Analysis of HbA1c on an automated multicapillary zone electrophoresis system.

Niclas Rollborn1, Torbjörn Åkerfeldt1, Gunnar Nordin2, Xiao Yan Xu1, Aleksandra Mandic-Havelka3, Lars-Olof Hansson1, Anders Larsson1.   

Abstract

Hemoglobin A1c (HbA1c) is a frequently requested laboratory test and there is thus a need for high throughput instruments for this assay. We evaluated a new automated multicapillary zone electrophoresis instrument (Capillarys 3 Tera, Sebia, Lisses, France) for analysis of HbA1c in venous samples. Routine requested HbA1c samples were analyzed immunologically on a Roche c6000 instrument (n = 142) and then with the Capillarys 3 Tera instrument. The Capillarys 3 Tera instrument performed approximately 70 HbA1c tests/hour. There was a strong linear correlation between Capillarys 3 Tera and Roche Tina-Quant HbA1c Gen 3 assay (y = 1.003x - 0.3246 R2 = .996). The total CV for the 12 capillaries varied between 0.8 and 2.2% and there was a good agreement between duplicate samples (R2 = .997). In conclusion, the Capillarys 3 Tera instrument has a high assay capacity for HbA1c. It has a good precision and agreement with the Roche Tina-Quant HbA1c method and is well suited for high volume testing of HbA1c.

Entities:  

Keywords:  Capillary electrophoresis; HbA1c; human; whole blood

Mesh:

Substances:

Year:  2016        PMID: 27768851     DOI: 10.1080/00365513.2016.1238507

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  1 in total

1.  Assessing agreement between the three common clinical measurement methods of HbA1c.

Authors:  Mohammad Taha Jalali; Samaneh Salehipour Bavarsad; Saeed Hesam; Mohammad Reza Afsharmanesh; Narges Mohammadtaghvaei
Journal:  J Diabetes Metab Disord       Date:  2020-02-11
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.