Comparison of Enzymatic Assay for HBA1C Measurement (Abbott Architect) With Capillary Electrophoresis (Sebia Minicap Flex Piercing Analyser).

Andrea Tesija Kuna1, Kristina Dukic1, Nora Nikolac Gabaj1, Marijana Miler1, Ines Vukasovic1, Sanja Langer1, Ana-Maria Simundic2, Nada Vrkic1.   

Abstract

OBJECTIVE: To compare the analytical performances of the enzymatic method (EM) and capillary electrophoresis (CE) for hemoglobin A1c (HbA1c) measurement.
METHODS: Imprecision, carryover, stability, linearity, method comparison, and interferences were evaluated for HbA1c via EM (Abbott Laboratories, Inc) and CE (Sebia).
RESULTS: Both methods have shown overall within-laboratory imprecision of less than 3% for International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) units (<2% National Glycohemoglobin Standardization Program [NGSP] units). Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for EM, compared with CE, but were not clinically relevant (<5 mmol/mol; NGSP <0.5%). At the clinically relevant HbA1c concentration, stability observed with both methods was acceptable (bias, <3%). Triglyceride levels of 8.11 mmol per L or greater showed to interfere with EM and fetal hemoglobin (HbF) of 10.6% or greater with CE.
CONCLUSION: The enzymatic method proved to be comparable to the CE method in analytical performances; however, certain interferences can influence the measurements of each method.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 29528429     DOI: 10.1093/labmed/lmx090

Source DB:  PubMed          Journal:  Lab Med        ISSN: 0007-5027


  2 in total

1.  Assessing agreement between the three common clinical measurement methods of HbA1c.

Authors:  Mohammad Taha Jalali; Samaneh Salehipour Bavarsad; Saeed Hesam; Mohammad Reza Afsharmanesh; Narges Mohammadtaghvaei
Journal:  J Diabetes Metab Disord       Date:  2020-02-11

2.  Analytical evaluation of the new Seal Autoanalyzer 3 High Resolution for urinary iodine determination.

Authors:  Valentina Vidranski; Maja Franceschi; Dražena Krilić; Tomislav Jukić; Ivan Mihaljević; Zvonko Kusić
Journal:  Biochem Med (Zagreb)       Date:  2019-06-15       Impact factor: 2.313

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.