Literature DB >> 22905668

A clinically relevant validation method for implant placement after virtual planning.

Luc M Verhamme1, Gert J Meijer, Tiny Boumans, Filip Schutyser, Stefaan J Bergé, Thomas J J Maal.   

Abstract

PURPOSE: To design a relevant method to compare the virtual planned implant position to the ultimately achieved implant position and to evaluate, in case of discrepancy, the cause for this.
MATERIALS AND METHODS: Five consecutive edentulous patients with retention problems of the upper denture received four implants in the maxilla. Preoperatively, first a cone-beam CT (CBCT) scan was acquired, followed by virtual implant planning. Then, a surgical template was designed and endosseous implants were flapless installed using the template as a guide. To inventory any differences in position, the postoperative CBCT scan was matched to the preoperative scan. The accuracy of implant placement was validated three-dimensionally (3D) and the Implant Position Orthogonal Projection (IPOP) validation method was applied to project the results to a bucco-lingual and mesio-distal plane. Subsequently, errors introduced by virtual planning, surgical instruments, and validation process were evaluated.
RESULTS: The bucco-lingual deviations were less obvious than mesio-distal deviations. A maximum linear tip deviation of 2.84 mm, shoulder deviation of 2.42 mm, and angular deviation of 3.41° were calculated in mesio-distal direction. Deviations included errors in planning software (maximum 0.15 mm), for surgical procedure (maximum 2.94°), and validation process (maximum 0.10 mm).
CONCLUSIONS: This study provides the IPOP validation method as an accurate method to evaluate implant positions and to elucidate inaccuracies in virtual implant planning systems.
© 2012 John Wiley & Sons A/S.

Entities:  

Keywords:  Implantology; cone-beam CT; implant placement; oral and maxillofacial surgery; validation; validation method; virtual implant planning

Mesh:

Substances:

Year:  2012        PMID: 22905668     DOI: 10.1111/j.1600-0501.2012.02565.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  11 in total

1.  Accuracy of using computer-aided rapid prototyping templates for mandible reconstruction with an iliac crest graft.

Authors:  Da-long Shu; Xiang-zhen Liu; Bing Guo; Wei Ran; Xin Liao; Yun-yan Zhang
Journal:  World J Surg Oncol       Date:  2014-06-24       Impact factor: 2.754

2.  Accuracy of a direct drill-guiding system with minimal tolerance of surgical instruments used for implant surgery: a prospective clinical study.

Authors:  Du-Hyeong Lee; Seo-Young An; Min-Ho Hong; Kyoung-Bae Jeon; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2016-06-17       Impact factor: 1.904

3.  Accuracy Evaluation of a Stereolithographic Surgical Template for Dental Implant Insertion Using 3D Superimposition Protocol.

Authors:  Corina Marilena Cristache; Silviu Gurbanescu
Journal:  Int J Dent       Date:  2017-05-07

4.  Bone Height Changes of the Mandibular Edentulous Ridge in Screw Retained Versus Telescopic Restorations for Completely Edentulous Patients.

Authors:  Eman Helal; Mohamed El-Zawahry; Ayman Gouda; Amr Hosny Elkhadem; Samira Ibrahim Ibrahim
Journal:  Open Access Maced J Med Sci       Date:  2017-01-18

5.  Integration of digital dental casts in cone beam computed tomography scans-a clinical validation study.

Authors:  Frits A Rangel; Thomas J J Maal; Martien J J de Koning; Ewald M Bronkhorst; Stefaan J Bergé; Anne Marie Kuijpers-Jagtman
Journal:  Clin Oral Investig       Date:  2017-09-20       Impact factor: 3.573

Review 6.  Narrative review regarding the applicability, accuracy, and clinical outcome of flapless implant surgery with or without computer guidance.

Authors:  Emitis Natali Naeini; Mandana Atashkadeh; Hugo De Bruyn; Jan D'Haese
Journal:  Clin Implant Dent Relat Res       Date:  2020-05-13       Impact factor: 3.932

7.  Accuracy of Guided Implant Surgery in the Edentulous Jaw Using Desktop 3D-Printed Mucosal Supported Guides.

Authors:  Rani D'haese; Tom Vrombaut; Geert Hommez; Hugo De Bruyn; Stefan Vandeweghe
Journal:  J Clin Med       Date:  2021-01-20       Impact factor: 4.241

8.  Accuracy of mucosa supported guided dental implant surgery.

Authors:  Andoni Jones
Journal:  Clin Case Rep       Date:  2018-09-23

9.  Influence of bone condition on implant placement accuracy with computer-guided surgery.

Authors:  Ramadhan Hardani Putra; Nobuhiro Yoda; Masahiro Iikubo; Yoshihiro Kataoka; Kensuke Yamauchi; Shigeto Koyama; Upul Cooray; Eha Renwi Astuti; Tetsu Takahashi; Keiichi Sasaki
Journal:  Int J Implant Dent       Date:  2020-09-20

10.  Accuracy of implant surgical guides fabricated using computer numerical control milling for edentulous jaws: a pilot clinical trial.

Authors:  Jinyou Chai; Xiaoqian Liu; Ramona Schweyen; Jürgen Setz; Shaoxia Pan; Jianzhang Liu; Yongsheng Zhou
Journal:  BMC Oral Health       Date:  2020-10-21       Impact factor: 2.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.