Felipe García-Pinillos1,2, Diego Jaén-Carrillo3, Victor Soto Hermoso1, Pedro Latorre Román4, Pedro Delgado2, Antonio Carton3, Cristian Martinez2, Luis Roche Seruendo3.
Abstract
BACKGROUND: Markerless systems to capture body motion require no markers to be attached to the body, thereby improving clinical feasibility and testing time. However, the lack of markers might affect the accuracy of measurements.Entities:
Keywords: Kinect; OptoGait; accuracy; feasibility; gait; motion capture; runners; sport technology; validity; video
Mesh:
Year: 2020 PMID: 33095181 PMCID: PMC7647810 DOI: 10.2196/19498
Source DB: PubMed Journal: JMIR Mhealth Uhealth ISSN: 2291-5222 Impact factor: 4.773
Descriptive data of spatiotemporal parameters obtained from different systems (ie, Kinect with the MotionMetrix software, OptoGait, and high-speed video analysis at 1000 Hz).
| Variable | OGa, mean (SD) | VAb, mean (SD) | MMc, mean (SD) | MM-OG difference, mean (%)d | MM-VA difference, mean (%) |
| Contact tine (s) | 0.265 (0.015) | 0.262 (0.013) | 0.273 (0.021) | 0.008 (3.0) | 0.011 (4.2) |
| Flight time (s) | 0.089 (0.018) | 0.097 (0.017) | 0.082 (0.019) | −0.007 (−7.9) | −0.011 (−11.3) |
| Step frequency (spme) | 166.60 (6.80) | 165.76 (6.90) | 167.23 (7.28) | 0.63 (0.4) | 1.47 (0.9) |
| Step length (cm) | 115.53 (6.80) | 116.58 (5.92) | 115.44 (5.26) | −0.09 (−0.1) | −1.14 (−1.0) |
aOG: OptoGait system.
bVA: High-speed video analysis.
cMM: Kinect system with the MotionMetrix software.
d%: Refers to values reported by the reference system.
espm: Steps per minute.
Coefficients of variation (%) of spatiotemporal parameters during running at 12 km/h obtained from three different systems.
| Variable | OGa | VAb | MMc |
| Contact time, CVd | 5.7 | 5.2 | 8.1 |
| Flight time, CV | 16.4 | 17.3 | 23.2 |
| Step frequency, CV | 4.1 | 4.2 | 4.4 |
| Step length, CV | 4.6 | 4.6 | 5.02 |
aOG: OptoGait system.
bVA: High-speed video analysis.
cMM: Kinect system with the MotionMetrix software.
dCV: Coefficient of variation.
Pearson correlation analysis and intraclass correlation coefficients between spatiotemporal parameters obtained from the Kinect system with the MotionMetrix software and those obtained from two different systems (ie, OptoGait and high-speed video analysis) during running at a comfortable speed.
| Variables | MMa versus OGb | MM versus VAc | ||||
|
|
| ICCd (95% CI) |
| ICC (95% CI) | ||
| Contact time | 0.645 | <.001 | 0.712 (0.333-0.874) | 0.664 | <.001 | 0.667 (0.153-0.862) |
| Flight time | 0.901 | <.001 | 0.894 (0.420-0.967) | 0.928 | <.001 | 0.838 (0.159-0.961) |
| Step frequency | 0.962 | <.001 | 0.978 (0.951-0.990) | 0.940 | <.001 | 0.959 (0.873-0.984) |
| Step length | 0.971 | <.001 | 0.986 (0.968-0.994) | 0.954 | <.001 | 0.964 (0.881-0.986) |
aMM: Kinect system with the MotionMetrix software.
bOG: OptoGait system.
cVA: High-speed video analysis.
dICC: intraclass correlation coefficient.
Figure 1Bland-Altman plots for the measurement of spatiotemporal parameters (ie, from up to bottom: contact time, flight time, step frequency, and step length) during running at a comfortable speed obtained from Kinect with MotionMetrix software and two different systems (OptoGait and high-speed video analysis at 1000 Hz). The plot includes the mean difference (dotted line) and 95% limits of agreement (dashed line), along with the regression line (solid line).