A-Yong Yu1,2, Hui Shao1, Anpeng Pan1,2, Qinmei Wang1,2, Zixu Huang1, Benhao Song1, Colm McAlinden1,3, Jinhai Huang4,5,6, Sisi Chen7,8,9.
Abstract
BACKGROUND: To investigate the biomechanical properties of the cornea in myopic eyes using corneal visualization Scheimpflug technology (Corvis ST). The relationships between the biomechanical properties of the cornea and the degree of myopia were also investigated.Entities:
Keywords: Corneal biomechanics; Corvis ST; Myopia; Scheimpflug
Mesh:
Year: 2020 PMID: 32652982 PMCID: PMC7353814 DOI: 10.1186/s12886-020-01530-w
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Participant demographics
| Parameters | Low myopia | Moderate myopia | High myopia | Severe myopia | |
|---|---|---|---|---|---|
| Gender (male: female) | 24:25 | 21:40 | 54:37 | 30:34 | < 0.05 |
| Age (years) | 29.2 ± 8.85 | 24.58 ± 4.97 | 25.85 ± 5.98 | 29.34 ± 8.56 | < 0.05 |
| SE (D) | −1.92 ± 0.72 | −4.64 ± 0.85 | −7.84 ± 1.04 | − 13.14 ± 3.27 | < 0.001 |
| Kf (D) | 42.87 ± 1.83 | 42.49 ± 1.16 | 42.87 ± 1.31 | 43.34 ± 1.41 | 0.007 |
| Ks (D) | 43.85 ± 1.70 | 43.46 ± 1.23 | 44.02 ± 1.44 | 44.56 ± 1.41 | < 0.001 |
| CCT (μm) | 537.22 ± 30.06 | 536.72 ± 32.95 | 538.92 ± 32.47 | 531.89 ± 31.72 | 0.599 |
| CV (mm3) | 60.12 ± 3.07 | 59.89 ± 3.49 | 60.57 ± 3.36 | 59.55 ± 3.42 | 0.291 |
| ACD (mm) | 3.12 ± 0.38 | 3.37 ± 0.31 | 3.25 ± 0.30 | 3.19 ± 0.31 | < 0.001 |
| AL (mm) | 24.61 ± 1.28 | 25.42 ± 1.09 | 26.36 ± 1.13 | 28.39 ± 1.52 | < 0.001 |
| IOP (mmHg) | 12.91 ± 2.21 | 13.07 ± 1.86 | 13.40 ± 2.39 | 13.67 ± 2.43 | 0.262 |
SE spherical equivalent, AL axial length, Ks steepest keratometry, Kf flattest keratometry, CCT central corneal thickness, CV corneal volume, ACD anterior chamber depth, AL Axial length, IOP intraocular pressure
All the parameters acquired by the Corvis ST
| Parameters | Low myopia | Moderate myopia | High myopia | Severe myopia (n = 64) | |
|---|---|---|---|---|---|
| A1-time (m/s) | 7.07 ± 0.21 | 7.07 ± 0.18 | 7.12 ± 0.24 | 7.14 ± 0.26 | 0.212 |
| A1-length (mm) | 1.76 ± 0.16 | 1.73 ± 0.12 | 1.73 ± 0.08 | 1.72 ± 0.09 | 0.293 |
| A1-velocity (m/s) | 0.15 ± 0.01 | 0.15 ± 0.01 | 0.15 ± 0.01 | 0.15 ± 0.01 | 0.602 |
| A2-time (m/s) | 21.93 ± 0.45 | 21.79 ± 0.36 | 21.76 ± 0.43 | 21.66 ± 0.42 | 0.011 |
| A2-length (mm) | 1.72 ± 0.28 | 1.71 ± 0.30 | 1.60 ± 0.25 | 1.55 ± 0.28 | 0.001 |
| A2-velocity (m/s) | −0.38 ± 0.08 | −0.40 ± 0.07 | − 0.42 ± 0.07 | − 0.46 ± 0.08 | < 0.001 |
| HC-time (m/s) | 16.89 ± 0.38 | 16.69 ± 0.35 | 16.72 ± 0.32 | 16.59 ± 0.30 | < 0.001 |
| PD (mm) | 4.16 ± 0.79 | 4.35 ± 0.66 | 4.23 ± 0.72 | 4.24 ± 0.79 | 0.619 |
| HC radius (mm) | 6.51 ± 0.79 | 6.63 ± 0.69 | 7.02 ± 0.86 | 6.52 ± 1.02 | < 0.001 |
| DA (mm) | 1.14 ± 0.11 | 1.12 ± 0.09 | 1.13 ± 0.10 | 1.18 ± 0.12 | 0.007 |
A1-time and A2-time: time from start until the first and second applanation; A1-length and A2-length: length of flattened cornea at the first and second applanation; A1-velocity and A2-velocity: corneal velocity during the first and second applanation; HC-time: time from start until highest concavity is arrived; PD: distance between the two peaks of the cornea at highest concavity; HC radius: central curvature at highest concavity; DA: maximum deformation amplitude at highest concavity. One-way analysis of variance (ANOVA) and LSD-t test was used to analyze the differences of corneal biomechanical parameters between the four groups. P < 0.05 was considered statistically significant
Stepwise multivariate regression model results
| Parameters | Age (years) | SE (D) | Kf (D) | Ks(D) | CCT (μm) | CV (mm3) | ACD (mm) | AL (mm) | IOP (mmHg) |
|---|---|---|---|---|---|---|---|---|---|
| β | β | β | β | β | β | β | β | β | |
| A1-time (m/s) | −0.020 0.025 | 0.989 < 0.001 | |||||||
| A1-length (mm) | 0.221 < 0.001 | ||||||||
| A1-velocity (m/s) | 0.133 0.022 | 0.222 < 0.001 | −0.156 0.004 | −0.373 < 0.001 | |||||
| A2-time (m/s) | 0.111 0.005 | 0.129 0.001 | 0.169 < 0.001 | −0.789 < 0.001 | |||||
| A2-length (mm) | −0.136 0.017 | 0.289 < 0.001 | −0.248 < 0.001 | 0.177 0.004 | |||||
| A2-velocity (m/s) | 0.408 < 0.001 | −0.119 0.008 | 0.597 < 0.001 | ||||||
| HC-time (m/s) | 0.289 < 0.001 | 0.143 0.016 | −0.234 < 0.001 | ||||||
| HC radius (mm) | 0.207 0.001 | 0.355 < 0.001 | |||||||
| DA (mm) | 0.155 < 0.001 | −0.119 0.039 | 0.084 0.026 | 0.151 0.008 | −0.785 < 0.001 |
SE spherical equivalent, Ks steepest keratometry, Kf flattest keratometry, CCT central corneal thickness, CV corneal volume, ACD anterior chamber depth, AL Axial length, IOP intraocular pressure; A1-time and A2-time: time from start until the first and second applanation; A1-length and A2-length: length of flattened cornea at the first and second applanation; A1-velocity and A2-velocity: corneal velocity during the first and second applanation moment; HC-time: time from start until highest concavity is arrived; HC radius: central curvature at highest concavity; DA: maximum deformation amplitude at highest concavity