| Literature DB >> 22319287 |
Wenbin Chung1, Chiachung Chen.
Abstract
Infrared tympanic thermometers (ITTs) are easy to use and have a quick response time. They are widely used for temperature measurement of the human body. The accuracy and uncertainty of measurement is the importance performance indicator for these meters. The performance of two infrared tympanic thermometers, Braun THT-3020 and OMRON MC-510, were evaluated in this study. The cell of a temperature calibrator was modified to serve as the standard temperature of the blackbody. The errors of measurement for the two meters were reduced by the calibration equation. The predictive values could meet the requirements of the ASTM standard. The sources of uncertainty include the standard deviations of replication at fixed temperature or the predicted values of calibration equation, reference standard values and resolution. The uncertainty analysis shows that the uncertainty of calibration equation is the main source for combined uncertainty. Ambient temperature did not have the significant effects on the measured performance. The calibration equations could improve the accuracy of ITTs. However, these equations did not improve the uncertainty of ITTs.Entities:
Keywords: calibration equation; infrared tympanic thermometer; uncertainty
Mesh:
Year: 2010 PMID: 22319287 PMCID: PMC3257961 DOI: 10.3390/s100403073
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Specification of the infrared clinical thermometers.
| Sensing element | thermopile | thermopile |
| Measuring range | 34–42.2 °C | 34–42.2 °C |
| Resolution | 0.1 °C | 0.1 °C |
| Nonlinearity and repeatability | 1. 36.0–39 °C, ±0.2 °C | 1. 37.0–39 °C, ±0.1 °C |
Figure 1.The relationship between reading values of MC-510 thermometer versus standard values.
Figure 2.The error distribution of OMRON MC-510 thermometer at three ambient temperatures.
Figure 3.The predicted errors of calibration equation of OMRON MC-510 thermometer.
Figure 4.The relationship between reading values of BRAUN IRT-3020 thermometer versus standard values.
Figure 5.The error distribution of BRAUN IRT-3020 thermometer at three ambient temperatures.
Figure 6.The predicted errors of calibration equation of BRAUN IRT-3020 thermometer.
Figure 7.The standard deviation of five measurements of OMRON MC-510 thermometer at different standard temperatures and three leaves of ambient temperatures.
Analysis of variance ANOVA table for the effect of standard temperature and ambient temperature on the measurement of OMRON MC-510 thermometer.
| Standard temp. | 0.003717 | 4 | 0.000929 | 5.9318 | 0.01614 | 3.8379 |
| Ambient temp | 0.000832 | 2 | 0.000416 | 2.6552 | 0.1305 | 4.4590 |
| Errors | 0.001253 | 8 | 0.000157 | |||
| Total | 0.005802 | 14 |
Note: SS is the sum of square, df is the degree of freedom, MS is the mean square and F-critical value is the smallest level of significance that be to reject the hypothesis of F test.
Figure 8.The standard deviation of five measurements of BRAUN IRT-3020 thermometer at different standard temperatures and three leaves of ambient temperatures.
Analysis of variance ANOVA table for the effect of standard temperature and ambient temperature on the measurement of BRAUN IRT-3020 thermometer.
| Standard temp. | 0.000133 | 4 | 3.321E-05 | 0.1926 | 0.9355 | 3.8379 |
| Ambient temp. | 0.000617 | 2 | 0.000309 | 1.7924 | 0.2274 | 4.4590 |
| Errors | 0.001377 | 8 | 0.000172 | |||
| Total | 0.002127 | 14 |
The type A uncertainty of several observations for two infrared tympanic thermometers.
| 34.5 | 36.0 | 37.5 | 39.0 | 40.5 | ||
|---|---|---|---|---|---|---|
| OMRON | None | 0.1799 | 0.2193 | 0.1767 | 0.1408 | 0.1506 |
| MC-510 | Polynomial equation | 0.2026 | 0.1990 | 0.1986 | 0.1989 | 0.2100 |
| BRAUN | None | 0.0884 | 0.0535 | 0.0567 | 0.0594 | 0.05606 |
| IRT-300 | Linear equation | 0.0711 | 0.0704 | 0.0701 | 0.0703 | 0.0709 |
The type B uncertainty analysis for OMRON MC-510 thermometer.
| Reference (uref) | 0.03 °C | 0.0153 | Normal |
| Resolution (ures) | 0.1 °C | 0.0289 | Rectangular |
| Nonlinear and repeatability | |||
| Unon1 36–39 °C | 0.2 °C | 0.1155 | Rectangular |
| Unon2 ≤ 36 °C, ≥39 °C | 0.3 °C | 0.1732 |
The type B uncertainty analysis for BRAUN IRT-3020 thermometer.
| Reference (uref) | 0.03 °C | 0.0153 | Normal |
| Resolution (ures) | 0.1 °C | 0.0289 | Rectangular |
| Nonlinear and repeatability | Rectangular | ||
| Unon1 37–39 °C | 0.1 °C | 0.0577 | |
| Unon2 ≤ 37 °C, ≥39 °C | 0.2 °C | 0.1155 |
The combined uncertainty for two IR thermometers.
| 34.5 | 36.0 | 37.5 | 39.0 | 40.5 | ||
|---|---|---|---|---|---|---|
| OMRON | None | 0.2 | 0.2 | 0.2 | 0.1 | 0.23 |
| MC-510 | Polynomial equation | 519 | 499 | 136 | 851 | 18 |
| 0.2 | 0.2 | 0.2 | 0.2 | 0.27 | ||
| BRAUN | None | 685 | 324 | 321 | 246 | 41 |
| IRT-300 | Linear equation | |||||
| 0.1 | 0.0 | 0.0 | 0.0 | 0.13 | ||
| 491 | 873 | 872 | 890 | 24 | ||
| 0.1 | 0.0 | 0.0 | 0.0 | 0.13 | ||
| 395 | 965 | 965 | 967 | 94 | ||