| Literature DB >> 29197856 |
Guy-Marino Hinnouho1, Maxwell A Barffour1, K Ryan Wessells1, Kenneth H Brown1,2, Sengchanh Kounnavong3, Bigphone Chanhthavong3, Kethmany Ratsavong3, Chidchamai Kewcharoenwong4, Sonja Y Hess1.
Abstract
BACKGROUND: Haemoglobin (Hb) assessment by Hemocue is used widely for anaemia screening in both adults and children. However, few studies have compared the diagnostic accuracy of Hemocue with an automated haematology analyser in young children. AIM: To compare Hb concentrations by Hemocue Hb301 and two automated haematology analysers in young children in rural communities of Lao PDR.Entities:
Keywords: diagnosis; epidemiology; evaluating instrument; paediatric haematology
Mesh:
Substances:
Year: 2017 PMID: 29197856 PMCID: PMC5969348 DOI: 10.1136/jclinpath-2017-204786
Source DB: PubMed Journal: J Clin Pathol ISSN: 0021-9746 Impact factor: 3.411
Haemoglobin concentration and anaemia prevalence in young Laotian children (n=1487) using two different methods of assessment
| Hemocue Hb301 | Automated haematology analysers combined* | P value | |
| Haemoglobin concentration, g/L | |||
| Mean±SD | 108.4±10.3 | 102.3±13.1 | <0.0001 |
| Range | 70–141 | 34–178 | |
| SE | 0.3 | 0.3 | |
| Anaemia prevalence, n (%) | 799 (53.7) | 1099 (73.9) | <0.0001 |
| Sensitivity of Hemocue Hb301 (%) | 68.7 | Ref | |
| Specificity of Hemocue Hb301 (%) | 85.8 | Ref | |
| Agreement of the two methods (%) | 71.9 |
*XT-1800i by Sysmex and BC-3000Plus by Mindray Medical International.
Figure 1Correlation of Hb concentrations assessed by Hemocue Hb301 and automated analysers combined (XT-1800i by Sysmex and BC-3000Plus by Mindray Medical International Ltd.) and concordance plots between the 2 methods.
Figure 2Bland-Altman plot showing agreement in haemoglobin concentration of young Laotian children assessed by HemoCue Hb301 and two automated haematology analysers combined (XT-1800i by Sysmex and BC-3000Plus by Mindray Medical International).
Paired comparisons of haemoglobin concentrations by HemoCue Hb301 (capillary and venous) and by Mindray BC-3000Plus in young Laotian children (n=129)
| Comparisons | Hb (mean±SD), g/L | P* | Bias (limits of agreement) | Anaemia prevalence n (%) | P† |
| Capillary HemoCue | 111.0±10.7 | 54 (41.9) | |||
| versus | 0.03 | −2.5 | 0.21 | ||
| Venous HemoCue | 113.6±14.0 | 47 (36.4) | |||
| Capillary HemoCue | 111.0±10.7 | 54 (41.9) | |||
| versus | <0.0001 | 8.7 | <0.0001 | ||
| Venous automated analyser | 102.3±17.4 | 85 (65.9) | |||
| Venous HemoCue | 113.6±14.0 | 47 (36.4) | |||
| versus | <0.0001 | 11.2 | <0.0001 | ||
| Venous automated analyser | 102.3±17.4 | 85 (65.9) |
*P value for difference of mean Hb concentration.
†P value for anaemia prevalence.
Hb, haemoglobin.