Organ deformation and dose coverage in robotic respiratory-tracking radiotherapy.

Xing-Qi Lu1, Lakshmi Narayan Shanmugham, Anand Mahadevan, Elena Nedea, Mary Ann Stevenson, Irving Kaplan, Eric T Wong, Salvatore La Rosa, Frank Wang, Stuart M Berman.   

Abstract

PURPOSE: Respiratory motion presents a significant challenge in stereotactic body radiosurgery. Respiratory tracking that follows the translational movement of the internal fiducials minimizes the uncertainties in dose delivery. However, the effect of deformation, defined as any changes in the body and organs relative to the center of fiducials, remains unanswered. This study investigated this problem and a possible solution. METHODS AND MATERIALS: Dose delivery using a robotic respiratory-tracking system was studied with clinical data. Each treatment plan was designed with the computed tomography scan in the end-expiration phase. The planned beams were applied to the computed tomography scan in end-inspiration following the shift of the fiducials. The dose coverage was compared with the initial plan, and the uncertainty due to the deformation was estimated. A necessary margin from the clinical target volume to the planning target volume was determined to account for this and other sources of uncertainty.
RESULTS: We studied 12 lung and 5 upper abdomen lesions. Our results demonstrated that for lung patients with properly implanted fiducials a 3-mm margin is required to compensate for the deformation and a 5-mm margin is required to compensate for all uncertainties. Our results for the upper abdomen tumors were still preliminary but indicated a similar result, although a larger margin might be required.
CONCLUSION: The effect of body deformation was studied. We found that adequate dose coverage for lung tumors can be ensured with proper fiducial placement and a 5-mm planning target volume margin. This approach is more practical and effective than a recent proposal to combine four-dimensional planning with respiratory tracking.

Entities:  

Mesh:

Substances:

Year:  2008        PMID: 18406892     DOI: 10.1016/j.ijrobp.2007.12.042

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  15 in total

1.  Correlation and prediction uncertainties in the cyberknife synchrony respiratory tracking system.

Authors:  Eric W Pepin; Huanmei Wu; Yuenian Zhang; Bryce Lord
Journal:  Med Phys       Date:  2011-07       Impact factor: 4.071

2.  Evaluation of reproducibility of tumor repositioning during multiple breathing cycles for liver stereotactic body radiotherapy treatment.

Authors:  Ludovic Bedos; Olivier Riou; Norbert Aillères; Antoine Braccini; Jessica Molinier; Carmen Llacer Moscardo; David Azria; Pascal Fenoglietto
Journal:  Rep Pract Oncol Radiother       Date:  2016-11-12

Review 3.  Improving radiotherapy planning, delivery accuracy, and normal tissue sparing using cutting edge technologies.

Authors:  Carri K Glide-Hurst; Indrin J Chetty
Journal:  J Thorac Dis       Date:  2014-04       Impact factor: 2.895

4.  Mediastinal surgery in connective tissue tunnels using flexible endoscopy.

Authors:  G O Spaun; C M Dunst; D V Martinec; B N Arnold; M Owens; L L Swanstrom
Journal:  Surg Endosc       Date:  2010-02-23       Impact factor: 4.584

5.  Toward the development of intrafraction tumor deformation tracking using a dynamic multi-leaf collimator.

Authors:  Yuanyuan Ge; Ricky T O'Brien; Chun-Chien Shieh; Jeremy T Booth; Paul J Keall
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

6.  Comparison of 3D and 4D Monte Carlo optimization in robotic tracking stereotactic body radiotherapy of lung cancer.

Authors:  Mark K H Chan; Rene Werner; Miriam Ayadi; Oliver Blanck
Journal:  Strahlenther Onkol       Date:  2014-09-20       Impact factor: 4.033

7.  Using 4D dose accumulation to calculate organ-at-risk dose deviations from motion-synchronized liver and lung tomotherapy treatments.

Authors:  William S Ferris; Edward H Chao; Jennifer B Smilowitz; Randall J Kimple; John E Bayouth; Wesley S Culberson
Journal:  J Appl Clin Med Phys       Date:  2022-04-29       Impact factor: 2.243

8.  Accuracy and sensitivity of four-dimensional dose calculation to systematic motion variability in stereotatic body radiotherapy (SBRT) for lung cancer.

Authors:  Mark K H Chan; Dora L W Kwong; Sherry C Y Ng; Anthony S M Tong; Eric K W Tam
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

9.  Dosimetric evaluation of four-dimensional dose distributions of CyberKnife and volumetric-modulated arc radiotherapy in stereotactic body lung radiotherapy.

Authors:  Mark K H Chan; Dora L W Kwong; Gilbert M L Law; Eric Tam; Anthony Tong; Venus Lee; Sherry C Y Ng
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

10.  Retrospective evaluation of CTV to PTV margins using CyberKnife in patients with thoracic tumors.

Authors:  Alejandro Floriano; Rafael García; Ramon Moreno; Alberto Sánchez-Reyes
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.