Ramadhan Hardani Putra1,2, Nobuhiro Yoda3, Masahiro Iikubo4, Yoshihiro Kataoka5, Kensuke Yamauchi5, Shigeto Koyama6, Upul Cooray7, Eha Renwi Astuti2, Tetsu Takahashi5, Keiichi Sasaki1.
Abstract
BACKGROUND: The impact of the jaw bone condition, such as bone quantity and quality in the implant placement site, affecting the accuracy of implant placement with computer-guided surgery (CGS) remains unclear. Therefore, this study aimed to evaluate the influence of bone condition, i.e., bone density, bone width, and cortical bone thickness at the crestal bone on the accuracy of implant placement with CGS.Entities:
Keywords: Computed tomography; Computer-assisted surgery; Dental implant; Mandible
Year: 2020 PMID: 32951152 PMCID: PMC7502099 DOI: 10.1186/s40729-020-00249-z
Source DB: PubMed Journal: Int J Implant Dent ISSN: 2198-4034
Fig. 1Bone condition analysis. a Bone density measurement. Red box indicates the measured area with an implant outer area as a reference. b Bone width measurement. Red bar indicates the measured area at 1 mm under the crest of alveolar bone. c Cortical bone thickness measurement. Red bar indicates the measured area parallel to the dental implant axis
Categories of bone condition measured in the implant site area
| Bone condition | Criteria |
|---|---|
| BD1 | < 500 HU |
| BD2 | > 500 HU |
| BW1 | < 6 mm |
| BW2 | > 6 mm |
| CBT1 | < 1.5 mm |
| CBT2 | > 1.5 mm |
HU Hounsfield units
Fig. 2Assessment of placed implant position. a Matching digital data between preoperative planned computed tomography image (gray-colored model) and scanned postoperative impression (pink-colored model). b Assessing placed implant position using implant collar and guide pin as references. c Comparing planned (yellow-colored) and placed (red-colored) implant position
Fig. 3Parameters for implant deviation measurement. VP, vertical deviation at implant platform (mm); VA, vertical deviation at implant apex (mm); LP, linear deviation at implant platform; LA, linear deviation at implant apex (mm)
Fig. 4Bucco-lingual section plane (red line) and mesio-distal section plane (blue line)
Distribution of implant according to the dimension and region of insertion
| Total implants | |
|---|---|
| 3.3 mm × 10 mm | 2 |
| 4.1 mm × 6 mm | 2 |
| 4.1 mm × 8 mm | 7 |
| 4.1 mm × 10 mm | 25 |
| 4.1 mm × 12 mm | 5 |
| 4.8 mm × 8 mm | 1 |
| 4.8 mm × 10 mm | 4 |
| 4.8 mm × 12 mm | 1 |
| Premolar region | 14 |
| Molar region | 33 |
Descriptive statistics of implant placement accuracy parameters on each bone condition groups
| Bone density | Bone width | Cortical bone thickness | ||||
|---|---|---|---|---|---|---|
| BD1 | BD2 | BW1 | BW2 | CT1 | CT2 | |
| 24 | 23 | 36 | 11 | 33 | 14 | |
| Mean | 1.63 | 1.30 | 1.56 | 1.16 | 1.53 | 1.32 |
| Median | 1.43 | 1.12 | 1.43 | 0.92 | 1.35 | 1.11 |
| SD | 0.84 | 0.71 | 0.75 | 0.86 | 0.84 | 0.66 |
| Min | 0.56 | 0.27 | 0.42 | 0.27 | 0.27 | 0.55 |
| Max | 3.32 | 2.66 | 3.32 | 2.98 | 3.32 | 2.31 |
| Mean | 1.94 | 1.56 | 1.88 | 1.34 | 1.46 | 1.32 |
| Median | 1.69 | 1.63 | 1.86 | 1.17 | 1.30 | 1.43 |
| SD | 0.95 | 0.83 | 0.90 | 0.82 | 0.87 | 0.63 |
| Min | 0.53 | 0.25 | 0.25 | 0.42 | 0.24 | 0.3 |
| Max | 4.23 | 3.01 | 4.23 | 2.9 | 3.24 | 2.32 |
| Mean | 1.63 | 1.19 | 1.52 | 1.08 | 1.84 | 1.54 |
| Median | 1.44 | 1.30 | 1.45 | 0.71 | 1.63 | 1.67 |
| SD | 0.84 | 0.71 | 0.75 | 0.91 | 0.94 | 0.82 |
| Min | 0.38 | 0.24 | 0.24 | 0.38 | 0.26 | 0.25 |
| Max | 3.24 | 2.65 | 3.24 | 2.93 | 4.23 | 2.82 |
| Mean | 2.40 | 1.39 | 1.93 | 1.82 | 1.98 | 1.72 |
| Median | 2.19 | 1.32 | 1.82 | 1.14 | 1.70 | 1.83 |
| SD | 0.94 | 0.74 | 0.91 | 1.22 | 1.03 | 0.86 |
| Min | 0.95 | 0.45 | 0.45 | 0.5 | 0.45 | 0.48 |
| Max | 4.04 | 3.41 | 4.04 | 3.67 | 4.04 | 3.34 |
| Mean | 1.50 | 0.96 | 1.34 | 0.88 | 1.31 | 1.06 |
| Median | 1.29 | 0.81 | 1.30 | 0.53 | 1.20 | 0.80 |
| SD | 0.85 | 0.78 | 0.81 | 0.94 | 0.92 | 0.67 |
| Min | 0.42 | 0.08 | 0.08 | 0.08 | 0.08 | 0.25 |
| Max | 3.25 | 2.65 | 3.25 | 2.79 | 3.25 | 2.25 |
| Mean | 1.68 | 0.96 | 1.43 | 0.98 | 1.43 | 1.09 |
| Median | 1.39 | 0.86 | 1.39 | 0.50 | 1.36 | 0.93 |
| SD | 0.90 | 0.84 | 0.88 | 1.05 | 0.96 | 0.86 |
| Min | 0.41 | 0.01 | 0.01 | 0.02 | 0.01 | 0.01 |
| Max | 3.46 | 2.68 | 3.46 | 3.26 | 3.46 | 2.65 |
BL bucco-lingual, LP linear deviation at the implant platform, LA linear deviation at the implant apical, MD Mesio-distal, VP vertical deviation at the implant platform, VA vertical deviation at the implant apical
Intra- and inter-observer measurement reliabilities and TEM for implant placement accuracy parameters
| Observer | BL-LP | MD-LP | BL-LA | MD-LA | VP | VA |
|---|---|---|---|---|---|---|
| 0.986 (0.095) | 0.990 (0.080) | 0.983 (0.114) | 0.993 (0.080) | 0.991 (0.079) | 0.991 (0.089) | |
| R.H.P | NA | NA | NA | NA | NA | NA |
| N.Y. | 0.937 (0.194) | 0.909 (0.223) | 0.936 (0.224) | 0.904 (0.243) | 0.896 (0.256) | 0.911 (0.268) |
| M.I. | 0.885 (0.238) | 0.831 (0.290) | 0.968 (0.163) | 0.889 (0.261) | 0.898 (0.241) | 0.920 (0.241) |
Measurements are made by R.H.P (Ramadhan H. Putra), N.Y. (Nobuhiro Yoda), and M.I. (Masahiro Iikubo). Table entries are intraclass correlation coefficient of intra- and inter-observer measurement using single-measure, absolute agreement, and two-way mixed model; TEM in millimeters (in parentheses)
NA not applicable, BL bucco-lingual, LP linear deviation at the implant platform, LA linear deviation at the implant apical, MD mesio-distal, VP vertical deviation at the implant platform, VA vertical deviation at the implant apical
Fig. 5Scatter plot showing a linear correlation between bone density and implant deviation. Left: deviation at the implant platform; right: deviation at the implant apical
Fig. 6Scatter plot showing a linear correlation between bone width and implant deviation. Left: deviation at the implant platform; right: deviation at the implant apical
Fig. 7Scatter plot showing a linear correlation between cortical bone thickness and implant deviation. Left: deviation at the implant platform; right: deviation at the implant apical
Multiple linear regression analysis of the correlation between bone condition and implant placement accuracy parameters
| BL-LP | MD-LP | BL-LA | MD-LA | VP | VA | ||
|---|---|---|---|---|---|---|---|
| Bone density | (B) | − 0.39 | − 0.50* | − 0.46 | − 1.05** | − 0.61* | − 0.79** |
| (I) | [− 0.84, 0.06] | [− 0.95, − 0.05] | [− 0.97, 0.04] | [− 1.55, − 0.55] | [− 1.08, − 0.15] | [− 1.28, − 0.30] | |
| Bone width | (B) | − 0.50 | − 0.55 | − 0.68* | − 0.33 | − 0.61* | − 0.65* |
| (I) | [− 1.04, 0.03] | [− 1.08, − 0.01] | [− 1.28, − 0.07] | [− 0.93, 0.27] | [− 1.16, − 0.05] | [− 1.23, − 0.06] | |
| Cortical bone thickness | (B) | − 0.28 | − 0.20 | − 0.38 | − 0.28 | − 0.32 | − 0.41 |
| (I) | [− 0.77, 0.21] | [− 0.69, 0.29] | [− 0.94, 0.17] | [− 0.83, 0.27] | [− 0.83, 0.19] | [− 0.95, 0.13] | |
| 0.13 | 0.16 | 0.17 | 0.31 | 0.21 | 0.26 | ||
| Overall | 0.099 | 0.051 | 0.045 | 0.001 | 0.016 | 0.004 | |
B: coefficient; 95% confidence intervals in brackets; confidence level *p < 0.05, **p < 0.01
BL bucco-lingual, LP linear deviation at the implant platform, LA linear deviation at the implant apical, MD mesio-distal, VP vertical deviation at the implant platform, VA vertical deviation at the implant apical