| Literature DB >> 32477744 |
Galen M Kabulski1,2, Amanda Northup3, Barbara S Wiggins1,2.
Abstract
Infections related to cardiac implantable electronic device (CIED) placement are associated with poor clinical outcomes. As such, preprocedural prophylactic antibiotic therapy is indicated for all patients prior to device insertion. However, the available data are less clear on the impact of postprocedural antibiotic therapy on rates of CIED infection when used in addition to preprocedural therapy. This is single-center, retrospective cohort study of 913 patients who underwent CIED-related procedures between October 2010 and August 2014 sought to compare the rate of CIED infections in patients receiving only preprocedural antibiotics with those receiving both preprocedural and postprocedural antibiotics. Univariate analysis was used to detect independent risk factors for CIED infection. After excluding patients receiving concomitant antibiotics for other conditions, those undergoing CIED extraction alone, and those with a lack of follow-up data and/or adequate documentation of clinical encounters, 569 patients were identified for inclusion in the final analysis. The majority of patients who received postprocedural antibiotics received three to five days of therapy, with the most common antibiotic used being cephalexin. There was no statistically significant difference in the incidence of infection between patients who did and did not receive postoperative antibiotics (4.5% versus 6.1%; p = 0.398). In a multivariate analysis, the use of postprocedural antibiotic therapy was not a significant risk factor for infection (adjusted odds ratio: 0.692; 95% confidence interval: 0.314-1.525; p = 0.361). It is therefore reasonable to withhold prescribing postoperative antibiotics in patients following CIED implantation. Individualized risk factor evaluation of patient comorbidities and procedural characteristics may be needed to aid in determining whether postoperative antibiotics are reasonable in different patients. The validity of these findings is contingent on further confirmation via a prospective, randomized clinical trial. Copyright:Entities:
Keywords: Antibiotic; cardiac implantable electronic device; implantable cardioverter defibrillator; permanent pacemaker; prophylaxis
Year: 2019 PMID: 32477744 PMCID: PMC7252755 DOI: 10.19102/icrm.2019.100804
Source DB: PubMed Journal: J Innov Card Rhythm Manag ISSN: 2156-3977
Comparison of Baseline Characteristics Between Patients Who Received Postoperative Antibiotic Prophylaxis and Those Who Did Not
| Postoperative Antibiotics (n = 401), n (%) | No Postoperative Antibiotics (n = 168), n (%) | p-value | |
|---|---|---|---|
| Patient characteristics | |||
| Male gender | 256 (63.8) | 104 (58.1) | 0.188 |
| Renal impairment* | 165 (41.3) | 62 (35) | 0.158 |
| Dialysis | 8 (2) | 6 (3.4) | 0.381 |
| Heart failure | 190 (52.6) | 92 (51.4) | 0.786 |
| Chronic liver disease | 18 (4.5) | 5 (2.8) | 0.331 |
| Diabetes | 122 (30.4) | 56 (31.3) | 0.835 |
| History of malignancy | 62 (15.5) | 23 (12.8) | 0.411 |
| Anticoagulation | 147 (36.7) | 64 (35.8) | 0.834 |
| Chronic steroids | 15 (3.8) | 7 (3.9) | 0.926 |
| Periprocedural factors | |||
| Fever** | 1 (0.4) | 5 (5) | 0.008*** |
| Prior CIED infection | 8 (2) | 6 (3.4) | 0.381 |
| Indwelling line | 15 (6.7) | 11 (11.7) | 0.140 |
| Periprocedural pacing | 11 (4.3) | 13 (14.4) | 0.001*** |
| Two or more new leads implanted | 258 (64.3) | 93 (52) | 0.005*** |
| Device | 0.518 | ||
| ICD | 149 (37.2) | 59 (33) | |
| PPM | 156 (38.9) | 72 (40.2) | |
| Combination | 2 (0.5) | 0 (0) | |
| CRT-D | 83 (20.7) | 45 (25.1) | |
| CRT-P | 11 (2.7) | 3 (1.7) | |
| Performing service | < 0.001*** | ||
| Cardiology | 380 (94.8) | 74 (41.3) | |
| Cardiothoracic surgery | 17 (4.2) | 88 (49.2) | |
| Pediatric cardiology | 2 (0.5) | 14 (7.8) | |
| Combined | 2 (0.5) | 3 (1.7) | |
| Procedure | < 0.001*** | ||
| Initial | 299 (74.6) | 82 (45.8) | |
| Revision | 87 (21.7) | 94 (52.5) | |
| Reimplantation | 14 (3.5) | 3 (1.7) | |
| Postoperative hematoma | 25 (9.9) | 19 (17.3) | 0.047*** |
| Reintervention within 90 days | 16 (4) | 15 (8.4) | 0.031*** |
CIED: cardiovascular implantable electronic device; CRT-D: cardiac resynchronization therapy defibrillator; CRT-P: cardiac resynchronization therapy pacemaker; ICD: implantable cardioverter-defibrillator; PPM: permanent pacemaker.
*Serum creatinine ≥ 1.5 mg/dL or estimated creatinine clearance < 60 mL/min upon admission.
**Fever within 24 hours prior to the procedure (defined as greater than 38°C) or hypothermia (defined as less than 35°C).
***Statistically significant.
Incidence of CIED Infection According to the Presence or Absence of Postoperative Antibiotics
| Postoperative Antibiotics (n = 401), n (%) | No Postoperative Antibiotics (n = 168), n (%) | p-value | |
|---|---|---|---|
| Any infection | 18 (4.5) | 11 (6.1) | 0.398 |
| Systemic infection | 2 (0.5) | 0 (0) | 1.000 |
CIED: cardiovascular implantable electronic device.
Risk Factors for CIED Infection by Univariate Logistic Regression
| Previously Identified Risk Factors[ | CIED Infection | |
|---|---|---|
| Odds Ratio (95% Confidence Interval) | p-value | |
| Patient characteristics | ||
| Renal impairment | 1.357 (0.633–2.908) | 0.433 |
| Dialysis | 3.327 (0.709–15.614) | 0.128 |
| Heart failure | 2.504 (1.090–5.750) | 0.030* |
| Chronic liver disease | 6.157 (2.108–17.985) | 0.001* |
| Diabetes | 1.901 (0.894–4.041) | 0.095 |
| Anticoagulation | 2.247 (1.059–4.768) | 0.035* |
| Chronic steroids | 3.225 (0.897–11.595) | 0.073 |
| Prior CIED infection | 1.478 (0.187–11.703) | 0.711 |
| Periprocedural factors | ||
| Indwelling line | 0.542 (0.070–4.209) | 0.558 |
| Periprocedural pacing | 1.199 (0.152–9.445) | 0.863 |
| Two or more new leads implanted | 1.253 (0.572–2.746) | 0.573 |
| Prior CIED procedure | 0.578 (0.273–1.222) | 0.151 |
| Postoperative antibiotics | 0.718 (0.332–1.553) | 0.400 |
CIED: cardiovascular implantable electronic device.
*Statistically significant.
Risk Factors for CIED Infection According to Multivariable Logistic Regression
| Risk Factor | CIED Infection | |
|---|---|---|
| Adjusted Odds Ratio (95% Confidence Interval) | p-value | |
| Heart failure | 2.533 (1.092–5.974) | 0.030* |
| Chronic liver disease | 6.502 (2.168–19.398) | 0.001* |
| Postoperative antibiotics | 0.692 (0.314–1.525) | 0.361 |
CIED: cardiovascular implantable electronic device.
*Statistically significant.