Tawfiq Khurayzi1, Fida Almuhawas1, Abdulrahman Sanosi1.
Abstract
BACKGROUND: Cochlear morphology and cochlear duct length (CDL) play important roles in the selection of appropriate electrodes. Cochlear parameters such as diameter (A value) and width (B value) are used as inputs for calculating the CDL. Current measurements of these parameters are inefficient and time consuming. Recently developed otological planning software (OTOPLAN) allows surgeons to directly measure these parameters and then automatically calculate the CDL.Entities:
Mesh:
Year: 2020 PMID: 32493102 PMCID: PMC7270618 DOI: 10.5144/0256-4947.2020.218
Source DB: PubMed Journal: Ann Saudi Med ISSN: 0256-4947 Impact factor: 1.526
Figure 1.(A) DICOM viewing software from the CT scanner, The A-value is the length of the cochlear base, indicated by the yellow line. B, C and D: Cochlear view in the OTOPLAN: The A-value, indicated by green dots. The B-value is the width of the cochlear base, indicated by blue dots. The H-value is the height of the cochlea, indicated by red dots.
Figure 2.3D SLICER software segmentation for one of our sample shows normal inner ear morphology including full 2.5 turns cochlea.
Figure 3.Cross-sectional of the cochlea shows the level of the organ of Corti (OC) as green dots and at level of lateral wall (LW) as blue dots. The organ of Corti (OC) is located 0.5 mm away from the lateral wall (LW).
Distribution of sex and ear side in the study sample.
| N | % | |
|---|---|---|
| Females | 46 | 52.3 |
| Males | 42 | 47.7 |
| Left | 37 | 42 |
| Right | 51 | 58 |
Cochlear duct length (CDL) at the level of the organ of corti(OC) and the first turn of the cochlea at the organ of corti(OC) and lateral wall(LW).
| Range (mm) | Mean (mm) | SD | N | Mean | SD | SE | Significance | |||
|---|---|---|---|---|---|---|---|---|---|---|
| CDL(OC) | 28.1–37.8 | 32.91 | 1.78 | Male | 42 | 33.39 | 1.52 | 0.23 | .014 | Significant |
| Female | 46 | 32.47 | 1.90 | 0.28 | ||||||
| 1.0 Turn(OC) | 15.39–21.08 | 18.22 | 1.05 | Male | 42 | 18.50 | 0.89 | 0.14 | .015 | Significant |
| Female | 46 | 17.96 | 1.12 | 0.16 | ||||||
| 1.0 Turn(LW) | 18.39–24.10 | 21.23 | 1.05 | Male | 42 | 21.51 | 0.89 | 0.14 | .015 | Significant |
| Female | 46 | 20.97 | 1.12 | 0.17 |
Cochlear duct length (CDL), organ of Corti (OC), lateral wall (LW).
Correlation of cochlear duct length CDL(OC), 1.0 turn(OC) and 1.0 turn(LW) to other parameters, n=88.
| A-value OTOPLAN | B-value OTOPLAN | H-value OTOPLAN | B/A | H/A | CDL(OC) | 1.0 Turn(OC) | 1.0 Turn(LW) | ||
|---|---|---|---|---|---|---|---|---|---|
| CDL(OC) | Pearson correlation | .636** | .961** | 0.189 | .498** | -0.068 | 1 | 1.000** | 1.000** |
| Sig. (2-tailed) | 0.000 | 0.000 | 0.078 | 0.000 | 0.531 | 0.000 | 0.000 | ||
| 1.0 Turn(OC) | Pearson correlation | .636** | .961** | 0.189 | .498** | -0.068 | 1.000** | 1 | 1.000** |
| Sig. (2-tailed) | 0.000 | 0.000 | 0.078 | 0.000 | 0.531 | 0.000 | 0.000 | ||
| 1.0 Turn(LW) | Pearson correlation | .634** | .962** | 0.188 | .501** | -0.067 | 1.000** | 1.000** | 1 |
| Sig. (2-tailed) | 0.000 | 0.000 | 0.079 | 0.000 | 0.534 | 0.000 | 0.000 |
Cochlear duct length (CDL), organ of Corti (OC), lateral wall (LW).
The different reported means of cochlear parameters values in the various studies that are done based on the radiological method in measurement including our current study.
| Author | Year | N | A (mm) | B (mm) | H (mm) |
|---|---|---|---|---|---|
| Escudé et al[ | 2006 | 84 | 9.23 | 6.9 | N/A |
| Martinez-Monedero et al[ | 2011 | 124 | 8.39 | 6.98 | N/A |
| Van der Marel et al[ | 2014 | 671 | 8.85 | 6.58 | N/A |
| Avci et al[ | 2014 | 16 | 9.2 | 7.0 | 4.4 |
| Pelliccia et al[ | 2014 | 482 | 9 | 6.8 | N/A |
| Meng J et al[ | 2016 | 178 men | 9.12 | 6.41 | N/A |
| 132 women | 8.93 | 6.22 | |||
| Liu et al[ | 2017 | 204 | 8.84 ± 0.29 | 6.30 ± 0.38 | 3.59 ± 0.12 |
| Rivas et al[ | 2017 | 309 | 9.22 ± 0.44 | N/A | |
| This study | 2019 | 88 | DICOM viewer 8.43 | N/A | N/A |
| OTOPLAN 8.44 | 6.53 | 2.71 |