Hsuan-Yu Chen1, Andrew Chen2, Chiachung Chen3.
Abstract
Many types of thermometers have been developed to measure body temperature. Infrared thermometers (IRT) are fast, convenient and ease to use. Two types of infrared thermometers are uses to measure body temperature: tympanic and forehead. With the spread of COVID-19 coronavirus, forehead temperature measurement is used widely to screen people for the illness. The performance of this type of device and the criteria for screening are worth studying. This study evaluated the performance of two types of tympanic infrared thermometers and an industrial infrared thermometer. The results showed that these infrared thermometers provide good precision. A fixed offset between tympanic and forehead temperature were found. The measurement values for wrist temperature show significant offsets with the tympanic temperature and cannot be used to screen fevers. The standard operating procedure (SOP) for the measurement of body temperature using an infrared thermometer was proposed. The suggestion threshold for the forehead temperature is 36 °C for screening of fever. The body temperature of a person who is possibly ill is then measured using a tympanic infrared thermometer for the purpose of a double check.Entities:
Keywords: COVID-19; body temperature; forehead temperature; infrared thermometer; tympanic temperature
Mesh:
Year: 2020 PMID: 32438729 PMCID: PMC7284737 DOI: 10.3390/s20102885
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Specifications of the infrared thermometers.
| BRAUN IRT-3020 | OMEON MC-510 | Tasco THI-301 | |
|---|---|---|---|
| Sensing elements | Pyroelectrical detector | Thermopile | Thermopile |
| Measuring range | 34 °C–42.2 °C | 34 °C–42.2 °C | 0 °C–50 °C |
| Resolution | 0.1 °C | 0.1 °C | 0.1 °C |
| Nonlinearity and repeatability | 37.0 °C–39.0 °C, ±0.1 °C | 36.0 °C–39.0 °C, ±0.2 °C | ±0.5 °C |
| ≤37.0 °C or ≥39 °C, ±0.2 °C | ≤37.0 °C or ≥39 °C, ±0.3 °C | After calibration |
Measurements of the tympanic temperature, forehead temperature and wrist temperature using a BRAUN infrared thermometer.
| Earleft | Earright | Forehead | Wrist | Earave | Eardiff | Earave–Forehead | Earave–Wrist | Forehead–Wrist | |
|---|---|---|---|---|---|---|---|---|---|
| Mean | 36.904 | 36.912 | 34.714 | 34.1644 | 36.911 | −0.258 | 2.1966 | 2.7467 | 0.5501 |
| Standard deviation | 0.286 | 0.259 | 0.392 | 0.455 | 0.264 | 0.139 | 0.411 | 0.415 | 0.532 |
| CV (%) | 0.775 | 0.712 | 1.129 | 1.332 | |||||
| Min | 35.8 | 35.9 | 34.0 | 34.0 | 35.85 | −0.3 | 0.15 | 0.95 | 0 |
| Max | 37.5 | 37.8 | 37.3 | 36.1 | 37.7 | 0.4 | 3.4 | 3.7 | 2.1 |
Note: 1. Earleft: the reading values for the left ear. 2. Earright: the reading values for the right ear. 3. Earave: the average of reading values for both ears. 4. Eardiff: the difference between the measurements for two ears. 5. Earave–Forehead: the difference between the average tympanic temperature and forehead temperature. 6. Earave–Wrist: the difference between the average tympanic temperature and wrist temperature. 7. Forehead–Wrist: the difference between the forehead and wrist temperature. 8. CV: coefficient of variance, CV = (standard deviation)/mean.
The correlation between measurement results for the first experiment.
| Air Temp | Earleft | Earright | Forehead | Wrist | |
|---|---|---|---|---|---|
| Air temp | 1 | ||||
| Earleft | 0.2627 | 1 | |||
| Earright | 0.3958 | 0.8602 * | 1 | ||
| Forehead | −0.018 | 0.1004 | 0.1488 | 1 | |
| Wrist | 0.0982 | 0.0656 | 0.1313 | 0.1005 | 1 |
Note: * represented the relationship is significant at p < 0.05 according to Duncan’s multiple range test.
Figure 1The relationship between the air temperatures and body temperatures measured at the ear, forehead and wrist.
Figure 2The relationship between Earave–forehead and Earave (the average temperature of both ears). Earave–forehead is the difference between average ear temperature (Earave) and forehead temperature.
Figure 3The relationship between Earave–wrist and Earave (the average temperature of both ears). Earave–wrist is the difference between the average ear temperature (Earave) and wrist temperature.
Statistics of the measure results for tympanic temperatures for the second experiment.
| Bright | Bleft | Bave | Bdiff | Oright | Oleft | Oave | Odiff | |
|---|---|---|---|---|---|---|---|---|
| Mean | 36.937 | 36.881 | 36.909 | 0.0461 | 36.740 | 36.679 | 36.719 | 0.0403 |
| Standard deviation | 0.301 | 0.293 | 0.286 | 0.155 | 0.264 | 0.266 | 0.250 | 0.174 |
| CV (%) | 0.815 | 0.794 | 0.719 | 0.725 | ||||
| Min | 36.3 | 36.2 | 36.25 | 0.3 | 36.2 | 36.0 | 36.1 | −0.5 |
| Max | 37.7 | 37.5 | 37.6 | 0.4 | 37.2 | 37.4 | 37.25 | 0.4 |
Note: 1. Bright: the reading values for the right ear using a BRAUN tympanic infrared thermometer. 2. Bleft: the reading values for the left ear using a BRAUN tympanic infrared thermometer. 3. Bave: the average of the right and left temperatures using a BRAUN tympanic infrared thermometer. 4. Bdiff: the difference between the right and left temperature using a BRAUN tympanic infrared thermometer. 5. Oright: the reading values for the right ear using an OMRON tympanic infrared thermometer. 6. Oleft: the reading values for the left ear using an OMRON tympanic infrared thermometer. 7. Oave: the average of the right and left temperatures by OMRON tympanic infrared thermometer. 8. Odiff: the difference between the right and left temperatures using an OMRON tympanic infrared thermometer.
The statistics of the measurement results for a THI-301 thermometer.
| Forehead | Wristright | Wristleft | |
|---|---|---|---|
| Mean | 34.802 | 33.601 | 33.605 |
| Standard deviation | 0.343 | 0.602 | 0.554 |
| CV (%) | 0.986 | 1.791 | 1.640 |
| Min | 33.8 | 32.4 | 32.2 |
| Max | 35.7 | 34.8 | 34.7 |
1. Wristright: the right wrist temperature; 2. Wristleft: the left wrist temperature; 3. CV: Coefficient of variance, CV = (standard deviation)/mean.
The statistics of the comparison of measurement.
| Bave–Forehead | Bave–Wristright | Bave–Wristleft | Wristright–Wristleft | |
|---|---|---|---|---|
| Mean | 2.1070 | 3.2781 | 3.3035 | 0.02544 |
| Standard deviation | 0.311 | 0.527 | 0.534 | 0.385 |
| Min | 1.32 | 1.8 | 2.05 | −0.9 |
| Max | 3.10 | 4.8 | 4.65 | 0.9 |
Bave–Forehead: the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and the forehead temperatures using a THI-301 infrared thermometer. Bave–Wristright: the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and the right wrist temperatures using a THI-301 infrared thermometer. Bave–Wristleft: the difference between the average tympanic temperatures using a BRAUN tympanic infrared thermometer and left wrist temperatures using a THI-301 infrared thermometer. Wristright–Wristleft: the difference between the wrist temperatures of two hands using a THI-301 infrared thermometer.
The correlation for the measurements for the second experiment.
| Bright | Bleft | Oright | Oleft | THIforehead | THIright wrist | THIleft wrist | |
|---|---|---|---|---|---|---|---|
| Bright | 1 | ||||||
| Bleft | 0.864 | 1 | |||||
| Oright | 0.670 | 0.617 | 1 | ||||
| Oleft | 0.693 | 0.668 | 0.784 | 1 | |||
| THIforehead | 0.424 | 0.472 | 0.191 | 0.246 | 1 | ||
| THIright wrist | 0.206 | 0.167 | 0.075 | 0.041 | 0.448 | 1 | |
| THIleft wrist | 0.3526 | 0.279 | 0.0518 | 0.0930 | 0.4758 | 0.738 | 1 |
Bright: the reading values for the right ear using a BRAUN tympanic infrared thermometer. Bleft: the reading values for the left ear using a BRAUN tympanic infrared thermometer. Oright: the reading values for the right ear using an OMRON tympanic infrared thermometer. Oleft: the reading values for the left ear using an OMRON tympanic infrared thermometer. THIforehead: the reading values for the forehead using a THI-301 infrared thermometer. THIright wrist: the reading values for the right hand using a THI-301 infrared thermometer. THIleft wrist: the reading values for the left hand using a THI-301 infrared thermometer.
Figure 4The relationship between Bave–Forehead and Bave: Bave–Forehead is the difference between the average era temperature (Earave) using a BRAUN infrared thermometer and the forehead temperature using a THI-301 infrared thermometer.
Figure 5The relationship between the Bave–Wristright, the Bave–Wristleft and the Bave. Bave–Wristright is the difference between the average ear temperature (Earave) using a BRAUN infrared thermometer and the right wrist temperature using a THI-301.