Abstract
INTRODUCTION: Haemoglobin A1c (HbA1c) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important.Entities:
Keywords: Haemoglobin A1c; affinity chromatography; capillary electrophoresis (CE); high-performance liquid chromatography (HPLC); immunoassay
Mesh:
Substances:
Year: 2016 PMID: 27812304 PMCID: PMC5082223 DOI: 10.11613/BM.2016.039
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Precision and linearity of HbA1c values analyzed by the four analyzers
| Low | High | Low | High | Low | High | Low | High | low | high | A | B | |||
| 1.76 | 1.11 | 1.94 | 1.15 | 1.20 | 0.95 | 1.32 | 0.99 | 3.5 | 15.3 | 0.95–1.09 | - 1.06–0.30 | 0.9978 | y = 1.0193 x – 0.3812 | |
| 0.71 | 0.32 | 1.33 | 0.61 | 0.47 | 0.39 | 0.87 | 0.53 | 3.6 | 14.9 | 0.88–1.11 | - 0.67–1.62 | 0.9932 | y = 0.9938 x + 0.4744 | |
| 0.98 | 0.84 | 1.41 | 1.61 | 0.65 | 0.62 | 0.93 | 0.68 | 4.2 | 14.9 | 0.87–1.11 | - 0.75–1.79 | 0.9920 | y = 0.9900 x + 0.5207 | |
| 1.90 | 0.99 | 1.76 | 1.65 | 1.21 | 0.82 | 1.16 | 1.40 | 4.0 | 13.3 | - 0.62–0.20 | 0.97–1.06 | 0.9990 | y = 1.0107 x – 0.2094 | |
| Low - whole blood sample with 5.0% (31 mmol/mol) HbA1c concentration. High - whole blood sample with 8.5% (69 mmol/mol) HbA1c concentration. CV – coefficient of variation. Linearity was calculated using NGSP units (%). 95% CI – confidence intervals of 95%. R2 – coefficient of correlation. The regression line equation is presented as y = A x + B. A – regression line slope. B - regression line intercept. | ||||||||||||||
Accuracy of HbA1c values analyzed by the four analyzers
| 1 | 5.4 | 5.4 | 5.7 | 5.2 | 5.2 | 0.0 | 5.6 | -3.7 | -3.7 |
| 2 | 6.1 | 6.4 | 6.2 | 6.0 | 6.2 | 4.9 | 1.6 | -1.6 | 1.6 |
| 3 | 7.8 | 7.9 | 8.0 | 7.6 | 7.6 | 1.3 | 2.6 | -2.6 | -2.6 |
| 4 | 10.4 | 10.7 | 10.5 | 10.3 | 10.1 | 2.9 | 1.0 | -1.0 | -2.9 |
| The CAP proficiency testing acceptance limit of ± 6% was set as the accuracy limit (relative bias was calculated by NGSP units). | |||||||||
Sample interferences for common interferents
| Low | High | Low | High | Low | High | Low | High | |
| 3.64 | 0.00 | 1.89 | 0.00 | 1.89 | - 2.63 | 1.13 | 0.26 | |
| - 1.82 | 0.00 | 0.00 | 1.28 | 1.89 | - 1.32 | 1.89 | - 0.79 | |
| - 3.64 | 3.85 | 0.00 | 0.00 | 1.89 | 0.00 | 1.32 | - 1.18 | |
| - 1.82 | 1.28 | 1.89 | 0.00 | 1.89 | - 1.32 | - 0.19 | - 2.76 | |
| - 1.82 | 1.28 | 1.89 | 0.00 | 1.89 | 1.32 | - 0.38 | - 3.55 | |
| - 3.64 | 1.28 | 1.89 | 0.00 | 0.00 | - 1.32 | - 1.13 | 3.42 | |
| - 3.64 | 2.56 | 0.00 | 0.00 | 0.00 | - 2.63 | - 2.45 | 1.32 | |
| - 3.64 | 1.28 | 0.00 | - 1.28 | - 1.89 | - 2.63 | - 1.51 | - 1.45 | |
| - 3.64 | - 1.28 | 0.00 | - 1.28 | - 1.89 | - 3.95 | - 3.77 | 1.18 | |
| - 1.82 | - 1.28 | 0.00 | 0.00 | 1.89 | - 1.32 | - 0.38 | - 1.71 | |
| - 3.64 | - 2.56 | 0.00 | 0.00 | 1.89 | - 1.32 | - 2.83 | - 3.68 | |
| - 3.64 | - 3.85 | 1.89 | 0.00 | 0.00 | - 1.32 | - 2.83 | - 3.16 | |
| - 3.64 | - 1.28 | 0.00 | - 1.28 | 1.89 | - 1.32 | - 4.53 | - 3.16 | |
| - 3.64 | - 3.85 | 0.00 | - 1.28 | 1.89 | - 1.32 | - 4.15 | - 1.58 | |
| 0.00 | - 1.28 | - 1.89 | - 1.28 | 0.00 | - 1.32 | - 3.21 | - 2.50 | |
| - 3.64 | - 1.28 | 1.89 | 0.00 | 3.77 | - 1.32 | - 1.70 | - 3.55 | |
| 0.00 | - 1.28 | 0.00 | 1.28 | 1.89 | - 2.63 | - 2.83 | - 3.42 | |
| 0.00 | 0.00 | 1.89 | - 1.28 | 1.89 | - 1.32 | - 4.15 | - 3.42 | |
| 0.00 | 2.56 | 1.89 | 0.00 | 3.77 | - 2.63 | - 5.13 | ||
| 0.00 | 1.28 | 1.89 | 0.00 | 3.77 | - 2.63 | - 5.26 | ||
| / | / | / | / | 3.8 | 4.8 | / | / | |
| 1.92 | 0.00 | 3.70 | 1.23 | 4.53 | -1.45 | |||
| 0.00 | - 1.27 | 5.56 | 1.23 | 6.42 | 6.32 | |||
| - 5.77 | 1.27 | 4.94 | 56.60 | 32.89 | 10.00 | 7.37 | ||
| 1.92 | - 2.53 | 6.04 | 5.92 | |||||
| 0.00 | 0.00 | 4.72 | 3.42 | |||||
| - 1.82 | - 2.56 | 3.70 | 0.00 | 1.89 | - 1.32 | 3.58 | - 0.53 | |
| - 1.82 | - 3.85 | 5.56 | 1.23 | 3.77 | - 1.32 | 2.08 | 1.18 | |
| - 3.64 | - 5.13 | 5.56 | 2.47 | 1.89 | - 1.32 | 2.08 | - 1.18 | |
| - 3.64 | 0.00 | 2.47 | 1.89 | 0.00 | - 0.19 | 0.13 | ||
| - 3.64 | - 1.28 | 3.70 | 0.00 | - 2.63 | 2.45 | 1.45 | ||
| 1.82 | 1.16 | 0.00 | 0.00 | 0.00 | 1.19 | 2.82 | 4.81 | |
| - 1.82 | 1.16 | 0.00 | 0.00 | 0.00 | 0.00 | - 0.56 | 4.44 | |
| - 3.64 | 0.00 | 0.00 | - 3.49 | 1.82 | - 2.38 | 0.00 | 0.99 | |
| - 3.64 | - 1.16 | 0.00 | - 3.49 | 1.82 | - 2.38 | 1.13 | 0.74 | |
| - 3.64 | 0.00 | 0.00 | - 4.65 | 5.45 | - 1.19 | 2.07 | 0.00 | |
| Low - whole blood sample with HbA1c concentration of 5.3% measured by Premier Hb9210. High - whole blood sample with HbA1c concentration of 7.6% measured by Premier Hb9210. TBIL - total bilirubin. TG – triglycerides. CHOL – cholesterol. Vit C - vitamin C. Glu - glucose. “/” - not performed. The bold results show relative bias > ± 7% (calculated by NGSP units). Relative bias > ± 7% was considered clinically significant (NGSP criterion). | ||||||||
Figure 1Comparison of HbA1c values obtained with four analyzer using Passing and Bablok (PB) regression analysis and Bland-Altman (BA) plot. (A) Comparison of Capillarys 2FP and Premier Hb9210 by PB. (B) Comparison of Capillarys 2FP and Premier Hb9210 by BA. (C) Comparison of Tosoh G8 and Premier Hb9210 by PB. (D) Comparison of Tosoh G8 and Premier Hb9210 by BA. (E) Comparison of Roche c501 and Premier Hb9210 by PB. (F) Comparison of Roche c501 and Premier Hb9210 by BA. In the Passing-Bablok regression analyses the dotted lines show the 95% confidence interval (CI); the gray lines represent the identity line (X = Y). In the Bland-Altman plots the solid lines show the mean difference, while the dotted lines show the mean difference ± 1.96 SD.
HbA1c values for HbE on the Capillarys 2FP, Tosoh G8, Premier Hb9210, and Roche c501 systems
| Premier Hb9210 | Capillarys 2FP | Tosoh G8 | Roche c501 | Capillarys 2FP | Tosoh G8 | Roche c501 | |
| 1 | 5.5 | 5.6 | 4.2 | 5.4 | 1.8 | -1.8 | |
| 2 | 5.0 | 4.8 | 4.2 | 5.0 | -4.0 | 0.0 | |
| 3 | 5.3 | 5.3 | 4.7 | 5.6 | 0.0 | 5.6 | |
| 4 | 5.1 | 5.2 | 4.1 | 5.2 | 2.0 | 2.0 | |
| 5 | 5.6 | 5.9 | 4.6 | 5.7 | 5.4 | 1.8 | |
| 6 | 5.6 | 5.3 | 4.6 | 5.4 | -5.4 | -3.6 | |
| The boronate affinity HPLC method with the Premier Hb9210 instrument was used as the comparative method. The bold results show the relative bias > ± 7% (calculated by NGSP units). Relative bias > ± 7% was considered clinically significant (NGSP criterion). | |||||||
Figure 2Detection of HbE with the Capillarys 2FP and Tosoh G8 analyzers. The arrow indicates the presence of the variant (HbE).