D)[23] have provided more precise measurements. Initial CT measurements of FNA
differed considerably from measurements done on radiographs,[24] and different CT measuring methods gave widely different results.[25],[26] It has been shown that FNA measurements are dependent on the positioning of
the femur for radiographic analysis as well as for two-dimensional CT analysis.[27] Measurements performed using 3D data sets have shown high consistency for
both intra- and interobserver agreements.[28]The current study aimed to evaluate pre- and postoperative proximal femoral symmetry
by semi-automated 3D CT measurements of FNA and the different offsets in a cohort
scheduled for THA. A secondary aim was to validate the software measurements by
inter- and intraobserver agreement calculations.
Material and Methods
Patients
The Regional Ethical Review Board approved the study (2009/369). The current
cohort was based on a clinical study, reported elsewhere.[29] In that study, 75 patients were recruited for THA after informed consent
and operated with the uncemented ABG II prosthesis (Stryker Orthopedics, Mahwah,
NJ, USA). Due to incomplete imaging for various reasons, there were missing data
for two patients preoperatively and two other patients postoperatively. For 71
patients with complete pre- and postoperative CT (45 males, 26 females), mean
age was 59.1 years (SD 8.1 years).
CT examinations
Pre- and postoperative imaging included CT of the hips and knees for 3D
assessment of AO, true and functional FO, and FNA. CT was performed on a
multidetector helical Brilliance 64 CT scanner (Philips, Eindhoven, The
Netherlands). Low-dose settings were used for the preoperative study (hips, CT
dose index by volume (CTDIvol) 4.8; knees, 4.2) and slightly higher dose
settings for the postoperative study to compensate for the implant (hips,
CTDIvol 16.4; knees, unchanged).[30] Helical CT was performed from the mid-pelvis, including the anterior
superior iliac spine to about 6 cm distal to the lesser trochanter, and from
directly proximal to the femoral condyles to directly distal to the knee joint.
Postoperative imaging covered the same area. Images were archived in the local
picture archiving and communication system.
Image evaluation—3D assessments
The pre- and postoperative 3D-CT examinations were evaluated using a CT-based 3D
templating software (Ortoma Plan™, Ortoma AB, Gothenburg, Sweden), giving
measurements for AO, true and functional FO, and FNA. The templating software
assigned the pelvis and knees CT scan volumes to a combined 3D volume. Thick
slab multiplanar reformations (MPRs) were provided in the orthogonal planes by
the software (Fig. 2).
The pelvis volume was rotated to the anterior pelvic plane[31] by first defining the bi-ischial line as a horizontal line between the
inferior points of the ischial tuberosities or the teardrops on an AP thick slab
MPR, secondly the sagittal plane as perpendicular to a line drawn between the
anterior superior iliac spines on an axial reconstruction, and thirdly the
anterior pelvic plane as a line between the anterior superior iliac spines and
the anterior point of the symphysis pubis on a lateral reconstruction. The
reconstructed volume was stepwise automatically rotated to the described
planes.
Fig. 2.
Screen capture showing the best-fit circles defining the femoral head,
the definition of the long axis of the proximal femur, acetabular
offset, and femoral offset (FO). Functional FO was measured on the
coronal reformation, true FO was measured in three dimensions. The
external objects at the right hip are bone density phantoms.
Screen capture showing the best-fit circles defining the femoral head,
the definition of the long axis of the proximal femur, acetabular
offset, and femoral offset (FO). Functional FO was measured on the
coronal reformation, true FO was measured in three dimensions. The
external objects at the right hip are bone density phantoms.The long axis of the proximal femur was defined by assigning best-fit spheres on
the AP MPR at the level of the distal border of the lesser trochanter in the CT
volume (Figs 2 and 3(a), point i), touching
the inner surface of the medial and lateral cortices, and one in the proximal
femur about 6 cm distal to the first, proximal to any visible bowing (Fig. 3(a), point ii). The
position of the spheres was then adjusted on the lateral and axial views. The
rotational center of the femoral head was defined by assigning a best-fit circle
on all three thick slab MPRs (Fig. 3(a), point iii). The condylar line of the knee was found by
assigning a line tangential to the posterior subchondral joint surface of both
femoral condyles (Fig.
3). Both points were then adjusted in the craniocaudal direction on the
lateral view to the most dorsal point of the femoral condyles. The condylar
plane was defined by the condylar line and the intersection on the long axis for
true FO (Fig. 3(a),
point iv). A central point in the knee was found at the midpoint of a line drawn
between the lateral and medial femoral epicondyle, with the point adjusted in
height to the same level as the posterior condylar line. The software then
calculated the respective measurements.
Fig. 3.
Schematic 3D illustration of the definition of femoral neck anteversion
(FNA) in the current study. The long axis of the femur is defined by two
points in the proximal femur: (i) at the inferior border of the lesser
trochanter and (ii) at a point about 6 cm distal in the femoral shaft.
The FNA is the angle between a perpendicular line between the proximal
femoral long axis and (iii) the femoral head and a line between the
femoral condyles.
Schematic 3D illustration of the definition of femoral neck anteversion
(FNA) in the current study. The long axis of the femur is defined by two
points in the proximal femur: (i) at the inferior border of the lesser
trochanter and (ii) at a point about 6 cm distal in the femoral shaft.
The FNA is the angle between a perpendicular line between the proximal
femoral long axis and (iii) the femoral head and a line between the
femoral condyles.AO was defined as the distance from the midline of the symphysis pubis to the
rotational center of the femoral head. True FO was defined as the perpendicular
distance from the long axis of the proximal femur to the rotational center of
the femoral head measured in 3D12 (Fig. 1). Functional FO was defined as the
perpendicular distance from the long axis of the proximal femur to the
rotational center of the femoral head on the coronal MPR, i.e. not taking into
account the shortening of the offset due to the FNA, rotation, or flexion of the
femur, corresponding to the “classical” FO as measured on conventional radiographs.[32] FNA was defined as the angle (Fig. 3(c)) between the true FO line and
the condylar plane.Side-to-side comparisons of AO, true and functional FO, and FNA between the
non-affected and the osteoarthritic sides were performed on all patients.
Postoperatively, the same measurements were repeated to evaluate the surgical
outcome.The measurement technique was validated by performing an interobserver agreement
analysis on all pre- and postoperative measurements. For interobserver agreement
measurements, two observers (MG and SK) performed the above-mentioned
measurements on 71 pairs of hips on all pre- and postoperative studies, 284
measurements per observer.For intraobserver measurements, both observers
repeated the measurements on both hips on the pre- and postoperative CT of 15
randomly selected patients after about two months; 60 measurements per
observer.
Statistics
Continuous data are expressed as means and standard deviation. Qualitative data
are expressed as frequency and percentage. Observer agreement was analyzed with
intraclass correlation (ICC) with 95% confidence intervals (CI). The strength of
observer agreement was translated according to definitions proposed by Landis
and Koch for kappa values,[33] as 0.00–0.20, slight; 0.21–0.40, fair; 0.41–0.60, moderate; 0.61–0.80,
substantial; and 0.81–1.00, almost perfect. Furthermore, Lee et al.[34] stated that the lower 95% CI should be above 0.75 for an agreement to
exist. The software package R version 3.5.3 was used for statistical computations.[35]
Results
Preoperatively, there was a significantly larger AO on the osteoarthritic side of
2.5 mm (95% CI: 1.97–2.95), resulting in correspondingly larger true and functional
GO of 2.2 (CI: 1.52–2.88) and 2.6 mm (CI: 1.83–3.47), respectively. There were no
significant differences in preoperative true or functional FO or FNA between the
non-affected and the osteoarthritic sides (Table 1, Fig. 4).
Table 1.
Preoperative differences between non-arthritic and osteoarthritic hips in 71
patients. Combined data for two observers.
Mean
95% CI
Femoral anteversion angle (degrees)
–0.86
–2.14 to 0.42
Acetabular offset (mm)
2.46
1.97 to 2.95
True femoral offset (mm)
–0.26
–0.76 to 0.25
True global offset (mm)
2.20
1.52 to 2.88
Functional femoral offset (mm)
0.19
–0.53 to 0.91
Functional global offset (mm)
2.65
1.83 to 3.47
Fig. 4.
Bar chart showing the differences in measurements between the non-affected
side and the side planned for total hip arthroplasty for femoral neck
anteversion, acetabular offset, true femoral offset, true global offset,
functional femoral offset, and functional global offset. Data for the two
observers combined.
Preoperative differences between non-arthritic and osteoarthritic hips in 71
patients. Combined data for two observers.Bar chart showing the differences in measurements between the non-affected
side and the side planned for total hip arthroplasty for femoral neck
anteversion, acetabular offset, true femoral offset, true global offset,
functional femoral offset, and functional global offset. Data for the two
observers combined.Postoperatively, there was a significantly smaller AO on the operated side (–2.0 mm,
CI: –2.58 to –1.52) with a concomitantly increased true FO (2.5 mm, CI: 1.82–3.10)
and functional FO (1.4 mm, CI: 0.47–2.24) (Table 2, Fig. 5). This resulted in symmetry for both
true GO and functional GO, without significant side-to-side differences. There were
no significant side-to-side differences in FNA.
Table 2.
Postoperative differences between non-arthritic and osteoarthritic hips in 71
patients. Combined data for two observers.
Mean
95% CI
Femoral anteversion angle (degrees)
0.52
–1.65 to 2.68
Acetabular offset (mm)
–2.05
–2.58 to –1.52
True femoral offset (mm)
2.46
1.82 to 3.10
True global offset (mm)
0.41
–0.29 to 1.10
Functional femoral offset (mm)
1.35
0.47 to 2.24
Functional global offset (mm)
–0.70
–1.61 to 0.26
Fig. 5.
Bar chart showing the differences in measurements between the non-affected
side and the postoperative values for the operated side for femoral neck
anteversion, acetabular offset, true femoral offset, true global offset,
functional femoral offset, and functional global offset. Data for the two
observers combined.
Postoperative differences between non-arthritic and osteoarthritic hips in 71
patients. Combined data for two observers.Bar chart showing the differences in measurements between the non-affected
side and the postoperative values for the operated side for femoral neck
anteversion, acetabular offset, true femoral offset, true global offset,
functional femoral offset, and functional global offset. Data for the two
observers combined.Observer agreements measured by ICC for the two observers were high. Observer
agreement for 213 native hips, i.e. 71 pairs of hips on the preoperative CT and the
71 non-operated hips on the postoperative examination, was good with almost perfect
ICC scores and narrow CI (Table
3). Observer agreement for the 71 operated hips was equally good with
almost perfect ICC scores and narrow CI (Table 3), without differences between pre-
and postoperative results. Thus, the results showed no increased difficulty in
measuring on postoperative hip examinations. Intraobserver agreements were almost
perfect for both observers, with narrow CI (Table 4).
Table 3.
Interobserver agreement assessed with intraclass correlation (ICC) for
measurements of acetabular, femoral anteversion angle, true and functional
femoral offset, and true and functional global offset on 213 non-operated
hips and 71 operated hips in 71 patients.
Type
Measurement
ICC
95% CI
213 non-operated hips
Acetabular offset
0.94
0.88–0.96
Femoral anteversion angle
0.93
0.90–0.95
True femoral offset
0.94
0.92–0.96
True global offset
0.97
0.96–0.98
Functional femoral offset
0.96
0.94–0.97
Functional global offset
0.97
0.97–0.98
71 operated hips
Acetabular offset
0.97
0.96–0.98
Femoral anteversion angle
0.95
0.93–0.97
True femoral offset
0.96
0.94–0.98
True global offset
0.98
0.96–0.98
Functional femoral offset
0.97
0.95–0.98
Functional global offset
0.98
0.96–0.98
Table 4.
Intraobserver agreement assessed by intraclass correlation coefficient (ICC)
in pre- and postoperative computed tomo-graphy in 15 patients, i.e. totally
45 non-operated and 15 operated hips.
Type
Measurement
ICC
95 % CI
Observer 1, 45 non-operated hips
Acetabular offset
0.94
0.90–0.97
Femoral anteversion angle
0.97
0.95–0.98
True femoral offset
0.97
0.94–0.98
Functional femoral offset
0.98
0.96–0.99
15 operated hips
Acetabular offset
0.99
0.97–1.00
Femoral anteversion angle
0.95
0.85–0.98
True femoral offset
0.93
0.81–0.98
Functional femoral offset
0.96
0.88–0.98
All 60 hips
Acetabular offset
0.97
0.93–0.98
Femoral anteversion angle
0.97
0.95–0.98
True femoral offset
0.96
0.94–0.98
Functional femoral offset
0.97
0.96–0.98
Observer 2, 45 non-operated hips
Acetabular offset
0.96
0.93–0.98
Femoral anteversion angle
0.94
0.90–0.97
True femoral offset
0.93
0.87–0.96
Functional femoral offset
0.96
0.93–0.98
15 operated hips
Acetabular offset
0.97
0.92–0.99
Femoral anteversion angle
0.95
0.85–0.98
True femoral offset
0.92
0.77–0.97
Functional femoral offset
0.98
0.95–0.99
All 60 hips
Acetabular offset
0.97
0.94–0.98
Femoral anteversion angle
0.95
0.91–0.97
True femoral offset
0.93
0.87–0.96
Functional femoral offset
0.96
0.94–0.98
Interobserver agreement assessed with intraclass correlation (ICC) for
measurements of acetabular, femoral anteversion angle, true and functional
femoral offset, and true and functional global offset on 213 non-operated
hips and 71 operated hips in 71 patients.Intraobserver agreement assessed by intraclass correlation coefficient (ICC)
in pre- and postoperative computed tomo-graphy in 15 patients, i.e. totally
45 non-operated and 15 operated hips.Measurements for pre- and postoperative CT of the non-operated hips were compared to
evaluate the robustness of the method, i.e. to determine whether repeated CT
examinations and measurements on the same hip would yield comparable results. The
ICC scores were almost perfect for both observers, with narrow CI (Table 5), and linear
regression analyses showed a significant correlation for AO, true and functional FO,
and FNA (p < 0.001; Fig. 6).
Table 5.
Comparison of pre- and postoperative assessments in 71 non-operated hips by
two observers by intraclass correlation coefficient (ICC).
Type
Measurement
ICC
95% CI
Observer 1, 71 non-operated hips pre- vs. postoperative
examination
Acetabular offset
0.94
0.91–0.96
Femoral anteversion angle
0.93
0.84–0.96
True femoral offset
0.96
0.93–0.97
True global offset
0.99
0.98–0.99
Functional femoral offset
0.97
0.95–0.98
Functional global offset
0.98
0.97–0.99
Observer 2, 71 non-operated hips pre- vs. postoperative
examination
Acetabular offset
0.94
0.91–0.96
Femoral anteversion angle
0.92
0.88–0.95
True femoral offset
0.93
0.89–0.96
True global offset
0.97
0.96–0.98
Functional femoral offset
0.94
0.90–0.97
Functional global offset
0.97
0.94–0.98
Fig. 6.
Correlation plots for measurements of acetabular offset, femoral neck
anteversion, true and functional femoral offset for two observers in 213
non-operated hips.
Comparison of pre- and postoperative assessments in 71 non-operated hips by
two observers by intraclass correlation coefficient (ICC).Correlation plots for measurements of acetabular offset, femoral neck
anteversion, true and functional femoral offset for two observers in 213
non-operated hips.
Discussion
In the current study, measurements of AO, true GO, and functional GO showed expected
results with significant side-to-side differences between the non-affected and the
osteoarthritic hips. Postoperatively, AO had been reduced with normalization of true
GO and functional GO. The inter- and intraobserver agreements in measuring AO, true
FO, functional FO, and FNA with a new hip arthroplasty templating software were
generally near-perfect with narrow CI. Measurements on repeated CT examinations were
close to identical.Offset measurements have traditionally been done on AP pelvis or hip radiographs,[12] where femoral rotation[16] and flexion[17] have been shown to influence measurements. In the current study, also true FO
was measured, which is the perpendicular distance from the long axis of the femur to
the center of the femoral head, measured in 3D. This measurement is always longer
than the AP measurement of functional FO and nonaffected by patient positioning.Measurements of FNA differ between studies, depending on how the measurements were
made. Distally, four different methods may be used to define the femoral condylar
axis; a line drawn between the most posterior points of the condyles, a line drawn
between the most medial and lateral points on the condyles, a line drawn between the
center of the centroids of the condyles in cross-section,[36] or a line bisecting a line through the most posterior points of the femoral
condyles and a line through the most anterior points of the distal femur in the same section.[37] In the current study, the most posterior points of the femoral condyles were
used to define the condylar line.The method for measurements of FNA used in the current analysis is not identical to
the definition by Billing.[38] Billing defined anteversion based on the long axis of the femur as defined by
a point centrally in the distal femur. In the current study, anteversion was
measured relative to the long axis of the femur proximally, to better correspond to
the true insertion site and final position of a femoral prosthetic stem, and not
having to compensate for the physiological bowing of the femoral shaft at
templating. The proximal point for defining the long axis of the femur has also been
defined in different ways. Billing, despite a detailed description of the geometry
of the femur and the definition of FNA,[38] did not define this point. Murphy et al.[25] defined it as “a centroid of a cross section of the femur at the base of the
femoral neck” on an axial CT section “3 per cent” distal to the middle of the lesser
trochanter. In the current study, the condylar plane was defined by the condylar
line and the intersection of the proximal long axis line and the line for the true
FO measurement.When using conventional CT images, measurements are improved if the sections are
corrected for hip and knee flexion as well as adduction of the hip.[39] This has up until recently been problematic and one reason why previous
comparisons of radiography and CT measurements of FNA have not shown superiority for CT.[27]Since previous studies have used slightly different techniques and measuring points
for FNA, the comparison of measurements between studies is difficult (Fig. 7). The most crucial
issue, however, is to develop a measuring method with high reproducibility and low
observer variation such as in the current study. The high observer agreement rates
are consistent with other studies reporting on CT assessment of measurements using
3D images. In one study, reporting on FNA in children using CT and MRI,[28] ICC values were also near-perfect with narrow CI. In another study measuring
FNA directly on 3D reconstructions,[40] ICC values were high but slightly lower than in the current study. In the
current study, CT was performed using a low-dose technique, with an effective dose
close to that of radiography,[30] showing that even with increased image noise, excellent results can be
achieved. The results from the current study further support the use of 3D data
sets. With the use of 3D data sets, the need for exact patient positioning is
practically eliminated.
Fig. 7.
Femoral neck anteversion (FNA) is different depending on the choice of long
axis of the femur. (a) In the current study, the proximal part of the femur
was chosen for the long axis definition since that corresponds to the length
of the femoral stem in hip arthroplasty. FNA is 35°. (b) In previous
definitions, e.g. by Billing[38] and Murphy et al.,[25] the long axis was defined by a point in the knees and a point in the
proximal femur, giving a different measurement value of FNA, 20°.
Femoral neck anteversion (FNA) is different depending on the choice of long
axis of the femur. (a) In the current study, the proximal part of the femur
was chosen for the long axis definition since that corresponds to the length
of the femoral stem in hip arthroplasty. FNA is 35°. (b) In previous
definitions, e.g. by Billing[38] and Murphy et al.,[25] the long axis was defined by a point in the knees and a point in the
proximal femur, giving a different measurement value of FNA, 20°.Various levels of symmetry in normal (non-arthritic) populations have previously been
reported regarding femoral hip orientation measurements, seemingly dependent on the
evaluated populations. Previously, an asymmetry > 2% has been reported for the
femoral neck-shaft angle, FO, FNA, femoral length, and femoral head radius[8] with substantial differences reported for FNA, FO, and femoral head center location,[7] all being important measurements for contralateral templating of THA. On the
other hand, another study, using high-resolution photographs on cadaveric specimens,
reported no significant side-to-side differences for proximal femoral measurements
of the femoral head diameter, the minimal neck diameter, and the femoral shaft diameter.[10]In osteoarthritic hips, reactive bone formation in the acetabular socket often leads
to an increase in AO but does not affect FO or FNA. The surgical aim in the current
study was to decrease AO with a compensatory increase in FO to restore symmetry in
GO. Increasing FO has been suggested to improve abductor strength, reduce limping,
and counteract polyethylene wear.[41] In the current study GO symmetry was achieved according to this
principle.The limitation of the study is mainly the small number of observers. However, the
high ICC and narrow CI showed high inter- and intraobserver agreements. There was no
reference standard for the measurements, but due to the use of different reference
points for measurements in the literature, this was impossible to find.In conclusion, using low-dose CT with 3D measurements with a templating software
yielded excellent repeatability of measurements with near-perfect observer
agreement. The study supports the use of 3D data sets for measurements in the pre-
and postoperative evaluation in THA.
Authors: Kevin A Cassidy; Manish S Noticewala; William Macaulay; Jonathan H Lee; Jeffrey A Geller Journal: J Arthroplasty Date: 2012-07-17 Impact factor: 4.757
Authors: Markus Weber; Michael L Woerner; Hans-Robert Springorum; Alexander Hapfelmeier; Joachim Grifka; Tobias F Renkawitz Journal: J Arthroplasty Date: 2014-03-28 Impact factor: 4.757
Authors: Nick J Little; Constant A Busch; John A Gallagher; Cecil H Rorabeck; Robert B Bourne Journal: Clin Orthop Relat Res Date: 2009-05-02 Impact factor: 4.176