Meer Rownaq Ali Abbasi1, Dileep Nag Vinnakota1, Vijaya Sankar1, Rekhalakshmi Kamatham2.
Abstract
INTRODUCTION: One of the principle factors for the success of implant supported/retained overdentures (IOs) is the manner in which the stresses are transferred to the surrounding bone. Hence, the aim of the present study is to compare the stress induced in the mandible around IOs, using two different attachment systems, locator and telescopic.Entities:
Keywords: attachment; locator; overdenture; stress; telescopic
Year: 2020 PMID: 32478325 PMCID: PMC7243882 DOI: 10.15386/mpr-1312
Source DB: PubMed Journal: Med Pharm Rep ISSN: 2602-0807
Material properties incorporated in the study.
| S.No | Material | Young’s Modulus (GPa) | Poisson’s Ratio | Density |
|---|---|---|---|---|
| 1 | Cortical bone | 13700 | 0.3 | 850–1250 HU (D2 bone) |
| 2 | Cancellous bone | 1370 | 0.3 | - |
| 3 | Titanium alloy (Ti6Al4V-implant) | 135000 | 0.33 | 4.43 |
| 4 | Nylon Ring | 350 | 0.4 | - |
| 5 | Cobalt Chromium | 218000 | 0.33 | 8.5 |
| 6 | Mucosa/Gingiva | 1 | 0.37 | - |
| 7 | Type 3 Gold | 85000 | 0.3 | 15 |
| 8 | Polymethyl Methacrylate (PMMA) | 2400 | 0.35 | 1.19 |
GPa: Gigapascal, g/cm3: Grams per cubic millimeter, HU: Hounsfield unit
Number of elements and nodes for each model.
| S.No | Component | Element Type | Elements | Nodes |
|---|---|---|---|---|
| 1 | Cortical bone | C3D101 | 133093 | 196993 |
| 2 | Cancellous Bone | C3D101 | 37645 | 66882 |
| 3 | Gingiva | C3D101 | 57360 | 95009 |
| 4 | Implant | C3D101 | 73820 | 110234 |
| 5 | Nylon cap | C3D101 | 15159 | 22838 |
| 6 | Telescopic attachment | C3D101 | 5321 | 8229 |
| 7 | Screw | C3D101 | 38028 | 56653 |
| 8 | Splint | C3D101 | 21650 | 33229 |
| 9 | Denture | C3D101 | 77208 | 115810 |
| Total | 772559 | 503491 | ||
Contact stress induced for vertical compressive load of 35 Newtons.
| Stress (in Megapascal) | |||||
|---|---|---|---|---|---|
| S.No | Components | Non-Splinted | Splinted | ||
| Locator | Telescopic | Locator | Telescopic | ||
| 1 | Cortical Bone | 3.73 | 2.66 | 4.12 | 3.7 |
| 2 | Cancellous Bone | 0.295 | 0.49 | 1.17 | 1.05 |
| 3 | Gingiva | 0.14 | 0.138 | 0.21 | 0.22 |
| 4 | Cap | 5.7 | 40.5 | 7.42 | 61.47 |
| 5 | Implant | 2.1 | 8.73 | 25.35 | 24.2 |
| 6 | Attachment | 10.25 | 26.41 | 12.1 | 30.19 |
| 7 | Screw | - | - | 34.5 | 30.89 |
| 8 | Splinted Bar | - | - | 43.32 | 48.36 |
| 9 | Abutment | - | - | 57.65 | 54.22 |
Deflection for vertical compressive load of 35 Newtons.
| Deflection (in millimeters) | |||||
|---|---|---|---|---|---|
| S.No | Components | Non-splinted | Splinted | ||
| Locator | Telescopic | Locator | Telescopic | ||
| 1 | Cortical Bone | 6.15E-04 | 4.14E-04 | 4.52E-04 | 4.19E-04 |
| 2 | Cancellous Bone | 1.18E-03 | 1.09E-03 | 8.93E-04 | 7.89E-04 |
| 3 | Gingiva | 9.72E-02 | 9.25E-02 | 1.16E-01 | 1.18E-01 |
| 4 | Cap | 1.34E-02 | 8.36E-03 | 2.98E-02 | 5.85E-02 |
| 5 | Implant | 7.61E-04 | 4.58E-04 | 1.11E-03 | 1.00E-03 |
| 6 | Attachment / Attachment on splinted bar | 9.40E-04 | 3.59E-03 | 9.61E-03 | 8.73E-03 |
| 7 | Screw | - | - | 8.63E-04 | 8.11E-04 |
| 8 | Splinted Bar | - | - | 1.88E-02 | 1.77E-02 |
| 9 | Abutment | - | - | 4.82E-03 | 4.45E-03 |