Mengjiao Liu1, Xiaobo Wei1, Yun Ding1, Changhai Cheng1, Wenming Yin1, Jie Chen1, Kou Wang1, Wendong Gu2.
Abstract
It has been clearly confirmed that radiation therapy (RT) after breast conserving surgery (BCS) is an effective treatment modality comparable to mastectomy for early breast cancer. The purpose of this study was to further evaluate the accuracy of 3D surface imaging system (Sentinel) for breast cancer patients received BCS. The optical surface scans and CBCT scans were acquired before and immediately after couch movement correction. The deviation of the CBCT scans from the reference planning CT was considered an estimate for the residual errors for patient setup correction. The planning target volume (PTV) margins for treatment sessions was calculated according to the setup errors. We obtained a total of 245 sets of data collected from 49 breast cancer patients. Compared with Sentinel setup errors, the residual setup errors as determined by the CBCT scans after couch movement correction were reduced in the six directions. The PTV margins derived from the CBCT residual errors were all less than 5 mm in X, Y, and Z directions. Our results suggested that Optical surface imaging can be applied in positioning for breast cancer patient accurately without unnecessary imaging dose.Entities:
Mesh:
Year: 2020 PMID: 32371862 PMCID: PMC7200723 DOI: 10.1038/s41598-020-64496-1
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
The variation analysis of Sentinel errors and CBCT residual errors.
| SE | CBCT residual errors | P-value | |
|---|---|---|---|
| X(left-right) | −1.45 mm ± 4.44 mm | 0.94 mm ± 2.65 mm | <0.001 |
| Y(cranial-caudal) | −1.97 mm ± 4.72 mm | 1.64 mm ± 3.11 mm | <0.001 |
| Z(anteroposterior) | −2.62 mm ± 3.98 mm | 1.21 mm ± 2.57 mm | <0.001 |
| Rx(x-rotation) | 0.5° ± 0.9° | 0.3° ± 0.9° | <0.001 |
| Ry(y-rotation) | 0.2° ± 0.6° | 0.3° ± 1.0° | <0.001 |
| Rz(z-rotation) | 0.0° ± 0.5° | 0.0° ± 0.8° | <0.001 |
SE: sentinel errors; CBCT: cone-beam computerized tomography.
Setup errors of Sentinel and residual setup errors of CBCT.
| The group mean (M) | Systemic errors (∑) | Random errors (σ) | PTV margin | ||
|---|---|---|---|---|---|
| X-shift (mm) | CBCT (mm) | 0.71 | 1.73 | 1.34 | 4.40 |
| Sentinel (mm) | −1.83 | 2.70 | 2.14 | 6.89 | |
| Y-shift (mm) | CBCT (mm) | 1.62 | 1.90 | 1.70 | 4.99 |
| Sentinel (mm) | −2.00 | 2.62 | 2.30 | 6.85 | |
| Z-shift (mm) | CBCT (mm) | 0.84 | 1.60 | 2.03 | 4.62 |
| Sentinel (mm) | −3.00 | 2.63 | 2.41 | 6.95 | |
| X-rotation (°) | CBCT (°) | 0.5 | 0.8 | 0.6 | — |
| Sentinel (°) | 0.3 | 0.9 | 0.8 | — | |
| Y-rotation (°) | CBCT (°) | 0.2 | 0.6 | 0.8 | — |
| Sentinel (°) | 0.2 | 1.0 | 0.9 | — | |
| Z-rotation (°) | CBCT (°) | 0.0 | 0.5 | 0.7 | — |
| Sentinel (°) | 0.0 | 0.8 | 0.9 | — |
PTV: planning target volume.
Characteristics of the study population (n = 49).
| Clinical characteristics | Value |
|---|---|
| Intraductal carcinoma (n) | 10 |
| Invasive carcinoma (n) | 39 |
| pTisN0M0 (n) | 10 |
| pT1N0M0 (n) | 22 |
| pT2N0M0 (n) | 17 |
| Mammary gland volume(cm3) | 467.9 ± 173.4 |
| Age (years old) | 52.5 ± 7.0 |
| The average BMI (kg/m2) | 24.3 ± 2.9 |
BMI: body mass index.
Figure 1The image of Sentinel and CBCT positioning correction equipment.
Figure 2Flow chart of research steps.