Literature DB >> 23089707

First evaluation of Capillarys 2 Flex Piercing® (Sebia) as a new analyzer for HbA1c assay by capillary electrophoresis.

Stéphane Jaisson1, Nathalie Leroy, Julie Meurice, Emmanuelle Guillard, Philippe Gillery.   

Abstract

BACKGROUND: HbA1c is a key biomarker for the monitoring of glycemic balance in diabetic patients. It may be measured by various methods, including HPLC and immunoassays. Here we report the results of the first evaluation of Capillarys 2 Flex Piercing®, a new analyzer using capillary electrophoresis for the separation and the quantification of HbA1c.
METHODS: We have evaluated the analytical performances of the method as well as the influence of the most frequent analytical interferences regarding HbA1c assays (i.e., labile HbA1c, carbamylated hemoglobin, high HbF and hemoglobin variants).
RESULTS: Intra-assay and between-assays CVs were respectively lower than 1.98% and 2.68% on the pre-market version of the instrument, and lower than 1.62% and 1.45% on the commercial version. The linearity was excellent for HbA1c values ranging from 19 mmol/mol (3.9%) to 161 mmol/mol (16.9%) (r=0.999). The results were well correlated with those obtained by the HPLC method routinely used in the laboratory (Variant II® NU Kit - Bio-Rad): HbA1c[Capillarys 2]=0.9452×HbA1c[Variant II]+1.7279 (r=0.994, n=500). The use of titrated quality control samples indicated a good accuracy of the method. Neither the presence of HbF (until 15%), labile HbA1c or carbamylated hemoglobin, nor that of some typical hemoglobin variants, such as hemoglobin S, D, C and E, affected HbA1c measurement.
CONCLUSIONS: This evaluation showed that the analytical performances of Capillarys 2 Flex Piercing® analyzer for HbA1c assay fulfilled quality criteria requested for clinical use, and allowed to recommend its implementation in clinical chemistry laboratories for routine practice.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 23089707     DOI: 10.1515/cclm-2012-0017

Source DB:  PubMed          Journal:  Clin Chem Lab Med        ISSN: 1434-6621            Impact factor:   3.694


  9 in total

1.  Assessing agreement between the three common clinical measurement methods of HbA1c.

Authors:  Mohammad Taha Jalali; Samaneh Salehipour Bavarsad; Saeed Hesam; Mohammad Reza Afsharmanesh; Narges Mohammadtaghvaei
Journal:  J Diabetes Metab Disord       Date:  2020-02-11

2.  Performances of capillary electrophoresis and HPLC methods in HbA1c determination: diagnostic accuracy in HbS and HbD-Iran variants' presence.

Authors:  Mariarita Dessi; Massimo Pieri; Stefano Pignalosa; Francesca Gabriela Martino; Rossella Zenobi
Journal:  J Clin Lab Anal       Date:  2014-03-28       Impact factor: 2.352

3.  Evaluation of hemoglobin A1c measurement by Capillarys 2 electrophoresis for detection of abnormal glucose tolerance in African immigrants to the United States.

Authors:  Zhen Zhao; Jeffrey Basilio; Steven Hanson; Randie R Little; Anne E Sumner; David B Sacks
Journal:  Clin Chim Acta       Date:  2015-04-08       Impact factor: 3.786

Review 4.  Impact of haemoglobin variants on the use of haemoglobin A1c for the diagnosis and monitoring of diabetes: a contextualised review.

Authors:  Anne Marie Liddy; Stephan Grundy; Seamus Sreenan; William Tormey
Journal:  Ir J Med Sci       Date:  2022-04-01       Impact factor: 1.568

Review 5.  HbA1c: a review of analytical and clinical aspects.

Authors:  Cas Weykamp
Journal:  Ann Lab Med       Date:  2013-10-17       Impact factor: 3.464

6.  A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA1c determination.

Authors:  Xiaobin Wu; Yan Chao; Zemin Wan; Yunxiu Wang; Yan Ma; Peifeng Ke; Xinzhong Wu; Jianhua Xu; Junhua Zhuang; Xianzhang Huang
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

7.  Labile glycated haemoglobin and carbamylated haemoglobin are still critical points for HbA1c measurement.

Authors:  Aurore Desmons; Stéphane Jaisson; Nathalie Leroy; Philippe Gillery; Emmanuelle Guillard
Journal:  Biochem Med (Zagreb)       Date:  2017-06-15       Impact factor: 2.313

8.  Hemoglobin Camperdown [β104Arg→Ser] Detection During Hemoglobin A(1c) Measurement via Capillary Electrophoresis.

Authors:  Valéry Brunel; Patrick Caneiro; Agnès Lahary; Guy Hue; Christian Thuillez
Journal:  Ann Lab Med       Date:  2016-07       Impact factor: 3.464

9.  False measurement of glycated hemoglobin in patients without hemoglobin A.

Authors:  Minghuan Suo; Dongmei Wen; Weijia Wang; Decai Zhang; Shengnan Xu; Xia Wang; Ting Hu
Journal:  Biosci Rep       Date:  2019-01-30       Impact factor: 3.840

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.