Literature DB >> 18762347

Comparison of peripheral dose from image-guided radiation therapy (IGRT) using kV cone beam CT to intensity-modulated radiation therapy (IMRT).

Julian R Perks1, Jöerg Lehmann, Allen M Chen, Claus C Yang, Robin L Stern, James A Purdy.   

Abstract

PURPOSE: The growing use of IMRT with volumetric kilovoltage cone-beam computed tomography (kV-CBCT) for IGRT has increased concerns over the additional (typically unaccounted) radiation dose associated with the procedures. Published data quantify the in-field dose of IGRT and the peripheral dose from IMRT. This study adds to the data on dose outside the target area by measuring peripheral CBCT dose and comparing it with out-of-field IMRT dose.
MATERIALS AND METHODS: Measurements of the CBCT peripheral dose were made in an anthropomorphic phantom with TLDs and were compared to peripheral dose measurements for prostate IMRT, determined with MOSFET detectors.
RESULTS: Doses above 1cGy (per scan) were found outside the CBCT imaged volume, with 0.2cGy at 25 cm from the central axis. IMRT peripheral dose was 1cGy at 20 cm and 0.4cGy at 25 cm (per fraction).
CONCLUSIONS: An appreciable dose can be found beyond the edge of the IGRT field; of similar order of magnitude as peripheral dose from IMRT (mGy), and approximately half the dose delivered to the same point from the IMRT treatment (0.2cGy c.f. 0.4cGy 25 cm from the isocenter). This shows that peripheral dose, as well as the in-field dose from CBCT, needs to be taken into account when considering long term care of radiation oncology patients.

Entities:  

Mesh:

Year:  2008        PMID: 18762347     DOI: 10.1016/j.radonc.2008.07.026

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  9 in total

1.  Comparison of prostate positioning guided by three-dimensional transperineal ultrasound and cone beam CT.

Authors:  Minglun Li; Hendrik Ballhausen; Nina-Sophie Hegemann; Michael Reiner; Stefan Tritschler; Christian Gratzke; Farkhad Manapov; Stefanie Corradini; Ute Ganswindt; Claus Belka
Journal:  Strahlenther Onkol       Date:  2016-12-07       Impact factor: 3.621

2.  Acute Toxicity in Hypofractionated/Stereotactic Prostate Radiotherapy of Elderly Patients: Use of the Image-guided Radio Therapy (IGRT) Clarity System.

Authors:  Rossella DI Franco; Valentina Borzillo; Domingo Alberti; Gianluca Ametrano; Angela Petito; Andrea Coppolaro; Ilaria Tarantino; Sabrina Rossetti; Sandro Pignata; Gelsomina Iovane; Sisto Perdonà; Giuseppe Quarto; Giovanni Grimaldi; Alessandro Izzo; Luigi Castaldo; Raffaele Muscariello; Marcello Serra; Gaetano Facchini; Paolo Muto
Journal:  In Vivo       Date:  2021 May-Jun       Impact factor: 2.155

3.  Patient-based low dose cone beam CT acquisition settings for prostate image-guided radiotherapy treatments on a Varian TrueBeam linear accelerator.

Authors:  Maria Antonietta Piliero; Margherita Casiraghi; Davide Giovanni Bosetti; Simona Cima; Letizia Deantonio; Stefano Leva; Francesco Martucci; Marino Tettamanti; Francesco Pupillo; Luca Bellesi; Antonella Richetti; Stefano Presilla
Journal:  Br J Radiol       Date:  2020-08-27       Impact factor: 3.039

4.  Use of new radiochromic devices for peripheral dose measurement: potential in-vivo dosimetry application.

Authors:  S-T Chiu-Tsao; Mf Chan
Journal:  Biomed Imaging Interv J       Date:  2009-10-01

5.  An organ and effective dose study of XVI and OBI cone-beam CT systems.

Authors:  Daniel E Hyer; Christopher F Serago; Siyong Kim; Jonathan G Li; David E Hintenlang
Journal:  J Appl Clin Med Phys       Date:  2010-04-17       Impact factor: 2.102

6.  Analyses of the factors influencing the accuracy of three-dimensional ultrasound in comparison with cone-beam CT in image-guided radiotherapy for prostate cancer with or without pelvic lymph node irradiation.

Authors:  Sha Zhou; Liling Luo; Jibin Li; Maosheng Lin; Li Chen; Jianhui Shao; Shipei Lu; Yaru Ma; Yingting Zhang; Wenfen Chen; Mengzhong Liu; Shiliang Liu; Liru He
Journal:  Radiat Oncol       Date:  2019-01-29       Impact factor: 3.481

7.  Optimal parameters for clinical implementation of breast cancer patient setup using Varian DTS software.

Authors:  Sook Kien Ng; Piotr Zygmanski; Andrew Jeung; Hassan Mostafavi; Juergen Hesser; Jennifer R Bellon; Julia S Wong; Yulia Lyatskaya
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

8.  Peripheral dose from megavoltage cone-beam CT imaging for nasopharyngeal carcinoma image-guided radiation therapy.

Authors:  Ming X Jia; Xu Zhang; Na Li; En Y Wang; Da W Liu; Wei S Cai
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

Review 9.  The Use of Ultrasound Imaging in the External Beam Radiotherapy Workflow of Prostate Cancer Patients.

Authors:  Saskia M Camps; Davide Fontanarosa; Peter H N de With; Frank Verhaegen; Ben G L Vanneste
Journal:  Biomed Res Int       Date:  2018-01-24       Impact factor: 3.411

  9 in total