Validation of a hemoglobinometer for use in outdoor blood donation camps.

Anju Dubey1, Saurabh Murti1.   

Abstract

Entities:  

Year:  2016        PMID: 27605859      PMCID: PMC4993091          DOI: 10.4103/0973-6247.187943

Source DB:  PubMed          Journal:  Asian J Transfus Sci        ISSN: 0973-6247


× No keyword cloud information.
Sir, Hemoglobin estimation of blood donors is performed to prevent anemic donors from donating blood and ensuring a good quality of packed red blood cell component. Several devices are presently available for hemoglobin estimation of blood donors as a screening test before blood donation.[1] The desirable features of a device intended for mass screening include high accuracy, precision, rapidity, and ease of operation.[2] In our part of country, there is a wide variation in temperature conditions. Thus, any device which has to be used for hemoglobin screening in outdoor blood donation camps should be able to give its optimum performance even at highest temperature range in that region. In this study, we have compared the performance of HemoCue Hb301 System at indoor temperatures (18–22°C) and outdoor temperatures (>35°C) in the month of July 2014. Ethylenediaminetetraacetic acid anticoagulated venous samples of 100 patients which were sent for hemoglobin estimation were tested by manual cyanmethemoglobin method which was standardized using commercially available standards (Span diagnostics, Surat, India) as a routine testing method. For comparing the accuracy, these were tested by HemoCue Hb301 System both at in the air-conditioned departmental laboratory (T1) and at outdoor work station (T2) by the same operator. The samples initially placed at T1 for 1 h were tested by cyanmethemoglobin method and HemoCue Hb301 System. These were then kept at T2 for 1 h before retesting with HemoCue Hb301 System after ensuring their physical integrity. For comparing the precision, 5 samples tested by cyanmethemoglobin method were taken. Each was retested 10 times by HemoCue Hb301 System both at T1 and T2. The results obtained are summarized in Figure 1 and Table 1.
Figure 1

Results of precision determination for 10 replicated performed on 5 patient samples at T1 and T2

Table 1

Results of precision determination for 10 replicated performed on 5 patient samples at T1 and T2

Results of precision determination for 10 replicated performed on 5 patient samples at T1 and T2 Results of precision determination for 10 replicated performed on 5 patient samples at T1 and T2 HemoCue Hb301 is a quantitative point of care system based on photometric detection of hemoglobin level. A previous study has evaluated its performance against previous HemoCueHb 201+ system in diagnosing anemia of pregnancy and found it sufficiently precise and accurate.[3] The system is designed to operate at 10–40°C[4] but there are no studies to the best of our knowledge, for evaluating its performance at varied temperature conditions. The device has shown good accuracy at both T1 (correlation coefficient r = 0.903) as well as T2 (correlation coefficient r = 0.893). The precision of measurement was also good at both the temperatures with coefficient of variation ranging from 0.060 to 0.094 at T1 and from 0.041 to 0.085 at T2. Since the process of validation was performed before installation of the equipment in the department, it also allowed a chance for hands-on training of the technical staff. We conclude that HemoCue Hb301 system is adequately apt for both routine uses and use in outdoor blood donation camps.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.
  3 in total

1.  Evaluation of various methods of point-of-care testing of haemoglobin concentration in blood donors.

Authors:  Abhay Singh; Anju Dubey; Atul Sonker; Rajendra Chaudhary
Journal:  Blood Transfus       Date:  2014-10-29       Impact factor: 3.443

2.  Validation and implementation of a new hemoglobinometer for donor screening at Canadian Blood Services.

Authors:  Mindy Goldman; Samra Uzicanin; Qi-Long Yi; Jason Acker; Sandra Ramirez-Arcos
Journal:  Transfusion       Date:  2012-07       Impact factor: 3.157

3.  [The new HemoCue system Hb 301 for the haemoglobin measurement in pregnant women].

Authors:  C Tayou Tagny; L Kouam; D Mbanya
Journal:  Ann Biol Clin (Paris)       Date:  2008 Jan-Feb       Impact factor: 0.459

  3 in total
  3 in total

Review 1.  Techniques used for the screening of hemoglobin levels in blood donors: current insights and future directions.

Authors:  Rajendra Chaudhary; Anju Dubey; Atul Sonker
Journal:  J Blood Med       Date:  2017-07-03

2.  Diagnostic accuracy of point-of-care devices for detection of anemia in community settings in India.

Authors:  Sutapa Bandyopadhyay Neogi; Jyoti Sharma; Shivam Pandey; Nausheen Zaidi; Maitreyee Bhattacharya; Rakhee Kar; Sitanshu Sekhar Kar; Abhishek Purohit; Sanjib Bandyopadhyay; Renu Saxena
Journal:  BMC Health Serv Res       Date:  2020-05-26       Impact factor: 2.655

3.  Digital Hemoglobinometers as Point-of-Care Testing Devices for Hemoglobin Estimation: A Validation Study from India.

Authors:  Kapil Yadav; Shashi Kant; Gomathi Ramaswamy; Farhad Ahamed; Kashish Vohra
Journal:  Indian J Community Med       Date:  2020-10-28
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.