Tsutomu Sugiura1, Kazuhiko Yamamoto1, Satoshi Horita1, Kazuhiro Murakami1, Tadaaki Kirita1.
Abstract
AIMS: The purposes of this study were to investigate the primary stability of cylindrical and tapered implants in different bone types by measuring implant displacement and to examine the relationship between insertion torque value (ITV) and implant displacement.Entities:
Keywords: Bone density; dental implants; torque
Year: 2019 PMID: 32308322 PMCID: PMC7150554 DOI: 10.4103/ccd.ccd_788_18
Source DB: PubMed Journal: Contemp Clin Dent ISSN: 0976-2361
Figure 1Different bone types of artificial bone. Bone blocks had a low-density or low- to medium-density cancellous bone layer with or without a 1-mm thick cortical bone layer
Figure 2(a) Implant design used in the present study. Cylindrical implant (left) and tapered implant (right). (b) Experimental setup
Differences in insertion torque value and implant displacement between cylindrical and tapered implants
| Cancellous bone density/cortical bone thickness | ITV (Ncm) | Displacement (µm) | ||||
|---|---|---|---|---|---|---|
| Cylindrical | Tapered | Cylindrical | Tapered | |||
| Low/0 mm | 4.3±0.9 | 4.6±0.7 | 0.382 | 372.6±52.9 | 345.1±38.7 | 0.233 |
| Low/1 mm | 10.2±2.8 | 13.4±5.9 | 0.307 | 183.3±29.1 | 120.7±22.9 | <0.01 |
| Low to medium/0 mm | 33.7±8.7 | 34.6±5.2 | 0.701 | 74.4±9.4 | 76.5±12.4 | 0.529 |
| Low to medium/1 mm | 34.6±4.8 | 45.8±10.7 | <0.01 | 64.1±6.3 | 58.1±9.3 | 0.025 |
Values are presented as the mean±SD. SD: Standard deviation; ITV: Insertion torque value
Insertion torque value in different bone types (Ncm)
| Cylindrical implant | Tapered implant | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Cortical bone thickness (mm) | Cancellous bone density (g/cm3) | LM/L | Cortical bone thickness (mm) | Cancellous bone density (g/cm3) | LM/L | ||||
| 0.16 | 0.32 | 0.16 | 0.32 | ||||||
| 0 | 4.3 | 33.7 | <0.01 | 7.84 | 0 | 4.6 | 34.6 | <0.01 | 7.52 |
| 1 | 10.2 | 34.6 | <0.01 | 3.39 | 1 | 13.4 | 45.8 | <0.01 | 3.42 |
| <0.01 | 0.561 | <0.01 | <0.01 | ||||||
| Tcor 1/Tcor 0 | 2.37 | 1.03 | Tcor 1/Tcor 0 | 2.91 | 1.32 | ||||
L: Low-density cancellous bone; LM: Low- to medium-density cancellous bone; Tcor 0: Cortical bone thickness of 0 mm; Tcor 1: Cortical bone thickness of 1 mm
Implant displacement in different bone types (μm)
| Cylindrical implant | Tapered implant | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| Cortical bone thickness (mm) | Cancellous bone density (g/cm3) | L/LM | Cortical bone thickness (mm) | Cancellous bone density (g/cm3) | L/LM | ||||
| 0.16 | 0.32 | 0.16 | 0.32 | ||||||
| 0 | 372.6 | 74.4 | <0.01 | 5.01 | 0 | 345.1 | 76.5 | <0.01 | 4.51 |
| 1 | 183.3 | 64.1 | <0.01 | 2.86 | 1 | 120.7 | 58.1 | <0.01 | 2.08 |
| <0.01 | <0.01 | <0.01 | <0.01 | ||||||
| Tcor 0/Tcor 1 | 2.03 | 1.16 | Tcor 0/Tcor 1 | 2.86 | 1.32 | ||||
L: Low-density cancellous bone; LM: Low- to medium-density cancellous bone; Tcor 0: Cortical bone thickness of 0 mm; Tcor 1: Cortical bone thickness of 1 mm
Figure 3Displacement/insertion torque plot. (a) Cylindrical implant. (b) Tapered implant