Literature DB >> 27996097

Prevention of Cardiac Implantable Electronic Device Infections: Single Operator Technique with Use of Povidone-Iodine, Double Gloving, Meticulous Aseptic/Antiseptic Measures and Antibiotic Prophylaxis.

Antonis S Manolis1, Helen Melita2.   

Abstract

BACKGROUND: Cardiac implantable electronic device (CIED) implantation is complicated by infection still at a worrisome rate of 2-5%. Since early on during device implantation procedures, we have adopted an infection-preventive technique which has hitherto resulted in effective prevention of infections. Herein we present our results of applying this technique by a single operator in a prospective series of 762 consecutive patients undergoing device implantation.
METHODS: A meticulous search for and treatment of active, occult, or smoldering infection was undertaken preoperatively. An aseptic/antiseptic technique was used for implantation of each device. Skin preparation is thorough with initial cleansing performed with alcohol followed by povidone-iodine 10% solution, which is also used in the wound and inside the pocket. In addition, we routinely use double gloving, and IV antibiotic prophylaxis 1 hour before and for 48 hours afterwards followed by oral antibiotic for 2-3 days after discharge. The skin is closed with absorbable sutures. The study includes 382 patients having a new pacemaker (n = 333) or battery change, system upgrade or lead revision (n = 49), and 380 patients having a new implantable cardioverter-defibrillator (ICD) (n = 296) or device replacement/upgrade/lead revision (n = 84).
RESULTS: The pacemaker group, aged 70.2 ± 16.5 years, includes 18% VVI, 49% DDD, 29% VDD, and 4% cardiac resynchronization therapy (CRT) devices. The ICD group, aged 61.3 ± 13.0 years, with a mean ejection fraction of 36 ± 13%, includes 325 ICD and 55 CRT implants. Over 26.6 ± 33.4 months for the pacemaker group and 36.6 ± 38.3 months for the ICD group, infection occurred in one patient in each group (0.26%) having a device replacement.
CONCLUSION: A consistent and strict approach of aseptic/antiseptic technique with the use of double gloving and povidone-iodine solution within the pocket plus a 4-day regimen of antibiotic prophylaxis minimizes infections in CIED implants.
© 2016 Wiley Periodicals, Inc.

Entities:  

Keywords:  antibiotic prophylaxis; cardiac implantable electronic devices; cardiac resynchronization therapy; defibrillators; endocarditis; infections; pacemakers

Mesh:

Substances:

Year:  2017        PMID: 27996097     DOI: 10.1111/pace.12996

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  7 in total

1.  Real-world effectiveness of infection prevention interventions for reducing procedure-related cardiac device infections: Insights from the veterans affairs clinical assessment reporting and tracking program.

Authors:  Archana Asundi; Maggie Stanislawski; Payal Mehta; Anna E Baron; Hillary J Mull; P Michael Ho; Peter J Zimetbaum; Kalpana Gupta; Westyn Branch-Elliman
Journal:  Infect Control Hosp Epidemiol       Date:  2019-06-04       Impact factor: 3.254

Review 2.  A Roadmap for Reducing Cardiac Device Infections: a Review of Epidemiology, Pathogenesis, and Actionable Risk Factors to Guide the Development of an Infection Prevention Program for the Electrophysiology Laboratory.

Authors:  Westyn Branch-Elliman
Journal:  Curr Infect Dis Rep       Date:  2017-08-16       Impact factor: 3.725

3.  Care Bundle Approach to Minimizing Infection Rates after Neurosurgical Implants for Neuromodulation: A Single-Surgeon Experience.

Authors:  Elsa V Arocho-Quinones; Chiang-Ching Huang; Barney D Ward; Peter A Pahapill
Journal:  World Neurosurg       Date:  2019-04-12       Impact factor: 2.104

4.  "Real life" longevity of implantable cardioverter-defibrillator devices.

Authors:  Antonis S Manolis; Themistoklis Maounis; Spyridon Koulouris; Vassilios Vassilikos
Journal:  Clin Cardiol       Date:  2017-05-22       Impact factor: 2.882

5.  Cardiac implantable electronic device lead extraction using the lead-locking device system: keeping it simple, safe, and inexpensive with mechanical tools and local anesthesia.

Authors:  Antonis S Manolis; Georgios Georgiopoulos; Sofia Metaxa; Spyridon Koulouris; Dimitris Tsiachris
Journal:  Anatol J Cardiol       Date:  2017-08-11       Impact factor: 1.596

6.  Postoperative Antibiotic Prophylaxis Following Cardiac Implantable Electronic Device Placement.

Authors:  Galen M Kabulski; Amanda Northup; Barbara S Wiggins
Journal:  J Innov Card Rhythm Manag       Date:  2019-08-15

7.  Post procedural complications of cardiac implants done in a resource limited setting under 'C' arm: A single centre experience.

Authors:  A Jayachandra; Vivek Aggarwal; Sandeep Kumar; I V Nagesh
Journal:  Med J Armed Forces India       Date:  2017-12-09