| Literature DB >> 25017613 |
Myungmo Lee1, Changho Song1, Kyoungjin Lee1, Doochul Shin1, Seungho Shin1.
Abstract
BACKGROUND: Treadmill gait analysis was more advantageous than over-ground walking because it allowed continuous measurements of the gait parameters. The purpose of this study was to investigate the concurrent validity and the test-retest reliability of the OPTOGait photoelectric cell system against the treadmill-based gait analysis system by assessing spatio-temporal gait parameters.Entities:
Mesh:
Year: 2014 PMID: 25017613 PMCID: PMC4106927 DOI: 10.12659/MSM.890658
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 1Lateral view of the 2 measuring systems. The transmitting bar and receiving bar of the OPTOGait system were placed on the side bars of the Zebris treadmill.
Mean (S.D.) and level of agreement of the gait parameters for subjects using the OPTOGait system and Zebris treadmill system.
| Gait parameters | Young subjects | Stroke patients | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| OPTOGait | Zebris treadmill | ICC (95%CI) | CV% | 95%LOA | OPTOGait | Zebris treadmill | ICC (95%CI) | CV% | 95%LOA | |
| Speed (m/s) | 0.85 (0.08) | 0.85 (0.08) | 0.995 (0.990~0.998) | 0.63 | −0.012~0.014 | 0.71 (0.09) | 0.72 (0.08) | 0.908 (0.844~0.946) | 3.83 | −0.033~0.026 |
| Cadence (steps/m) | 94.93 (8.67) | 94.75 (8.75) | 0.997 (0.994~0.998) | 0.50 | −1.087~0.665 | 97.60 (10.98) | 97.63 (11.27) | 0.988 (0.980~0.993) | 1.23 | −1.987~1.820 |
| Gait cycle (s) | 1.28 (0.12) | 1.28 (0.12) | 0.999 (0.998~1.000) | 0.27 | −0.006~0.012 | 1.24 (0.14) | 1.25 (0.15) | 0.990 (0.983~0.995) | 1.15 | −0.023~0.039 |
| Lt. step (cm) | 53.42 (3.24) | 53.53 (3.33) | 0.958 (0.920~0.978) | 1.25 | −1.362~1.807 | 44.76 (5.73) | 44.50 (5.58) | 0.975 (0.957~0.986) | 1.99 | −2.542~1.772 |
| Rt. step (cm) | 53.66 (3.26) | 53.58 (3.25) | 0.932 (0.870~0.965) | 1.59 | −2.406~1.962 | 45.90 (5.18) | 45.53 (4.86) | 0.947 (0.910~0.969) | 2.42 | −3.193~1.943 |
| Lt. step time (s) | 0.64 (0.06) | 0.64 (0.06) | 0.990 (0.981~0.995) | 0.96 | −0.008~0.018 | 0.61 (0.08) | 0.62 (0.08) | 0.958 (0.928~0.976) | 2.67 | −0.019~0.027 |
| Rt. step time (s) | 0.64 (0.06) | 0.64 (0.06) | 0.985 (0.971~0.992) | 1.14 | −0.020~0.015 | 0.61 (0.08) | 0.62 (0.08) | 0.852 (0.756~0.912) | 5.01 | −0.012~0.027 |
| Stride (cm) | 106.97 (5.54) | 107.00 (6.12) | 0.969 (0.939~0.984) | 0.97 | −1.986~2.319 | 90.64 (9.01) | 89.98 (8.69) | 0.976 (0.959~0.986) | 1.51 | −5.017~2.575 |
| Lt. SLS (%) | 31.39 (1.69) | 34.82 (1.50)*** | 0.370 (0.052~0.620) | 3.83 | −0.165~7.143 | 29.79 (2.25) | 33.20 (2.00)*** | 0.297 (0.029~0.525) | 5.68 | −1.378~7.987 |
| Rt. SLS (%) | 31.44 (1.64) | 34.76 (1.36)*** | 0.238 (−0.099~0.526) | 3.92 | 0.026~7.074 | 29.63 (2.27) | 33.00 (2.39)*** | 0.098 (−0.177~0.359) | 7.07 | −2.811~9.386 |
| TDLS (%) | 37.19 (3.21) | 30.39 (2.60)*** | 0.310 (−0.016~0.577) | 7.18 | −13.975~−0.158 | 40.64 (3.91) | 33.80 (2.41)*** | −0.294 (−0.523~−0.026) | 10.05 | −16.894~3.571 |
| Lt. swing phase (%) | 31.44 (1.62) | 34.84 (1.44)*** | 0.243 (−0.088~0.526) | 4.02 | −0.047~7.147 | 29.98 (2.35) | 32.92 (2.31)*** | 0.042 (−0.232~0.309) | 7.20 | −3.596~9.061 |
| Rt. swing phase (%) | 31.38 (1.70) | 34.77 (1.43)*** | 0.378 (0.061~0.626) | 3.75 | −0.097~7.109 | 30.07 (2.32) | 33.31 (2.06)*** | 0.315 (0.048~0.539) | 5.78 | −1.358~8.206 |
| Lt. stance phase (%) | 68.56 (1.62) | 65.23 (1.37)*** | 0.239 (−0.093~0.523) | 1.96 | −7.165~0.077 | 70.02 (2.35) | 67.09 (2.32)*** | 0.041 (−0.232~0.308) | 3.32 | −9.063~3.605 |
| Rt. stance phase (%) | 68.36 (2.15) | 65.18 (1.49)*** | 0.468 (0.169~0.688) | 2.02 | −7.139~0.627 | 69.93 (2.32) | 66.68 (2.05)*** | 0.319 (0.053~0.543) | 2.66 | −8.000~1.344 |
ICC– intra correlation coefficient; CVME – coefficients of variation of method error; LOA – limits of agreement; SLS – single limb support; TDLS – total double limb support. Significant difference between the two measuring instruments (***, p≤0.001).
Figure 2Relationship between OPTOGait system and Zebris treadmill system for main spatio-temporal gait parameters in healthy young adults.
Figure 3Relationship between OPTOGait system and Zebris treadmill system for main spatio-temporal gait parameters in stroke patients.
Mean (S.D.) and test-retest reliability of the gait parameters for the two measurement sessions in healthy young adults.
| Gait parameters | Session 1 | Session 2 | ICC (95% CI) | 95% LOA | CV % | SEM % | MDC % |
|---|---|---|---|---|---|---|---|
| Speed (m/s) | 0.84 (0.07) | 0.85 (0.07) | 0.978 (0.943–0.992) | −0.026~0.037 | 1.35 | 2.03 | 5.63 |
| Cadence (steps/m) | 95.13 (8.35) | 94.73 (9.21) | 0.962 (0.901–0.948) | −5.173~4.373 | 1.81 | 0.37 | 1.02 |
| Gait cycle (s) | 1.27 (0.11) | 1.27 (0.13) | 0.961 (0.900–0.985) | −0.057~0.073 | 1.84 | 0.40 | 1.11 |
| Lt. step (cm) | 53.44 (2.91) | 53.39 (3.42) | 0.805 (0.668–0.929) | −5.101~4.324 | 3.19 | 1.89 | 5.24 |
| Rt. step (cm) | 54.22 (3.84) | 54.55 (3.61) | 0.835 (0.613–0.935) | −3.865~4.532 | 2.78 | 1.17 | 3.23 |
| Lt. step time (s) | 0.64 (0.06) | 0.64 (0.07) | 0.920 (0.799–0.969) | −0.049~0.050 | 2.79 | 0.88 | 2.43 |
| Rt. step time (s) | 0.64 (0.07) | 0.64 (0.06) | 0.940 (0.848–0.977) | −0.050~0.043 | 2.58 | 0.66 | 1.82 |
| Stride (cm) | 106.28 (4.46) | 107.66 (6.49) | 0.812 (0.566–0.925) | −5.035~8.083 | 2.25 | 1.14 | 3.16 |
| Lt. SLS (%) | 31.31 (1.52) | 31.47 (1.89) | 0.873 (0.694–0.951) | −1.527~1.861 | 1.94 | 0.76 | 2.12 |
| Rt. SLS (%) | 31.15 (1.50) | 31.74 (1.71) | 0.857 (0.658–0.944) | −1.100~2.277 | 1.93 | 0.78 | 2.15 |
| TDLS (%) | 37.56 (2.93) | 36.81 (3.52) | 0.911 (0.778–0.966) | −3.430~1.930 | 2.60 | 0.84 | 2.34 |
| Lt. swing phase (%) | 31.17 (1.51) | 31.72 (1.72) | 0.855 (0.655–0.943) | −1.151~2.262 | 1.95 | 0.79 | 2.20 |
| Rt. swing phase (%) | 31.29 (1.53) | 31.46 (1.89) | 0.872 (0.691–0.950) | −1.548~1.870 | 1.96 | 0.77 | 2.14 |
| Lt. stance phase (%) | 68.84 (1.52) | 68.28 (1.71) | 0.856 (0.657–0.944) | −2.265~1.142 | 0.89 | 0.36 | 1.00 |
| Rt. stance phase (%) | 68.72 (1.52) | 68.28 (2.31) | 0.774 (0.492–0.909) | −3.018~2.140 | 1.35 | 0.76 | 2.11 |
ICC – intra correlation coefficient; LOA – limits of agreement; CVME – coefficients of variation of method error; SEM – standard error of measurement; MDC – minimum detectable change; SLS – single limb support; TDLS – total double limb support.
Mean (S.D.) and test-retest reliability of the gait parameters for the two measurement sessions in stroke patients.
| Gait parameters | Session 1 | Session 2 | ICC (95% CI) | 95% LOA | CV % | SEM % | MDC % |
|---|---|---|---|---|---|---|---|
| Speed (m/s) | 0.71 (0.10) | 0.71 (0.10) | 0.966 (0.926–0.985) | −0.049~0.051 | 2.97 | 0.47 | 1.31 |
| Cadence (steps/m) | 97.65 (11.36) | 97.56 (10.81) | 0.970 (0.935–0.986) | −5.408~5.224 | 2.27 | 0.35 | 0.97 |
| Gait cycle (s) | 1.23 (0.15) | 1.23 (0.14) | 0.962 (0.918–0.983) | −0.080~0.076 | 2.52 | 0.46 | 1.28 |
| Lt. step (cm) | 44.56 (5.76) | 44.96 (5.80) | 0.985 (0.968–0.996) | −1.537~2.344 | 1.63 | 0.19 | 0.54 |
| Rt. step (cm) | 45.99 (5.35) | 45.81 (5.11) | 0.929 (0.848–0.967) | −4.060~3.694 | 3.18 | 0.83 | 2.29 |
| Lt. step time (s) | 0.61 (0.07) | 0.61 (0.08) | 0.911 (0.812–0.959) | −0.067~0.066 | 3.54 | 1.17 | 3.24 |
| Rt. step time (s) | 0.61 (0.08) | 0.61 (0.08) | 0.926 (0.843–0.966) | −0.063~0.064 | 3.56 | 0.97 | 2.69 |
| Stride (cm) | 90.45 (9.28) | 90.83 (8.90) | 0.973 (0.941–0.988) | −3.750~4.520 | 1.86 | 0.28 | 0.77 |
| Lt. SLS (%) | 29.73 (2.34) | 29.85 (2.20) | 0.901 (0.792–0.954) | −1.855~2.117 | 2.61 | 0.78 | 2.16 |
| Rt. SLS (%) | 29.55 (2.26) | 29.71 (2.31) | 0.949 (0.890–0.977) | −1.279~1.586 | 1.78 | 0.40 | 1.10 |
| TDLS (%) | 40.74 (4.12) | 40.52 (3.89) | 0.957 (0.907–0.980) | −2.529~2.083 | 2.49 | 0.44 | 1.21 |
| Lt. swing phase (%) | 30.01 (2.23) | 30.12 (2.44) | 0.920 (0.830–0.963) | −1.730~1.940 | 2.21 | 0.65 | 1.80 |
| Rt. swing phase (%) | 29.91 (2.35) | 30.04 (2.39) | 0.854 (0.701–0.932) | −2.378~2.647 | 3.25 | 1.16 | 3.23 |
| Lt. stance phase (%) | 69.98 (2.23) | 69.87 (2.44) | 0.920 (0.830–0.963) | −1.940~1.730 | 0.97 | 0.28 | 0.77 |
| Rt. stance phase (%) | 70.09 (2.34) | 69.95 (2.39) | 0.854 (0.701–0.932) | −2.647~2.378 | 1.41 | 0.50 | 1.38 |
ICC – intra correlation coefficient; LOA – limits of agreement; CVME – coefficients of variation of method error; SEM – standard error of measurement; MDC – minimum detectable change; SLS – single limb support; TDLS – total double limb support.
Figure 4Agreement for the main spatial and temporal gait parameters between session 1 and session 2 of OPTOGait system in healthy young adults.
Figure 5Agreement for the main spatial and temporal gait parameters between session 1 and session 2 of OPTOGait system in stroke patients.