| Literature DB >> 33058408 |
Shunsuke Ono1, Yoshihiro Ueda1, Shingo Ohira1, Masaru Isono1, Iori Sumida2, Shoki Inui1, Masahiro Morimoto1, Reiko Ashida3, Masayoshi Miyazaki1, Kazuhiko Ogawa2, Teruki Teshima1.
Abstract
PURPOSE: To investigate the detectability of fiducial markers' positions for real-time target tracking system equipping with a standard linac. The hypothesis is that the detectability depends on the type of fiducial marker and the gantry angle of acquired triggered images.Entities:
Keywords: fiducial marker; real-time target tracking; respiratory management; triggered image
Mesh:
Year: 2020 PMID: 33058408 PMCID: PMC7700931 DOI: 10.1002/acm2.13050
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.243
Fig. 1Nine types of fiducials embed in water‐equivalent phantom. The name, diameter as the interval between blue lines in mm, and length as the interval between red lines in mm for each fiducial are summarized in Table 1. (a) iGold (Medikit, Tokyo, Japan), 2, (b) Acculoc (Civco Medical Solutions, Iowa, USA), 1, 3. (c) VISICOIL (IBA, Bartlett, TN), 0.5, 5. (d) GoldAnchor of length in 10 mm, straight type (Naslund Medical AB, Huddinge, Sweden), 0.3, 10. (e) GoldAnchor of length in 10 mm, folded type (Naslund Medical AB), 2, 2. (f) GoldAnchor length in 20 mm, straight type (Naslund Medical AB), 0.3, 20. (g) GoldAnchor length in 20 mm, folded type (Naslund Medical AB), 2, 2. (h) LumiCoil straight type, (Boston Scientific, Boston, MA), 0.5, 5. (i) LumiCoil form 8 type, (Boston Scientific), 2, 5.
The form, measured, and set diameters and length in ABH for each fiducial.
| Fiducial | Form | Measurement [mm] | Marker setting in ABH [mm] | ||
|---|---|---|---|---|---|
| Plate_A/A45 | Plate_B | Plate_B | |||
| Dia./Leng. | Dia./Leng. | Dia./Leng. | Dia./Leng. | ||
| iGold | Ball | 2.0/2.0 | 2.0/2.0 | ||
| Acculoc | Slim | 1.0/3.0 | 1.0/3.0 | 1.0/3.0 | |
| VISICOIL | Slim | 0.5/5.0 | 1.3/5.0 | 1.3/5.0 | 1.0/3.0 |
| GA 10 mm | Slim | 0.35/10 | |||
| GA 10 mm | Folded | 2.0/2.0 | 2.0/2.0 | ||
| GA 20 mm | Slim | 0.35/20 | |||
| GA 20 mm | Folded | 2.0/2.0 | 2.0/2.0 | ||
| LC | Slim | 0.45/5.0 | 1.3/5.0 | 1.3/5.0 | 1.0/3.0 |
| LC | Slim (Form 8) | 2.0/5.0 | 2.5/5.0 | 2.5/5.0 | 1.0/3.0 |
Abbreviations: Dia., Diameter; Leng., Length; ABH, auto beam holding; GA, GoldAnchor; LC, LumiCoil.
Fig. 2(a) Setting for fiducials. All fiducials turn to the gantry side. The distance of centers for fiducials was 3.16 cm (). (b) Experiment view. The I'mRT phantom and QUASAR platform were prepared on the treatment couch. The isocenter and the center of I'mRT phantom coincided. Infrared reflective marker was set on the QUASAR platform to use the respiratory management system of TrueBeam STx.
Fig. 3Shift of fiducial and triggered image. The red point represents isocenter, and the orange point the shift of fiducial. The yellow shade represents the kV x‐ray direction to the detector. In the triggered image, the green circle represents the tolerance of fiducial position error, the red point represents isocenter, and the orange square represents fiducial position. When fiducials move in the superior–inferior direction, the fiducials on the triggered image at the direction move vertically (a), when they move anterior‐posterior and right‐left, those on the triggered image move horizontally (b, c).
Fig. 4Comparisons between expected and detected horizontal positions for GoldAnchor, when setup error occurs in anterior direction (a) and left direction (d). The blue lines represent the expected horizontal positions, and the orange dots represent detected horizontal positions in the triggered images. Comparisons between expected and detected horizontal positions for VISICOIL in Plate_A, when setup error occurs in anterior direction (b) and left direction (e). Comparisons between expected and detected horizontal positions for VISICOIL in Plate_B, when setup error occurs in anterior direction (c) and left direction (f).
The mean value, standard deviation, and maximum values of differences in mm between the expected and detected positions in the triggered images when setup errors occurred for Plate_A in anterior, left, and superior directions.
| Dir. | SE | Makers | ||||||
|---|---|---|---|---|---|---|---|---|
| iGold | Acculoc | VISICOIL | GA | GA | LC | LC | ||
| Shift | 10 mm | 20 mm | Slim | Form 8 | ||||
| Ant. | 5 mm |
0.1 ± 0.1 (0.3) |
0.3 ± 0.2 (0.4) |
0.1 ± 0.2 (0.1) |
0.0 ± 0.1 (0.1) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.3) |
| 3 mm |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.1) |
0.0 ± 0.1 (0.2) |
0.0 ± 0.0 (0.1) |
0.2 ± 0.2 (0.4) |
0.1 ± 0.1 (0.4) | |
| 1 mm |
0.0 ± 0.0 (0.1) |
0.0 ± 0.1 (0.2) |
0.0 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.3) | |
| Lt. | 5 mm |
0.1 ± 0.2 (0.3) |
0.0 ± 0.2 (0.3) |
0.0 ± 0.1 (0.2) |
0.0 ± 0.1 (0.3) |
0.1 ± 0.1 (0.1) |
0.2 ± 0.2 (0.4) |
0.1 ± 0.2 (0.3) |
| 3 mm |
0.0 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.2) | |
| 1 mm |
0.0 ± 0.0 (0.1) |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.0 ± 0.0 (0.1) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.3) | |
| Sup. | 5 mm |
0.1 ± 0.2 (0.3) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.2 ± 0.1 (0.3) |
0.0 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.4) |
| 3 mm |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.2) | |
| 1 mm |
0.1 ± 0.1 (0.1) |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.1) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) | |
Abbreviations: SE, setup errors; GA, GoldAnchor; LC, LumiCoil; Dir., Directions; Ant., anterior; Lt., left; Sup., superior.
The mean value, standard deviation, and maximum value of differences in mm between the expected and detected positions in the triggered images when setup errors occurred for Plate_B and Plate_A45 in anterior, left, and superior directions.
| Dir. | SE | Plate_B | Plate_ A45 | |||
|---|---|---|---|---|---|---|
| Markers | ||||||
| Acculoc | VISICOIL | LC | LC | VISICOIL | ||
| Shift | Slim | Form 8 | ||||
| Ant. | 5 mm |
0.3 ± 0.2 (0.4) |
0.0 ± 0.5 (0.9) |
0.9 ± 1.1 (2.2) |
0.4 ± 0.8 (3.2) |
0.0 ± 0.1 (0.2) |
| 3 mm |
0.3 ± 0.4 (0.9) |
0.7 ± 0.9 (2.1) |
0.7 ± 0.9 (2.4) |
0.7 ± 1.1 (3.2) |
0.0 ± 0.1 (0.2) | |
| 1 mm |
0.1 ± 0.1 (0.3) |
0.1 ± 1.0 (2.7) |
0.3 ± 0.5 (1.5) |
0.5 ± 0.9 (2.8) |
0.0 ± 0.1 (0.3) | |
| Lt. | 5 mm |
0.0 ± 0.2 (0.3) |
0.4 ± 0.7 (1.7) |
0.7 ± 0.9 (3.1) |
0.3 ± 0.5 (2.9) |
0.0 ± 0.1 (0.3) |
| 3 mm |
0.4 ± 0.5 (1.3) |
0.7 ± 0.9 (2.8) |
0.6 ± 0.9 (3.0) |
0.5 ± 0.8 (3.0) |
0.0 ± 0.1 (0.2) | |
| 1 mm |
0.1 ± 0.1 (0.3) |
0.4 ± 1.0 (2.7) |
0.4 ± 0.5 (1.5) |
0.3 ± 0.6 (1.5) |
0.0 ± 0.1 (0.3) | |
| Sup. | 5 mm |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.1) |
0.1 ± 0.1 (0.1) |
0.1 ± 0.2 (0.8) |
0.1 ± 0.1 (0.2) |
| 3 mm |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.1 ± 0.2 (0.7) |
0.1 ± 0.1 (0.3) | |
| 1 mm |
0.1 ± 0.1 (0.2) |
0.1 ± 0.1 (0.2) |
0.0 ± 0.1 (0.2) |
0.1 ± 0.1 (0.3) |
0.2 ± 0.1 (0.4) |
Abbreviations: SE, setup errors; LC, LumiCoil; Dir., Directions; Ant., anterior; Lt, left; Sup., superior.
P‐values for each difference in fiducial type.
|
| |
|---|---|
| GoldAnchor vs. VISICOIL for Plate_A | .133 |
| Plate_A vs. Plate_B for Acculoc | <.001 |
| Plate_A vs. Plate_B for VISICOIL | <.001 |
| Plate_A vs. Plate_A45 for VISICOIL | .614 |
| Appropriate setting vs. inappropriate setting | <.001 |
Fig. 5Comparisons between the expected and detected horizontal positions for VISICOIL in Plate_A45 when setup error occurs in anterior direction (a) and left direction (b). The blue lines represent the expected horizontal positions and the orange dots represent detected horizontal positions in the triggered images.
Fig. 6Detected vertical positions in the appropriate (a) and inappropriate settings (b) when VISICOIL was used in the plate for Plate_A. The appropriate setting was 5 mm in length in the ABH system. The inappropriate setting was 3 mm in length in the ABH system. There were no setup errors in any direction. The blue dots were of the vertical position in the triggered images detected in the ABH system.