Detection of setup uncertainties with 3D surface registration system for conformal radiotherapy of breast cancer.

Letizia Deantonio1, Laura Masini1, Gianfranco Loi2, Giuseppina Gambaro1, Cesare Bolchini1, Marco Krengli3.   

Abstract

AIM: To investigate the clinical application of a technique for patient set-up verification in breast cancer radiotherapy based on a 3D surface image registration system.
BACKGROUND: Accurate and reproducible patient set-up is a prerequisite to correctly deliver fractionated radiotherapy. Various approaches are available to verify and correct patient setup for 3D image acquisition in a radiation treatment room.
MATERIALS AND METHODS: The study analyzed the setup reproducibility of 15 patients affected by breast cancer and candidates for conformal radiotherapy by using the AlignRT system (VisionRT, London, UK). At the initial setup, electronic portal imaging device (EPID) images were compared with Digitally Reconstructed Radiographs (DRRs) and a reference three-dimensional (3D) surface image was obtained by AlignRT. Surface images were acquired prior to every subsequent setup procedure. The systematic and random errors along longitudinal and vertical directions were measured and compared for the two systems.
RESULTS: The procedure for surface registration, image acquisition and comparison with the reference image took less than 1 min on average. The T test for systematic error showed no significant difference between the 2 verification systems along the longitudinal (p = 0.69) and vertical (p = 0.67) axes. The T-test for random error showed a significant difference between the 2 systems along the vertical axis (p = 0.05).
CONCLUSION: AlignRT is fast, simple, non-invasive and seems to be reliable in detecting patient setup errors. Our results suggest that it could be used to assess the setup reproducibility for breast cancer patients.

Entities:  

Keywords:  Breast cancer; Patient setup; Radiotherapy; Surface imaging

Year:  2011        PMID: 24376961      PMCID: PMC3863244          DOI: 10.1016/j.rpor.2011.02.003

Source DB:  PubMed          Journal:  Rep Pract Oncol Radiother        ISSN: 1507-1367


  17 in total

1.  Evaluation of ultrasound-based prostate localization for image-guided radiotherapy.

Authors:  K M Langen; J Pouliot; C Anezinos; M Aubin; A R Gottschalk; I-C Hsu; D Lowther; Y-M Liu; K Shinohara; L J Verhey; V Weinberg; M Roach
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-11-01       Impact factor: 7.038

2.  A performance comparison of flat-panel imager-based MV and kV cone-beam CT.

Authors:  B A Groh; J H Siewerdsen; D G Drake; J W Wong; D A Jaffray
Journal:  Med Phys       Date:  2002-06       Impact factor: 4.071

3.  An application framework for computer-aided patient positioning in radiation therapy.

Authors:  T Liebler; M Hub; C Sanner; W Schlegel
Journal:  Med Inform Internet Med       Date:  2003-09

4.  A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup.

Authors:  Christoph Bert; Katherine G Metheany; Karen Doppke; George T Y Chen
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

5.  Patient set-up verification by infrared optical localization and body surface sensing in breast radiation therapy.

Authors:  Maria Francesca Spadea; Guido Baroni; Marco Riboldi; Barbara Tagaste; Cristina Garibaldi; Roberto Orecchia; Antonio Pedotti
Journal:  Radiother Oncol       Date:  2006-03-30       Impact factor: 6.280

6.  Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients.

Authors:  Christoph Bert; Katherine G Metheany; Karen P Doppke; Alphonse G Taghian; Simon N Powell; George T Y Chen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-03-15       Impact factor: 7.038

7.  3D optoelectronic analysis of interfractional patient setup variability in frameless extracranial stereotactic radiotherapy.

Authors:  Guido Baroni; Cristina Garibaldi; Marco Riboldi; Maria F Spadea; Gianpiero Catalano; Barbara Tagaste; Giampiero Tosi; Roberto Orecchia; Antonio Pedotti
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-09-28       Impact factor: 7.038

8.  Effects of setup errors and shape changes on breast radiotherapy.

Authors:  Anke van Mourik; Simon van Kranen; Suzanne den Hollander; Jan-Jakob Sonke; Marcel van Herk; Corine van Vliet-Vroegindeweij
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-08       Impact factor: 7.038

9.  Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax.

Authors:  Philipp J Schöffel; Wolfgang Harms; Gabriele Sroka-Perez; Wolfgang Schlegel; Christian P Karger
Journal:  Phys Med Biol       Date:  2007-06-06       Impact factor: 3.609

Review 10.  Deep inspiration breath hold and respiratory gating strategies for reducing organ motion in radiation treatment.

Authors:  Gikas S Mageras; Ellen Yorke
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

View more
  10 in total

1.  The importance of accurate treatment planning, delivery, and dose verification.

Authors:  Julian Malicki
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

2.  Radiotherapeutic management of the axillae in early stage breast cancer: Perspective of the radiation oncologist.

Authors:  Arantxa Eraso; Ferran Guedea
Journal:  Rep Pract Oncol Radiother       Date:  2012-05-17

3.  Three-dimensional surface imaging for detection of intra-fraction setup variations during radiotherapy of pelvic tumors.

Authors:  Giuseppina Apicella; Gianfranco Loi; Sara Torrente; Silvia Crespi; Debora Beldì; Marco Brambilla; Marco Krengli
Journal:  Radiol Med       Date:  2016-06-14       Impact factor: 3.469

4.  Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (Catalyst™).

Authors:  F Walter; P Freislederer; C Belka; C Heinz; M Söhn; F Roeder
Journal:  Radiat Oncol       Date:  2016-11-24       Impact factor: 3.481

5.  Feasibility of surface guided radiotherapy for patient positioning in breast radiotherapy versus conventional tattoo-based setups- a systematic review.

Authors:  Wesley Naidoo; Michelle Leech
Journal:  Tech Innov Patient Support Radiat Oncol       Date:  2022-04-16

6.  Prone versus supine position for adjuvant breast radiotherapy: a prospective study in patients with pendulous breasts.

Authors:  Marco Krengli; Laura Masini; Tina Caltavuturo; Carla Pisani; Giuseppina Apicella; Eleonora Negri; Letizia Deantonio; Marco Brambilla; Giuseppina Gambaro
Journal:  Radiat Oncol       Date:  2013-10-08       Impact factor: 3.481

7.  Three-dimensional surface and ultrasound imaging for daily IGRT of prostate cancer.

Authors:  Marco Krengli; Gianfranco Loi; Carla Pisani; Debora Beldì; Giuseppina Apicella; Valentina Amisano; Marco Brambilla
Journal:  Radiat Oncol       Date:  2016-12-13       Impact factor: 3.481

8.  Surface imaging-based analysis of intrafraction motion for breast radiotherapy patients.

Authors:  David B Wiant; Stacy Wentworth; Jacqueline M Maurer; Caroline L Vanderstraeten; Jonathon A Terrell; Benjamin J Sintay
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

9.  Optical Surface Scanning for Patient Positioning in Radiation Therapy: A Prospective Analysis of 1902 Fractions.

Authors:  G Carl; D Reitz; S Schönecker; M Pazos; P Freislederer; M Reiner; F Alongi; M Niyazi; U Ganswindt; C Belka; S Corradini
Journal:  Technol Cancer Res Treat       Date:  2018-01-01

10.  Impact of use of optical surface imaging on initial patient setup for stereotactic body radiotherapy treatments.

Authors:  Brian Leong; Laura Padilla
Journal:  J Appl Clin Med Phys       Date:  2019-12       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.