Sutapa Bandyopadhyay Neogi1, Jyoti Sharma2, Shivam Pandey2, Nausheen Zaidi2, Maitreyee Bhattacharya3, Rakhee Kar4, Sitanshu Sekhar Kar4, Abhishek Purohit5, Sanjib Bandyopadhyay3, Renu Saxena6.
Abstract
BACKGROUND: Accurate diagnosis of anemia by community workers using a point-of-care device is a challenge. The objective of the study was to establish the diagnostic accuracy of point-of-care devices for detecting anemia in community settings.Entities:
Keywords: Anemia; Community; Devices; Diagnosis; Hemoglobin; Invasive; Non invasive
Mesh:
Year: 2020 PMID: 32456639 PMCID: PMC7249358 DOI: 10.1186/s12913-020-05329-9
Source DB: PubMed Journal: BMC Health Serv Res ISSN: 1472-6963 Impact factor: 2.655
Fig. 1Flowchart depicting recruitment of participants in the study for various tests
Distribution of study participants as per ICMR classification of anemia across two sites
| Hb (gm/dl) | Total ( | Puducherry ( | Kolkata ( |
|---|---|---|---|
| 11.64 (±2.7) | 12.31 (±2.4) | 10.8 (±2.8) | |
| 2–20.2 | 2–20.2 | 4.2–18.6 | |
| 938 (67.1%) | 580 (77.1%) | 358 (55.4%) | |
| 124 (8.8%) | 58 (7.7%) | 66 (10.2%) | |
| 240 (17.2%) | 94 (12.5%) | 146 (22.6%) | |
| 96 (6.8%) | 20 (2.6%) | 76 (11.7%) |
Fig. 2Diagnostic Accuracy Parameters for testing anemia and no anemia as per ICMR classification
Fig. 3Diagnostic Accuracy Parameters for testing severe anemia as per ICMR classification
Difference in Hb readings (mean, 95% CI) between autoanalyzer (reference) and index tests in different categories of anemia
| Type of device | Device | All readings (Hb 2–20 g/dl) | Moderate and severe anemia (Hb < 10 g/dl) | Severe anemia (Hb < 7 g/dl) |
|---|---|---|---|---|
| 0.47 (0.42 to 0.52) | 0.82 (0.71 to 0.92) | 0.92 (0.81 to 1.02) | ||
| 0.18 (0.11 to 0.25) | 0.70 (0.58 to 0.82) | 0.84 (0.66 to 1.03) | ||
| −0.28(− 0.36 to 0.20) | −0.23(− 0.41 to − 0.04) | 0.05(− 0.52 to 0.63) | ||
| −0.22(− 0.35 to − 0.09) | −0.23(− 0.43 to − 0.03) | −0.06(− 0.71 to 0.58) | ||
| 0.06(−0.10 to 0.24) | 1.7 (1.55 to 1.97) | −0.39(−1.37 to 0.57) | ||
| 0.47 (0.37 to 0.58) | 1.7 (1.56 to 1.98) | 1.10 (0.20 to 1.99) | ||
| 0.43 (0.17 to 0.69) | 2.6 (2.07 to 3.2) | 3.84 (2.62 to 5.06) | ||
| 0.71 (0.46 to 0.96) | 2.5 (1.88 to 3.12) | 4.1 (2.6 to 5.6) |
Fig. 4Bland-Altman Plot (Concordance analysis) showing agreement between auto-analyser V/s venous sample for Hemocue (Hb values expressed as gm/dl)
Fig. 5Bland-Altman Plot (Concordance analysis) showing agreement between auto-analyzer V/s venous sample for TrueHb (Hb values expressed as gm/dl)