Assessing agreement between the three common clinical measurement methods of HbA1c.

Mohammad Taha Jalali1,2, Samaneh Salehipour Bavarsad2, Saeed Hesam3, Mohammad Reza Afsharmanesh2, Narges Mohammadtaghvaei1,2.   

Abstract

Reliable measurement of hemoglobin A1c (HbA1c) has great importance in the diagnosis and monitoring of diabetes mellitus. The aim of the present study was to compare the performance parameters of the three common methods of HbA1c assay, including the Roche, Sebia and TOSOH G8 systems. We studied 120 patients referred to a clinical laboratory for HbA1c assay. The blood samples were analyzed with the Roche, Sebia and TOSOH G8 systems based on immunoassay, capillary electrophoresis, and ion-exchange chromatography techniques, respectively. The Spearman and the Passing-Bablok regression,as well as the Bland-Altman plots, were used to compare these methods. For each assay, the patients' classification was evaluated at the three cut-points of 6.5, 7, and 8% and the sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of the methods were estimated. Our results showed that there were good correlations and agreement between the methods. We found a mean difference of 0.07% for the TOSOH G8 vs. Roche, 0.06% for the TOSOH G8 vs. Sebia and - 0.01% for the Roche vs. Sebia. The methods represented very low bias, indicating the good accuracy of the results. The sensitivity and specificity of the methods were comparable as well. The three methods also performed similarly in the classification of patients at the proposed cut-off points. Based on our results, the Roche, Sebia and TOSOH G8 systems showed a very high level of agreement with comparable performance parameters and yielded similar and accurate classification of diabetic patients. Therefore, these methods can be used interchangeably. © Springer Nature Switzerland AG 2020.

Entities:  

Keywords:  Cut-offs; HbA1c; Misclassification; Roche; Sebia; TOSOH G8

Year:  2020        PMID: 32550176      PMCID: PMC7270288          DOI: 10.1007/s40200-020-00503-6

Source DB:  PubMed          Journal:  J Diabetes Metab Disord        ISSN: 2251-6581


  11 in total

1.  The effect of elevated fetal hemoglobin on hemoglobin A1c results: five common hemoglobin A1c methods compared with the IFCC reference method.

Authors:  Curt L Rohlfing; Shawn M Connolly; Jack D England; Steven E Hanson; Christina M Moellering; Janielle R Bachelder; Randie R Little
Journal:  Am J Clin Pathol       Date:  2008-05       Impact factor: 2.493

2.  Analysis of HbA1c on an automated multicapillary zone electrophoresis system.

Authors:  Niclas Rollborn; Torbjörn Åkerfeldt; Gunnar Nordin; Xiao Yan Xu; Aleksandra Mandic-Havelka; Lars-Olof Hansson; Anders Larsson
Journal:  Scand J Clin Lab Invest       Date:  2016-10-21       Impact factor: 1.713

3.  First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis.

Authors:  Stéphane Jaisson; Nathalie Leroy; Julie Meurice; Emmanuelle Guillard; Philippe Gillery
Journal:  Clin Chem Lab Med       Date:  2012-10-01       Impact factor: 3.694

4.  A comparison between turbidimetric inhibition immunoassay and capillary electrophoresis in glycated hemoglobin (HbA1c) measurement.

Authors:  Fehime Benli Aksungar; Mustafa Serteser; Abdurrahman Coşkun; İbrahim Ünsal
Journal:  Clin Chem Lab Med       Date:  2013-08       Impact factor: 3.694

5.  Performances of capillary electrophoresis and HPLC methods in HbA1c determination: diagnostic accuracy in HbS and HbD-Iran variants' presence.

Authors:  Mariarita Dessi; Massimo Pieri; Stefano Pignalosa; Francesca Gabriela Martino; Rossella Zenobi
Journal:  J Clin Lab Anal       Date:  2014-03-28       Impact factor: 2.352

6.  HbA1c analysis by capillary electrophoresis - comparison with chromatography and an immunological method.

Authors:  Olav Klingenberg; Torill Furuset; Camilla R Hestbråten; Marit H Hallberg; Astrid Steiro; Iren R Orset; Jens P Berg
Journal:  Scand J Clin Lab Invest       Date:  2017-06-23       Impact factor: 1.713

7.  Evaluation of the Sebia CAPILLARYS 2 flex piercing for the measurement of HbA(1c) on venous and capillary blood samples.

Authors:  Olivier Heylen; Stefaan Van Neyghem; Sandra Exterbille; Charline Wehlou; Frans Gorus; Ilse Weets
Journal:  Am J Clin Pathol       Date:  2014-06       Impact factor: 2.493

8.  Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser).

Authors:  Andrea Tesija Kuna; Kristina Dukic; Nora Nikolac Gabaj; Marijana Miler; Ines Vukasovic; Sanja Langer; Ana-Maria Simundic; Nada Vrkic
Journal:  Lab Med       Date:  2018-07-05

9.  A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA1c determination.

Authors:  Xiaobin Wu; Yan Chao; Zemin Wan; Yunxiu Wang; Yan Ma; Peifeng Ke; Xinzhong Wu; Jianhua Xu; Junhua Zhuang; Xianzhang Huang
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

10.  Effect of the different assays of HbA1c on diabetic patients monitoring.

Authors:  Farideh Razi; Ensieh Nasli Esfahani; Marjan Rahnamaye Farzami; Ali Tootee; Mostafa Qorbani; Soltan Ahmad Ebrahimi; Mehrzad Nahid; Parvin Pasalar
Journal:  J Diabetes Metab Disord       Date:  2015-08-04
View more
  1 in total

1.  Cases of fulminant type 1 and type 2 diabetes mellitus whose HbA1c levels were unmeasurable due to increased labile HbA1c.

Authors:  Shunsuke Kato; Izumi Otaka; Hanako Toyama; Ryota Kusumi; Kazuyuki Takahashi; Mitsuhiko Nara; Yumi Suganuma; Takehiro Sato; Tsukasa Morii; Hiroki Fujita; Hironori Waki
Journal:  Diabetol Int       Date:  2022-07-22
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.